Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue

被引:45
作者
Mattison, Lars [1 ]
Verma, Atul [2 ]
Tarakji, Khaldoun G. [1 ]
Reichlin, Tobias [3 ]
Hindricks, Gerhard [4 ]
Sack, Kevin L. [1 ]
Onal, Birce [1 ]
Schmidt, Megan M. [1 ]
Miklavcic, Damijan [5 ]
Sigg, Daniel C. [1 ]
机构
[1] Medtronic, Minneapolis, MN USA
[2] McGill Univ, Hlth Ctr, Montreal, PQ, Canada
[3] Univ Bern, Univ Hosp Bern, Dept Cardiol, Inselspital, Bern, Switzerland
[4] Univ Leipzig, Dept Electrophysiol, Heart Ctr Leipzig, Leipzig, Germany
[5] Univ Ljubljana, Fac Elect Engn, Ljubljana, Slovenia
关键词
cardiac ablation; contact force; irreversible electroporation; pulsed field ablation; RADIOFREQUENCY ABLATION; ATRIAL-FIBRILLATION; ELECTROGRAM AMPLITUDE; CATHETER ABLATION; POOR PREDICTORS; LETHAL ISOTHERM; STEAM POP; IMPEDANCE; SIZE;
D O I
10.1111/jce.15813
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionContact force has been used to titrate lesion formation for radiofrequency ablation. Pulsed field ablation (PFA) is a field-based ablation technology for which limited evidence on the impact of contact force on lesion size is available. MethodsPorcine hearts (n = 6) were perfused using a modified Langendorff set-up. A prototype focal PFA catheter attached to a force gauge was held perpendicular to the epicardium and lowered until contact was made. Contact force was recorded during each PFA delivery. Matured lesions were cross-sectioned, stained, and the lesion dimensions measured. ResultsA total of 82 lesions were evaluated with contact forces between 1.3 and 48.6 g. Mean lesion depth was 4.8 +/- 0.9 mm (standard deviation), mean lesion width was 9.1 +/- 1.3 mm, and mean lesion volume was 217.0 +/- 96.6 mm(3). Linear regression curves showed an increase of only 0.01 mm in depth (depth = 0.01 x contact force + 4.41, R-2 = 0.05), 0.03 mm in width (width = 0.03 x contact force + 8.26, R-2 = 0.13) for each additional gram of contact force, and 2.20 mm(3) in volume (volume = 2.20 x contact force + 162, R-2 = 0.10). ConclusionIncreasing contact force using a bipolar, biphasic focal PFA system has minimal effects on acute lesion dimensions in an isolated porcine heart model and achieving tissue contact is more important than the force with which that contact is made.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 35 条
[11]   Reduction in Pulmonary Vein Stenosis and Collateral Damage With Pulsed Field Ablation Compared With Radiofrequency Ablation in a Canine Model [J].
Howard, Brian ;
Haines, David E. ;
Verma, Atul ;
Packer, Douglas ;
Kirchhof, Nicole ;
Barka, Noah ;
Onal, Birce ;
Fraasch, Steve ;
Miklavcic, Damijan ;
Stewart, Mark T. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2020, 13 (09) :E008337
[12]   Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine Heart Electrogram Amplitude, Impedance, and Electrode Temperature Are Poor Predictors of Electrode-Tissue Contact Force and Lesion Size [J].
Ikeda, Atsushi ;
Nakagawa, Hiroshi ;
Lambert, Hendrik ;
Shah, Dipen C. ;
Fonck, Edouard ;
Yulzari, Aude ;
Sharma, Tushar ;
Pitha, Jan V. ;
Lazzara, Ralph ;
Jackman, Warren M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (06) :1174-1180
[13]   Pulsed Field Ablation of Left Ventricular Myocardium in a Swine Infarct Model [J].
Im, Sung Il ;
Higuchi, Satoshi ;
Lee, Adam ;
Stillson, Carol ;
Buck, Eric ;
Morrow, Blake ;
Schenider, Kit ;
Speltz, Molly ;
Gerstenfeld, Edward P. .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2022, 8 (06) :722-731
[14]   Endocardial ventricular pulsed field ablation: a proof-of-concept preclinical evaluation [J].
Koruth, Jacob S. ;
Kuroki, Kenji ;
Iwasawa, Jin ;
Viswanathan, Raju ;
Brose, Richard ;
Buck, Eric D. ;
Donskoy, Elina ;
Dukkipati, Srinivas R. ;
Reddy, Vivek Y. .
EUROPACE, 2020, 22 (03) :434-439
[15]   Membrane Electroporation and Electropermeabilization: Mechanisms and Models [J].
Kotnik, Tadej ;
Rems, Lea ;
Tarek, Mounir ;
Miklavcic, Damijan .
ANNUAL REVIEW OF BIOPHYSICS, VOL 48, 2019, 48 :63-91
[16]   Pulsed-field ablation: Computational modeling of electric fields for lesion depth analysis [J].
Meckes, Daniel ;
Emami, Mehrdad ;
Fong, Ian ;
Lau, Dennis H. ;
Sanders, Prashanthan .
HEART RHYTHM O2, 2022, 3 (04) :433-440
[17]   A validated model of in vivo electric field distribution in tissues for electrochemotherapy and for DNA electrotransfer for gene therapy [J].
Miklavcic, D ;
Semrov, D ;
Mekid, H ;
Mir, LM .
BIOCHIMICA ET BIOPHYSICA ACTA-GENERAL SUBJECTS, 2000, 1523 (01) :73-83
[18]  
Nakagawa H., 2014, HEART RHYTHM, V11, pS548, DOI DOI 10.1016/j.hrthm.2013.10.057
[19]  
Nakagawa H., 2021, HEAR RHYTHM, V18, pS242
[20]   Locations of High Contact Force During Left Atrial Mapping in Atrial Fibrillation Patients Electrogram Amplitude and Impedance Are Poor Predictors of Electrode-Tissue Contact Force for Ablation of Atrial Fibrillation [J].
Nakagawa, Hiroshi ;
Kautzner, Josef ;
Natale, Andrea ;
Peichl, Petr ;
Cihak, Robert ;
Wichterle, Dan ;
Ikeda, Atsushi ;
Santangeli, Pasquale ;
Di Biase, Luigi ;
Jackman, Warren M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (04) :746-753