Electromagnetically spinning viscometer designed for measurement of low viscosity in low shear rate region

被引:0
|
作者
Hosoda, M. [1 ]
Yamakawa, Y. [2 ]
Sakai, K. [3 ,4 ]
机构
[1] Tokyo Denki Univ, Hatoyama, Saitama 3500394, Japan
[2] Triple Eye Co Ltd, Shimomaruyacho,Nakagyo Ku, Kyoto 6048006, Japan
[3] Univ Tokyo, Inst Ind Sci, Meguro Ku, Tokyo 1538505, Japan
[4] Univ Tokyo, Fac Engn, Bunkyo Ku, Tokyo 1138656, Japan
基金
日本学术振兴会;
关键词
viscosity; electromagnetically spinning method; low viscosity; BLOOD-VISCOSITY;
D O I
10.35848/1347-4065/ad2d0b
中图分类号
O59 [应用物理学];
学科分类号
摘要
The electromagnetically spinning method for measurement of fluid viscosity was improved to obtain accurate values of low viscosity in the low shear rate region. Harmful effects derived from the mechanical friction to the smooth rotation of the rotor are thoroughly eliminated by employing a viscosity probe suspended by a thin metal wire. In our previous study, the motion of the meniscus of the sample surface also acted as a troublesome resistant torque for probe rotation, which was addressed by employing a sandwiched structure of the disk probe between the bottom and top plates. The measurements were carried out in two procedures. In the freely oscillating operation, we could measure the viscosity of the atmosphere with a viscosity of approximately 1/100 of that of water. The second type of quasi-steady measurement enabled a measurement of pure water in the range of shear deformation rates smaller than 1 s-1.