Transvenous endocardial pacing with SelectSecure™ 3830 lead in pediatric patients: case series of two infants and a literature review

被引:0
|
作者
Yang, Chuan [1 ]
Qi, Jing [1 ]
Alam, Mahmood [1 ]
Zou, Deling [1 ]
机构
[1] China Med Univ, Dept Cardiol, Shengjing Hosp, 36 Sanhao St, Shenyang 110004, Peoples R China
关键词
Infants; Endocardial pacing; 3830; lead; Lumenless; Right ventricular septal pacing; CONGENITAL HEART; CHILDREN; LUMENLESS; PERFORMANCE; EXPERIENCE; PACEMAKER; IMPLANTATION; THERAPY; ADULTS;
D O I
10.1186/s12872-024-03820-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The SelectSecure (TM) 3830 lead is an innovative, lumenless, and thin active fixed lead with a nonretractable screw-in tip and a diameter of 4.1 Fr, making it the thinnest pacing lead available. Its high anti-extrusion properties and durability have shown favorable outcomes in cardiac pacing, especially in pediatric patients. The superfine design and easy implantation of the lead have rendered it a preferred choice in children, particularly in cases of congenital heart disease. Case presentation This case series presents two infant patients who underwent transvenous endocardial pacing using the SelectSecure (TM) 3830 lead, along with a comprehensive literature review on the topic. The study followed the patients for 5 years and 3 years, respectively, and observed stable pacing parameters, indicating a positive therapeutic outcome and safety.This article discusses the optimal age and body shape for transvenous lead implantation in infants and highlights the advantages and disadvantages of endocardial and epicardial pacing approaches. Although endocardial pacing offers several benefits such as minimal trauma, short hospital stay, and longer battery life, it may not be suitable for intracardiac shunts, and venous occlusion remains a concern. On the other hand, epicardial pacing may be considered for children with challenging endocardial access but comes with higher risk of lead failure and coronary artery compression. This study emphasizes the importance of careful follow-up in pediatric patients with pacing, as lead failure can occur in young patients owing to growth and development, leading to syncope and battery depletion. The article also underscores the significance of selecting the appropriate pacing location to minimize the impact of cardiac function, with right ventricular septal pacing emerging as a preferable option. Conclusions The SelectSecure (TM) 3830 lead presents a promising solution for transvenous endocardial pacing in pediatric patients with high degree atrioventricular block and bradycardia, ensuring safe and effective pacing as they grow and develop.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Transvenous endocardial pacing with SelectSecure™ 3830 lead in pediatric patients: case series of two infants and a literature review
    Chuan Yang
    Jing Qi
    Mahmood Alam
    Deling Zou
    BMC Cardiovascular Disorders, 24
  • [2] Case series of late lead dislodgement of Medtronic SelectSecure 3830 pacing leads in growing paediatric patients
    Ng, Li Yen
    Gallagher, Sarah
    Walsh, Kevin P.
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2021, 5 (02)
  • [3] Transvenous pacing lead-induced thrombosis: A series of cases with a review of the literature
    Barakat, K
    Robinson, NM
    Spurrell, RAJ
    CARDIOLOGY, 2000, 93 (03) : 142 - 148
  • [4] Transvenous pacing in pediatric patients with bipolar lumenless lead: Ten-year clinical experience
    De Filippo, Paolo
    Giofre, Fabrizio
    Leidi, Cristina
    Senni, Michele
    Ferrari, Paola
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2018, 255 : 45 - 49
  • [5] 3830 lead extraction from two patients with left bundle branch area pacing
    Hayashi, Yusuke
    Shimeno, Kenji
    Tamura, Shota
    Naruko, Takahiko
    Fukuda, Daiju
    JOURNAL OF ARRHYTHMIA, 2023, 39 (04) : 623 - 626
  • [6] Tularemia in Pediatric Patients: A Case Series and Review of the Literature
    Kossadoum, Remadji Fiona
    Baron, Audrey
    Parizot, Marie
    Husain, Maya
    Poey, Nora
    Maurin, Max
    Caspar, Yvan
    Caseris, Marion
    Bidet, Philippe
    Bonacorsi, Stephane
    PEDIATRIC INFECTIOUS DISEASE JOURNAL, 2025, 44 (02) : 180 - 185
  • [7] Case of early right ventricular pacing lead perforation and review of the literature
    Nash, Gary
    Williams, John Mark
    Nekkanti, Rajasekhar
    Movahed, Assad
    WORLD JOURNAL OF CLINICAL CASES, 2014, 2 (06) : 206 - 208
  • [8] Case of early right ventricular pacing lead perforation and review of the literature
    Gary Nash
    John Mark Williams
    Rajasekhar Nekkanti
    Assad Movahed
    World Journal of Clinical Cases, 2014, (06) : 206 - 208
  • [9] Neurosurgical treatment of Cushing disease in pediatric patients: case series and review of literature
    Gazioglu, Nurperi
    Canaz, Huseyin
    Camlar, Mahmut
    Tanriover, Necmettin
    Kocer, Naci
    Islak, Civan
    Evliyaoglu, Olcay
    Ercan, Oya
    CHILDS NERVOUS SYSTEM, 2019, 35 (11) : 2127 - 2132
  • [10] Appendicitis in Neutropenic Pediatric Oncology Patients: A Case Series and Review of the Literature
    Pudela, Caileigh
    Lancaster, Scott
    McGahren, Eugene
    Petersen, William C.
    JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2020, 42 (08) : E835 - E838