Stem size and stem alignment affects periprosthetic fracture risk and primary stability in cementless total hip arthroplasty

被引:4
|
作者
Konow, Tobias [1 ]
Glismann, Katja [1 ]
Lampe, Frank [2 ]
Ondruschka, Benjamin [3 ]
Morlock, Michael M. [1 ]
Huber, Gerd [1 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, Denickestr 15, D-21073 Hamburg, Germany
[2] Asklepios Klin Barmbek, Hamburg, Germany
[3] Univ Med Ctr Hamburg Eppendorf, Inst Legal Med, Hamburg, Germany
关键词
periprosthetic fracture; preoperative planning; primary stability; stem alignment; stem size; FEMORAL FRACTURES; COMPONENTS;
D O I
10.1002/jor.25729
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The ideal stem size and stem position is important for the success of total hip arthroplasty, since it can affect early implant loosening and periprosthetic fractures (PPF). This study aimed to investigate how small deviations from the ideal stem size and position influences the PPF risk and primary stability. Six experienced surgeons performed preoperative templating based on which the benchmark size for each femur was determined. Consecutive implantations were performed in six cadaveric femur pairs-one side was implanted with an undersized stem followed by the benchmark size and the contralateral side with a benchmark size followed by an oversized stem (Corail, Depuy Synthes). Moreover, three different alignments (six varus, six neutral, six valgus-undersized) were compared using 18 femurs. Cortical strains during broaching and implantation were measured, and laser scans were used to determine final stem position. All specimens underwent dynamic loading. Primary stability was estimated from stem subsidence and pull-out forces. Templated stem size varied between surgeons (+/- 1 size; p = 0.005). Undersizing increased stem subsidence by 320% (p < 0.001). Oversized stems exhibited 52% higher pull-out forces (p = 0.001) and 240% higher cortical strains (p = 0.056). Cortex strains increased with varus alignment (R-2 = 0.356, p = 0.011) while primary stability decreased with valgus stem alignment (p = 0.043). Surgeons should be aware that small deviations from the ideal stem size and malalignments of the stem can significantly alter the mechanical situation and affect the success of their surgery.
引用
收藏
页码:829 / 836
页数:8
相关论文
共 50 条
  • [31] PLS cementless stem for total hip arthroplasty. Early results
    Bistolfi, A.
    Ravera, L.
    Giordano, A.
    Cimino, A.
    Collo, G.
    Massazza, G.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2013, 64 (06) : 589 - 595
  • [32] Periprosthetic fracture between a constrained total knee arthroplasty and a long-stem total hip arthroplasty - Treatment with a novel device
    Lachiewicz, Paul F.
    JOURNAL OF ARTHROPLASTY, 2007, 22 (03): : 449 - 452
  • [33] Stem alignment with short tapered-wedge cementless stems affects bone mineral density continuously for at least 5 years after primary total hip arthroplasty
    Masuda, Sho
    Iwakiri, Kentaro
    Ohta, Yoichi
    Minoda, Yukihide
    Kobayashi, Akio
    Nakamura, Hiroaki
    HIP INTERNATIONAL, 2024, 34 (01) : 57 - 65
  • [34] Single Wedge Femoral Stem Designs are Associated With a Higher Risk for Revision After Cementless Primary Total Hip Arthroplasty
    Reddy, Nithin C.
    Chang, Richard N.
    Prentice, Heather A.
    Paxton, Elizabeth W.
    Kelly, Matthew P.
    Khatod, Monti
    JOURNAL OF ARTHROPLASTY, 2023, 38 (05): : 855 - +
  • [35] Risk factors for intraoperative calcar fracture in cementless total hip arthroplasty
    Miettinen, Simo S. A.
    Makinen, Tatu J.
    Kostensalo, Inari
    Mekela, Keijo
    Huhtala, Heini
    Kettunen, Jukka S.
    Remes, Ville
    ACTA ORTHOPAEDICA, 2016, 87 (02) : 113 - 119
  • [36] Treatment of periprosthetic femoral fracture after cementless total hip arthroplasty with Ilizarov external fixation
    Sakai, Takashi
    Ohzono, Kenji
    Nakase, Takanobu
    Lee, Seung Bak
    Manaka, Tomoya
    Nishihara, Shunsaku
    JOURNAL OF ARTHROPLASTY, 2007, 22 (04): : 617 - 620
  • [37] Periprosthetic Femoral Fractures in Cementless Short Versus Straight Stem Total Hip Arthroplasty: A Propensity Score Matched Analysis
    Luger, Matthias
    Feldler, Sandra
    Pisecky, Lorenz
    Klasan, Antonio
    Gotterbarm, Tobias
    Schopper, Clemens
    JOURNAL OF ARTHROPLASTY, 2023, 38 (04): : 751 - 756
  • [38] Influence of Stem Size on Clinical Outcome of Primary Total Hip Arthroplasty with Cementless Extensively Porous-Coated Femoral Components
    Engh, C. Anderson, Jr.
    Mohan, Vivek
    Nagowski, Jarrod P.
    Terefenko, Christi J. Sychterz
    Engh, Charles A.
    JOURNAL OF ARTHROPLASTY, 2009, 24 (04): : 554 - 559
  • [39] Primary stability of an anatomical cementless hip stem: A statistical analysis
    Viceconti, Marco
    Brusi, Giovanni
    Pancanti, Alberto
    Cristofolini, Luca
    JOURNAL OF BIOMECHANICS, 2006, 39 (07) : 1169 - 1179
  • [40] Biomechanical evaluation of retrograde docking nailing to a total hip arthroplasty stem in a periprosthetic femur fracture model
    Lenz, Mark
    Hofmann-Fliri, Ladina
    Kasper, Lena Anna
    Varga, Peter
    Zderic, Ivan
    Gehweiler, Dominic
    Klos, Kajetan
    Hofmann, Gunther O.
    Stoffel, Karl
    Gueorguiev, Boyko
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2021, 52 (01): : 53 - 59