Open Port Interface for Coupling Capillary Electrophoresis and Mass Spectrometry: Performance Evaluation for Capillary Isoelectric Focusing

被引:0
作者
Beloborodov, Stanislav S. [1 ]
Schneider, Bradley B. [1 ]
Oleschuk, Richard D. [2 ]
Le Blanc, J. C. Yves [1 ]
机构
[1] SCIEX Ltd, Vaughan, ON L4K 4V8, Canada
[2] Queens Univ, Dept Chem, Kingston, ON K7L 3N6, Canada
关键词
ZONE-ELECTROPHORESIS; ELECTROSPRAY INTERFACE; SAMPLING INTERFACE; LIQUID-JUNCTION; HIGH-EFFICIENCY; CE; SEPARATION; NANOLITER; PROTEINS; POINTS;
D O I
10.1021/jasms.3c00060
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Capillary electrophoresis (CE) combined with mass spectrometry (MS) is a powerful analytical technique that utilizes the resolving power of CE and the mass-detection capabilities of MS. In many cases, CE is coupled to MS via a sheath-flow interface (SFI). This interface has a simple design and can be easily constructed; however, it often suffers from issues such as MS signal suppression, interference of MS and CE electrical circuits, and the inability to set an optical point of detection close to the capillary end due to the specific design of the coupling union. In this paper, we describe a novel coupling of CE and MS based upon the open port interface (OPI). The OPI differs from classical sheath flow interfaces by operating at flow rates at least 1 order of magnitude higher. In addition to the flow rate difference, the OPI provides more efficient mixing of the capillary eluates with the transport fluid and thus minimizes MS signal suppression. In this work, we compared the performance of OPI and SFI in a series of capillary isoelectric focusing (cIEF) experiments with 5 pI markers, carbonic anhydrase II and NIST antibody. The evaluation criteria for the comparison of the OPI and SFI were analytical sensitivity, reproducibility, and pI marker linearity. Given the extent of sample dilution in the OPI, we also compared the peak resolution determined using an upstream UV detector to those determined by the downstream mass spectrometer. The results suggested that the OPI configuration reduced signal suppression, with no adverse effect on peak resolution. In addition, the OPI provided better decoupling of the CE and MS potentials as well as reduced signal dependence upon the sheath liquid composition. While these results are preliminary, they suggest that the OPI is a viable approach for CE-MS coupling.
引用
收藏
页码:2107 / 2116
页数:10
相关论文
共 62 条
  • [1] Electrochemical properties of capillary electrophoresis-nanoelectrospray mass spectrometry
    Bateman, KP
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (04) : 309 - 317
  • [2] Bateman KP, 1997, RAPID COMMUN MASS SP, V11, P307
  • [3] Batruch, SIMPLIFIED ORTHOGONA
  • [4] Laser dissection sampling modes for direct mass spectral analysis
    Cahill, John F.
    Kertesz, Vilmos
    Van Berkel, Gary J.
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2016, 30 (05) : 611 - 619
  • [5] CAPILLARY ELECTROPHORESIS MASS-SPECTROMETRY
    CAI, JY
    HENION, J
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) : 667 - 692
  • [6] Practical implications of some recent studies in electrospray ionization fundamentals
    Cech, NB
    Enke, CG
    [J]. MASS SPECTROMETRY REVIEWS, 2001, 20 (06) : 362 - 387
  • [7] Recent advances (2019-2021) of capillary electrophoresis-mass spectrometry for multilevel proteomics
    Chen, Daoyang
    McCool, Elijah N.
    Yang, Zhichang
    Shen, Xiaojing
    Lubeckyj, Rachele A.
    Xu, Tian
    Wang, Qianjie
    Sun, Liangliang
    [J]. MASS SPECTROMETRY REVIEWS, 2023, 42 (02) : 617 - 642
  • [8] Tapered-Tip Capillary Electrophoresis Nano-Electrospray Ionization Mass Spectrometry for Ultrasensitive Proteomics: the Mouse Cortex
    Choi, Sam B.
    Zamarbide, Marta
    Manzini, M. Chiara
    Nemes, Peter
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2017, 28 (04) : 597 - 607
  • [9] Covey T. R., 2010, ELECTROSPRAY MALDI M, P441
  • [10] Dai Jun, 2022, Methods Mol Biol, V2500, P55, DOI 10.1007/978-1-0716-2325-1_6