Development of MEMS Flow and Pressure Sensor Device for Detection of Extravasation at an Early Stage

被引:1
作者
Hasegawa, Yoshihiro [1 ]
Shimohira, Chihiro [1 ]
Matsushima, Miyoko [2 ]
Sugiyama, Tomoshi [3 ]
Kawabe, Tsutomu [2 ]
Shikida, Mitsuhiro [1 ]
机构
[1] Hiroshima City Univ, Grad Sch Informat Sci, Dept Biomed Informat Sci, Asaminami Ku, 3-4-1 Ozukahigashi, Hiroshima 7313194, Japan
[2] Nagoya Univ, Grad Sch Med, Dept Integrated Hlth Sci,Higashi Ku, Div Host Def Sci,Om Hlth Sci,Tokai Natl Higher Ed, 1-20 Daiko Minami, Nagoya, Aichi 4618673, Japan
[3] Gifu Prefectural Tajimi Hosp, Departmet Thorac Surg, 5-161 Maebatacho, Tajimi Gifu 5078522, Japan
关键词
intravenous injection; extravasation; flow sensor; pressure sensor;
D O I
10.1002/tee.23706
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed a flow and pressure measurement device for detection of extravasation by intravenous injection of anticancer drugs at an early stage. Intravenous injection of a drug solution using a syringe pump enables a low-speed and highly accurate drug-administration rate. However, because the syringe pump forcibly delivers the drug solution, the drug solution is administrated at a constant administration rate regardless of the leakage of the drug solution to the outside of the blood vessel (extravasation) due to the indwelling needle coming out of the blood vessel. The proposed device is integrated into a flow-channel structure consisting of a PDMS layer in which a MEMS thermal-flow sensor formed on thin polyimide film and a pressure sensor are stacked on a plastic plate as a substrate. Therefore, it is possible to measure both the drug-administration rate and injection pressure. Early detection of extravasation can be achieved by identifying the state of needle-puncture based on changes in injection pressure. Then, the amount of leaked drug after leak occurs can be accurately measured with a flow sensor. We evaluated the flow and pressure detection function of the proposed device and confirmed that it is possible to detect extravasation early through an experiment involving a piece of raw chicken simulating human skin and flesh. (c) 2022 Institute of Electrical Engineers of Japan. Published by Wiley Periodicals LLC.
引用
收藏
页码:139 / 146
页数:8
相关论文
共 21 条
  • [1] Al-Benna S, 2013, ISRN Dermatol, V2013, P856541, DOI 10.1155/2013/856541
  • [2] Wearable Sensor Patch for Early Extravasation Detection
    Cheng, Ming-Yuan
    Damalerio, Ramona
    Lim, Ruiqi
    Chen, Weiguo
    Tan, Kwan Ling
    Bong, Choon Looi
    Tan, Swee Kim
    [J]. 2016 IEEE 66TH ELECTRONIC COMPONENTS AND TECHNOLOGY CONFERENCE (ECTC), 2016, : 1632 - 1637
  • [3] Choi W, 2011, PROC IEEE MICR ELECT, P17, DOI 10.1109/MEMSYS.2011.5734351
  • [4] Du Y-C., 2017, NOVEL DEVICE NON INV, DOI [10.1109/ICSENS.2017.8234371, DOI 10.1109/ICSENS.2017.8234371]
  • [5] Gingerich MD, 2001, TRANSDUCERS '01: EUROSENSORS XV, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, P416
  • [6] Analysis of measurement conditions for detecting change in vital signs with catheter flow sensor
    Hasegawa, Y.
    Mitsunari, Y.
    Kawaoka, H.
    Matsushima, M.
    Kawabe, T.
    Shikida, M.
    [J]. JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2018, 28 (10)
  • [7] Heartbeat Signal Detection From Analysis of Airflow in Rat Airway Under Different Depths of Anaesthesia Conditions
    Kawaoka, Hidetaka
    Yamada, Takayuki
    Matsushima, Miyoko
    Kawabe, Tsutomu
    Hasegawa, Yoshihiro
    Shikida, Mitsuhiro
    [J]. IEEE SENSORS JOURNAL, 2017, 17 (14) : 4369 - 4377
  • [8] Leonardi M, 2003, BOSTON TRANSDUCERS'03: DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, P1043
  • [9] Development of a Wireless Sensor for the Measurement of Chicken Blood Flow Using the Laser Doppler Blood Flow Meter Technique
    Nishihara, Kei
    Iwasaki, Wataru
    Nakamura, Masaki
    Higurashi, Eiji
    Soh, Tomoki
    Itoh, Toshihiro
    Okada, Hironao
    Maeda, Ryutaro
    Sawada, Renshi
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (06) : 1645 - 1653
  • [10] Nozaka H., 2016, P 2016 C ASIA PACIFI, P94