Risk of cardiac implantable device malfunction in cancer patients receiving proton therapy: an overview

被引:5
作者
Mirzaei, Milad [1 ,2 ]
Rowshanfarzad, Pejman [3 ]
Gill, Suki [1 ,3 ]
Ebert, Martin A. [1 ,3 ]
Dass, Joshua [1 ]
机构
[1] Sir Charles Gairdner Hosp, Dept Radiat Oncol, Nedlands, WA, Australia
[2] Monash Univ, Sch Biomed Sci, Dept Med Imaging & Radiat Sci, Clayton, Vic, Australia
[3] Univ Western Australia, Sch Phys Math & Comp, Crawley, WA, Australia
来源
FRONTIERS IN ONCOLOGY | 2023年 / 13卷
关键词
proton therapy; cardiac implantable electronic device; malfunction; secondary neutrons; pacemakers; implantable cardioverter defibrillators; implantable loop recorders; pacing leads; PASSIVE CARBON-ION; RADIATION-THERAPY; IONIZING-RADIATION; PACEMAKER FAILURE; DOSE-EQUIVALENT; RADIOTHERAPY; SECONDARY; MANAGEMENT; INTERFERENCE; IRRADIATION;
D O I
10.3389/fonc.2023.1181450
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Age is a risk factor for both cardiovascular disease and cancer, and as such radiation oncologists frequently see a number of patients with cardiac implantable electronic devices (CIEDs) receiving proton therapy (PT). CIED malfunctions induced by PT are nonnegligible and can occur in both passive scattering and pencil beam scanning modes. In the absence of an evidence-based protocol, the authors emphasise that this patient cohort should be managed differently to electron- and photon- external beam radiation therapy (EBRT) patients due to distinct properties of proton beams. Given the lack of a PT-specific guideline for managing this cohort and limited studies on this important topic; the process was initiated by evaluating all PT-related CIED malfunctions to provide a baseline for future reporting and research. In this review, different modes of PT and their interactions with a variety of CIEDs and pacing leads are discussed. Effects of PT on CIEDs were classified into a variety of hardware and software malfunctions. Apart from secondary neutrons, cumulative radiation dose, dose rate, CIED model/manufacturer, distance from CIED to proton field, and materials used in CIEDs/pacing leads were all evaluated to determine the probability of malfunctions. The importance of proton beam arrangements is highlighted in this study. Manufacturers should specify recommended dose limits for patients undergoing PT. The establishment of an international multidisciplinary team dedicated to CIED-bearing patients receiving PT may be beneficial.
引用
收藏
页数:14
相关论文
共 101 条
  • [1] Cardiac Implantable Electronic Miniaturized and Micro Devices
    Acha, Moshe Ray
    Soifer, Elina
    Hasin, Tal
    [J]. MICROMACHINES, 2020, 11 (10)
  • [2] [Anonymous], 2011, RAD THER BIOTRONIK C
  • [3] [Anonymous], 2010, PACEMAKERS IMPLANTAB
  • [4] The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
    Aslian, Hossein
    Kron, Tomas
    Watts, Troy
    Akalanli, Cagla
    Hardcastle, Nicholas
    Lonski, Peta
    Montaseri, Atousa
    Hay, Barry
    Korte, James
    Berk, Kemal
    Longo, Francesco
    Severgnini, Mara
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (06): : 121 - 131
  • [5] Production and subsequent separation of 47Sc of nuclear medicine applications using neutron-induced reactions on different natural targets
    Attallah, Mohamed F.
    Mohamed, Gehan Y.
    Breky, Mohamed M. E.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2022, 331 (04) : 1723 - 1730
  • [6] Current clinical practice in patients with cardiac implantable electronic devices undergoing radiotherapy: a literature review
    Azraai, Meor
    D'Souza, Daniel
    Lin, Yuan-Hong
    Nadurata, Voltaire
    [J]. EUROPACE, 2022, 24 (03): : 362 - 374
  • [7] Radiotherapy-Induced Cardiac Implantable Electronic Device Dysfunction in Patients With Cancer
    Bagur, Rodrigo
    Chamula, Mathilde
    Brouillard, Emilie
    Lavoie, Caroline
    Nombela-Franco, Luis
    Julien, Anne-Sophie
    Archambault, Louis
    Varfalvy, Nicolas
    Gaudreault, Valerie
    Joncas, Sebastien X.
    Israeli, Zeev
    Parviz, Yasir
    Mamas, Mamas A.
    Lavi, Shahar
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2017, 119 (02) : 284 - 289
  • [8] A Monte Carlo-based analytic model of neutron dose equivalent for a mevion gantry-mounted passively scattered proton system for craniospinal irradiation
    Baradaran-Ghahfarokhi, Milad
    Reynoso, Francisco
    Sun, Baozhou
    Darafsheh, Arash
    Prusator, Michael T.
    Mutic, Sasa
    Zhao, Tianyu
    [J]. MEDICAL PHYSICS, 2020, 47 (09) : 4509 - 4521
  • [9] The impact of particle radiotherapy on the functioning of cardiac implantable electronic devices: a systematic review of in vitro and in vivo studies according to PICO criteria
    Barcellini, Amelia
    Dusi, Veronica
    Mirandola, Alfredo
    Ronchi, Sara
    Riva, Giulia
    Dal Mas, Francesca
    Massaro, Maurizio
    Vitolo, Viviana
    Ciocca, Mario
    Rordorf, Roberto
    Orlandi, Ester
    [J]. RADIOLOGIA MEDICA, 2022, 127 (09): : 1046 - 1058
  • [10] BAUMANN R, 1995, 1995 IEEE INTERNATIONAL RELIABILITY PHYSICS PROCEEDINGS, 33RD ANNUAL, P297, DOI 10.1109/RELPHY.1995.513695