Effect of aortic curvature on bioprosthetic aortic valve performance

被引:3
|
作者
Vogl, Brennan [1 ]
Gadhave, Rajat [1 ]
Wang, Zhenyu [2 ,3 ]
El Shaer, Ahmed [4 ]
Ponce, Alejandra Chavez [4 ]
Alkhouli, Mohamad [4 ]
Hatoum, Hoda [1 ,5 ]
机构
[1] Michigan Technol Univ, Dept Biomed Engn, Houghton, MI 49931 USA
[2] Ohio State Univ, Dept Mech Engn, Columbus, OH USA
[3] Ohio State Univ, Simulat Innovat & Modeling Ctr, Columbus, OH USA
[4] Mayo Clin, Dept Cardiovasc Med, Rochester, MN USA
[5] Michigan Technol Univ, Hlth Res Inst, Inst Comp & Cybernet, Ctr Biocomp & Digital Hlth, Houghton, MI USA
关键词
Aortic valve; Bioprosthetic valve; Aortic curvature; Pressure gradient; Pressure recovery; PRESSURE RECOVERY; DOPPLER ULTRASOUND; CATHETER GRADIENTS; STENOSIS; IMPACT; OVERESTIMATION; MANIFESTATION; REPLACEMENT; TRENDS; AREA;
D O I
10.1016/j.jbiomech.2022.111422
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Transvalvular pressure gradient (AP) after aortic valve replacement is an important surrogate of aortic bio-prostheses performance. Invasive AP is often measured after transcatheter aortic valve replacement to exclude patient-prosthetic mismatch. However, invasive aortic pressures are usually recorded in the pressure recovery (PR) zone downstream of the valve, potentially resulting in AP underestimation compared to noninvasive measurements. PR was extensively studied in straight ascending aortas. However, the impact of various aortic arch configurations on AP has not been explored. PR was assessed in a pulse duplicating simulator at various cardiac conditions of cardiac output, heart rates and pressures. Three different aortic geometries with identical root dimensions but with different aortic arches were used: (1) curvature 1, (2) curvature 2, and (3) straight aortic models. Instantaneous pressure and peak AP measurements were recorded incrementally along the models for each cardiac condition. The models with aortic arches produced two distinct PR zones (after the valve and after the aortic arch), whereas the model without an aortic arch produced only one PR zone (after the valve). The trend of the pressure and AP curves for each model was independent of the cardiac condition used, but the individually measured pressure magnitudes did change with different conditions. In this study, we illustrated the differences in PR between distinct aortic curvatures and straight aorta. PR affects pressure and AP measurements. These effects are clear when recording aortic pressures by catheterization and echocardiography.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Bioprosthetic Aortic Valve Endocarditis in Association with Enterococcus durans
    Fallavollita, Luca
    Di Gioacchino, Lorena
    Balestrini, Fabrizio
    TEXAS HEART INSTITUTE JOURNAL, 2016, 43 (02): : 165 - 167
  • [42] Leaflet Kinematics of Mechanical and Bioprosthetic Aortic Valve Prostheses
    Vennemann, Bernhard
    Roesgen, Thomas
    Heinisch, Paul Philipp
    Obrist, Dominik
    ASAIO JOURNAL, 2018, 64 (05) : 651 - 661
  • [43] Antithrombotic therapy following bioprosthetic aortic valve replacement
    Nowell, Justin
    Wilton, Emma
    Markus, Hugh
    Jahangiri, Marjan
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2007, 31 (04) : 578 - 585
  • [44] Association of Bioprosthetic Aortic Valve Leaflet Calcification on Hemodynamic and Clinical Outcomes
    Zhang, Bin
    Salaun, Erwan
    Cote, Nancy
    Wu, Yongjian
    Mahjoub, Haifa
    Mathieu, Patrick
    Dahou, Abdellaziz
    Zenses, Anne-Sophie
    Clisson, Marine
    Pibarot, Philippe
    Clavel, Marie-Annick
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 76 (15) : 1737 - 1748
  • [45] Transcatheter valve-in-valve versus redo surgical aortic valve replacement for the treatment of degenerated bioprosthetic aortic valve: A systematic review and meta-analysis
    Tam, Derrick Y.
    Vo, Thin X.
    Wijeysundera, Harindra C.
    Dvir, Danny
    Friedrich, Jan O.
    Fremes, Stephen E.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2018, 92 (07) : 1404 - 1411
  • [46] Bioprosthetic Aortic Paravalvular Leak: Is Valve-in-Valve Another Solution?
    Tandar, Anwar
    Bull, David A.
    Welt, Frederick G. P.
    JOURNAL OF INVASIVE CARDIOLOGY, 2017, 29 (01) : E1 - E7
  • [47] The Perceval Sutureless Bioprosthetic Aortic Valve: Evolution of Surgical Valve Technology
    Fialka, Nicholas M.
    EL-Andari, Ryaan
    Wang, Shaohua
    Dokollari, Aleksander
    Kent, William D. T.
    Hassanabad, Ali Fatehi
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2024, 19 (02) : 125 - 135
  • [48] Bioprosthetic valve thrombosis developed shortly after aortic valve replacement
    Kuwano, Akito
    Yoshikai, Masaru
    Sato, Hisashi
    ASIAN CARDIOVASCULAR & THORACIC ANNALS, 2025, : 52 - 54
  • [49] Subclinical valve leaflet thrombosis following bioprosthetic aortic valve replacement
    Giuliani, Carlos
    Zanuttini, Antonela
    Salaun, Erwan
    Cote, Nancy
    Clavel, Marie Annick
    Pibarot, Philippe
    CURRENT OPINION IN CARDIOLOGY, 2024, 39 (05) : 457 - 464
  • [50] Imaging of Bioprosthetic Valve Dysfunction after Transcatheter Aortic Valve Implantation
    Alwan, Louhai
    Bernhard, Benedikt
    Brugger, Nicolas
    de Marchi, Stefano F.
    Praz, Fabien
    Windecker, Stephan
    Pilgrim, Thomas
    Grani, Christoph
    DIAGNOSTICS, 2023, 13 (11)