摘要
Purpose There are limited studies in the literature comparing the alignment accuracy of the large console, imageless, computer-assisted navigation (CAN) and portable, hand-held, accelerometer-based navigation (ABN) in total knee arthroplasty (TKA). This study was aimed to compare the operative time, blood loss, radiological, clinical and functional outcomes between CAN- and ABN-guided bilateral TKA. Methods From Jan 2016 to Dec 2017, 50 patients who underwent bilateral TKA were randomized to undergo either CAN-guided or ABN-guided TKA. Tourniquet time and blood loss were recorded, and intra-op complications were noted. Post-op radiological outcomes at 2 weeks were compared between the groups. The clinical and functional outcomes using the American Knee Society Scores (KSS) and Oxford Knee Score (OKS) were recorded pre-operatively and post-operatively at 3, 6, 12 months and at the end of the study with a minimum follow-up of 48 months. Results Both groups were well-matched in terms of patient demographic parameters. The mean surgical time per knee was significantly lower in the ABN group (54.5 +/- 5.6 min) compared to the CAN group (61.7 +/- 13.7 min; p < 0.01). Mean blood loss per knee in the ABN group was 592.1 +/- 245.3 mL compared to 682.8 +/- 322.0 in the CAN group (p = 0.11). In the ABN group, the mean post-op mechanical axis was 1.2 +/- 3.2 degrees (vs 1.5 +/- 2.2 degrees in the CAN group, p 0.6), the frontal femoral angle was 88.8 +/- 2.3 degrees (vs 88.8 +/- 1.8 degrees in the CAN group, p 1.0) and frontal tibial angle was 90.1 +/- 1.6 degrees (vs 89.7 +/- 1.1 degrees in the CAN group, p 0.14). At 48 month follow-up, the mean functional KSS in the ABN group was 89.0 +/- 5.7 (vs 88.1 +/- 4.5 in the CAN group, p 0.37) and the mean OKS was 40.5 +/- 2.8 (vs 39.6 +/- 3.2 in the CAN group, p 0.12). Conclusion Portable, hand-held ABN offers alignment accuracy and functional outcomes in TKA similar to that with CAN, with a reduced duration of surgery. There was no advantage of either of the techniques in terms of clinical or functional outcomes at 48 month follow-up.
引用
收藏
页码:786 / 792
页数:7
相关论文
共 50 条
  • [21] Accuracy of A Handheld Accelerometer-Based Navigation System for Femoral and Tibial Resection in Total Knee Arthroplasty
    Huang, Eddie H.
    Copp, Steven N.
    Bugbee, William D.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (11) : 1906 - 1910
  • [22] An accelerometer-based portable navigation system improved prosthetic alignment after total knee arthroplasty in 3D measurements
    Ueyama, Hideki
    Minoda, Yukihide
    Sugama, Ryo
    Ohta, Yoichi
    Yamamura, Kazumasa
    Nakamura, Suguru
    Takemura, Susumu
    Nakamura, Hiroaki
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2019, 27 (05) : 1580 - 1586
  • [23] Technical Considerations and Accuracy Improvement of Accelerometer-Based Portable Computer Navigation for Performing Distal Femoral Resection in Total Knee Arthroplasty
    Fujimoto, Eisaku
    Sasashige, Yoshiaki
    Nakata, Kyouhei
    Yokota, Gen
    Omoto, Takenori
    Ochi, Mitsuo
    JOURNAL OF ARTHROPLASTY, 2017, 32 (01) : 53 - 60
  • [24] Extramedullary Guides Versus Portable, Accelerometer-Based Navigation for Tibial Alignment in Total Knee Arthroplasty: A Randomized, Controlled Trial: Winner of the 2013 HAP PAUL Award
    Nam, Denis
    Cody, Elizabeth A.
    Nguyen, Joseph T.
    Figgie, Mark P.
    Mayman, David J.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (02) : 288 - 294
  • [25] Accelerometer-Based Portable Navigation System Is Useful for Tibial Bone Cutting in Modified Kinematically Aligned Total Knee Arthroplasty
    Tsubosaka, Masanori
    Kamenaga, Tomoyuki
    Kuroda, Yuichi
    Takayama, Koji
    Hashimoto, Shingo
    Ishida, Kazunati
    Hayashi, Shinya
    Kuroda, Ryosuke
    Matsumoto, Tomoyuki
    JOURNAL OF KNEE SURGERY, 2021, 34 (08) : 870 - 876
  • [26] Accuracy and outcome of a handheld accelerometer-based navigation device compared to conventional alignment method in total knee arthroplasty in a Chinese population
    Man, Sheryl Lok-Chi
    Chau, Wai-Wang
    Zheng, Xieyang
    Ong, Michael Tim-Yun
    Ho, Kevin Ki-Wai
    JOURNAL OF ORTHOPAEDICS TRAUMA AND REHABILITATION, 2024, 31 (01) : 34 - 47
  • [27] Efficacy of Portable Accelerometer-Based Navigation Devices versus Conventional Guides in Total Knee Arthroplasty: A Meta-analysis
    Sun, Han
    Li, Shuxiang
    Wang, Kun
    Wu, Guofeng
    Zhou, Jian
    Sun, Xiaoliang
    JOURNAL OF KNEE SURGERY, 2020, 33 (07) : 691 - 703
  • [28] Clinical outcomes of computer-assisted total knee arthroplasty using pinless navigation
    Jiang, Lei
    Chen, Jerry Yongqiang
    Chong, Hwei-Chi
    Chia, Shi-Lu
    Lo, Ngai-Nung
    Yeo, Seng Jin
    JOURNAL OF ORTHOPAEDIC SURGERY, 2017, 25 (01):
  • [29] Cadaveric analysis of an accelerometer-based portable navigation device for distal femoral cutting block alignment in total knee arthroplasty
    Nam, Denis
    Jerabek, Seth A.
    Cross, Michael B.
    Mayman, David J.
    COMPUTER AIDED SURGERY, 2012, 17 (04) : 205 - 210
  • [30] Does computer-assisted surgical navigation total knee arthroplasty reduce venous thromboembolism compared with conventional total knee arthroplasty?
    Ooi, L. H.
    Lo, N. N.
    Yeo, S. J.
    Ong, B. C.
    Ding, Z. P.
    Lefi, A.
    SINGAPORE MEDICAL JOURNAL, 2008, 49 (08) : 610 - 614