SPHIRD-Single Photon Counting Pixel Readout ASIC With Pulse Pile-Up Compensation Methods

被引:12
作者
Grybos, P. [1 ]
Kleczek, R. [1 ]
Kmon, P. [1 ]
Otfinowski, P. [1 ]
Fajardo, P. [2 ]
Magalhaes, D. [2 ]
Ruat, M. [2 ]
机构
[1] AGH Univ Sci & Technol, Dept Measurement & Elect, PL-30059 Krakow, Poland
[2] European Synchrotron Radiat Facil, Detector & Elect Grp, F-38043 Grenoble, France
关键词
Photonics; Detectors; Integrated circuits; Discharges (electric); Transistors; Resistance; Prototypes; X-ray imaging; high count-rate; single photon counting SPC; hybrid pixel detector; front-end electronics; charge sensitive amplifier CSA; DETECTORS; CIRCUIT; TIME;
D O I
10.1109/TCSII.2023.3267859
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This brief presents the design and measurement results of a prototype SPHIRD-1 ASIC in the CMOS 40 nm process. The chip is dedicated to high count rate single photon counting operation at the European Synchrotron Radiation Facility with Extremely Brilliant Source. The core of the prototype IC is the matrix of 64 x 32 pixels of 50 mu m pitch. Each pixel contains a Charge Sensitive Amplifier (CSA) with a fast discharge block and a detector leakage current compensation circuit. The CSA is connected to a set of three discriminators. The readout channel is equipped with additional circuits, which provide different pulse pile-up compensation methods. The priority of the analog front-end electronics is to process the input signal in a short time (the CSA output pulse time width is only 18 ns) and to keep low noise (the equivalent noise charge at the level of 188 el. rms) with a power consumption equals to 26 mu W/pixel. The chip is optimized for operation with a monochromatic X-ray beam with an energy of up to 30 keV. The measurements prove the possibility of counting up to 11.5 Gcps/mm(2) based on a 10% dead time loss input rate parameter.
引用
收藏
页码:3248 / 3252
页数:5
相关论文
共 16 条
[1]   Photon Counting Detectors for X-Ray Imaging With Emphasis on CT [J].
Ballabriga, R. ;
Alozy, J. ;
Bandi, F. N. ;
Campbell, M. ;
Egidos, N. ;
Fernandez-Tenllado, J. M. ;
Heijne, E. H. M. ;
Kremastiotis, I. ;
Llopart, X. ;
Madsen, B. J. ;
Pennicard, D. ;
Sriskaran, V. ;
Tlustos, L. .
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (04) :422-440
[2]   The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging [J].
Ballabriga, R. ;
Alozy, J. ;
Blaj, G. ;
Campbell, M. ;
Fiederle, M. ;
Frojdh, E. ;
Heijne, E. H. M. ;
Llopart, X. ;
Pichotka, M. ;
Procz, S. ;
Tlustos, L. ;
Wong, W. .
JOURNAL OF INSTRUMENTATION, 2013, 8
[3]   Time-over-threshold readout to enhance the high flux capabilities of single-photon-counting detectors [J].
Bergamaschi, Anna ;
Dinapoli, Roberto ;
Greiffenberg, Dominic ;
Henrich, Beat ;
Johnson, Ian ;
Mozzanica, Aldo ;
Radicci, Valeria ;
Schmitt, Bernd ;
Shi, Xintian ;
Stoppani, Laura .
JOURNAL OF SYNCHROTRON RADIATION, 2011, 18 :923-929
[4]   IBEX: Versatile Readout ASIC With Spectral Imaging Capability and High Count Rate Capability [J].
Bochenek, M. ;
Bottinelli, S. ;
Broennimann, Ch. ;
Livi, P. ;
Loeliger, T. ;
Radicci, V. ;
Schnyder, R. ;
Zambon, P. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2018, 65 (06) :1285-1291
[5]   EIGER characterization results [J].
Dinapoli, Roberto ;
Bergamaschi, Anna ;
Greiffenberg, Dominic ;
Henrich, Beat ;
Horisberger, Roland ;
Johnson, Ian ;
Mozzanica, Aldo ;
Radicci, Valeria ;
Schmitt, Bernd ;
Shi, Xintian ;
Tinti, Gemma .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 731 :68-73
[6]   ACTIVE D-C RESTORATION IN NUCLEAR PULSE SPECTROMETRY [J].
GERE, EA ;
MILLER, GL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1967, NS14 (01) :89-&
[7]   32k Channel Readout IC for Single Photon Counting Pixel Detectors with 75 μm Pitch, Dead Time of 85 ns, 9 e- rms Offset Spread and 2% rms Gain Spread [J].
Grybos, P. ;
Kmon, P. ;
Maj, P. ;
Szczygiel, R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (02) :1155-1161
[8]   Pole-zero cancellation circuit with pulse pile-up tracking system for low noise charge-sensitive amplifiers [J].
Grybos, Pawel ;
Szczygiel, Robert .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (01) :583-590
[9]  
Kleczek R., 2018, P 11 INT M FRONT END, P1
[10]   Single Photon-Counting Pixel Readout Chip Operating up to 1.2 Gcps/mm2 for Digital X-Ray Imaging Systems [J].
Kleczek, Rafal ;
Grybos, Pawel ;
Szczygiel, Robert ;
Maj, Piotr .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2018, 53 (09) :2651-2662