Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue

被引:45
作者
Mattison, Lars [1 ]
Verma, Atul [2 ]
Tarakji, Khaldoun G. [1 ]
Reichlin, Tobias [3 ]
Hindricks, Gerhard [4 ]
Sack, Kevin L. [1 ]
Onal, Birce [1 ]
Schmidt, Megan M. [1 ]
Miklavcic, Damijan [5 ]
Sigg, Daniel C. [1 ]
机构
[1] Medtronic, Minneapolis, MN USA
[2] McGill Univ, Hlth Ctr, Montreal, PQ, Canada
[3] Univ Bern, Univ Hosp Bern, Dept Cardiol, Inselspital, Bern, Switzerland
[4] Univ Leipzig, Dept Electrophysiol, Heart Ctr Leipzig, Leipzig, Germany
[5] Univ Ljubljana, Fac Elect Engn, Ljubljana, Slovenia
关键词
cardiac ablation; contact force; irreversible electroporation; pulsed field ablation; RADIOFREQUENCY ABLATION; ATRIAL-FIBRILLATION; ELECTROGRAM AMPLITUDE; CATHETER ABLATION; POOR PREDICTORS; LETHAL ISOTHERM; STEAM POP; IMPEDANCE; SIZE;
D O I
10.1111/jce.15813
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionContact force has been used to titrate lesion formation for radiofrequency ablation. Pulsed field ablation (PFA) is a field-based ablation technology for which limited evidence on the impact of contact force on lesion size is available. MethodsPorcine hearts (n = 6) were perfused using a modified Langendorff set-up. A prototype focal PFA catheter attached to a force gauge was held perpendicular to the epicardium and lowered until contact was made. Contact force was recorded during each PFA delivery. Matured lesions were cross-sectioned, stained, and the lesion dimensions measured. ResultsA total of 82 lesions were evaluated with contact forces between 1.3 and 48.6 g. Mean lesion depth was 4.8 +/- 0.9 mm (standard deviation), mean lesion width was 9.1 +/- 1.3 mm, and mean lesion volume was 217.0 +/- 96.6 mm(3). Linear regression curves showed an increase of only 0.01 mm in depth (depth = 0.01 x contact force + 4.41, R-2 = 0.05), 0.03 mm in width (width = 0.03 x contact force + 8.26, R-2 = 0.13) for each additional gram of contact force, and 2.20 mm(3) in volume (volume = 2.20 x contact force + 162, R-2 = 0.10). ConclusionIncreasing contact force using a bipolar, biphasic focal PFA system has minimal effects on acute lesion dimensions in an isolated porcine heart model and achieving tissue contact is more important than the force with which that contact is made.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 35 条
[1]   The effects of electrode-tissue contact on radiofrequency lesion generation [J].
Avitall, B ;
Mughal, K ;
Hare, J ;
Helms, R ;
Krum, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (12) :2899-2910
[2]  
Calkins H, 2017, HEART RHYTHM, V14, pE445, DOI [10.1016/j.hrthm.2017.05.012, 10.1016/j.hrthm.2017.07.009, 10.1093/europace/eux275]
[3]   Relationship Between Catheter Forces, Lesion Characteristics, "Popping," and Char Formation: Experience with Robotic Navigation System [J].
Di Biase, Luigi ;
Natale, Andrea ;
Barrett, Conor ;
Tan, Carmela ;
Elayi, Claude S. ;
Ching, Chi Keong ;
Wang, Paul ;
Al-Ahmad, Amin ;
Arruda, Mauricio ;
Burkhardt, J. David ;
Wisnoskey, Brian J. ;
Chowdhury, Punam ;
De Marco, Shari ;
Armaganijan, Luciana ;
Litwak, Kenneth N. ;
Schweikert, Robert A. ;
Cummings, Jennifer E. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (04) :436-440
[4]   EARLY PHASE ACUTE MYOCARDIAL INFARCT SIZE QUANTIFICATION - VALIDATION OF THE TRIPHENYL TETRAZOLIUM CHLORIDE TISSUE ENZYME STAINING TECHNIQUE [J].
FISHBEIN, MC ;
MEERBAUM, S ;
RIT, J ;
LANDO, U ;
KANMATSUSE, K ;
MERCIER, JC ;
CORDAY, E ;
GANZ, W .
AMERICAN HEART JOURNAL, 1981, 101 (05) :593-600
[5]   A Numerical Investigation of the Electric and Thermal Cell Kill Distributions in Electroporation-Based Therapies in Tissue [J].
Garcia, Paulo A. ;
Davalos, Rafael V. ;
Miklavcic, Damijan .
PLOS ONE, 2014, 9 (08)
[6]  
HAINES D E, 1991, Journal of Cardiovascular Electrophysiology, V2, P509, DOI 10.1111/j.1540-8167.1991.tb01353.x
[7]   Letter by Haines Regarding Article, "Direct Measurement of the Lethal Isotherm for Radiofrequency Ablation of Myocardial Tissue" [J].
Haines, David E. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2011, 4 (05) :E67-E67
[8]   Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: A prospective case control study [J].
Haldar, Shouvik ;
Jarman, Julian W. E. ;
Panikker, Sandeep ;
Jones, David G. ;
Salukhe, Tushar ;
Gupta, Dhiraj ;
Wynn, Gareth ;
Hussain, Wajid ;
Markides, Vias ;
Wong, Tom .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (02) :1160-1166
[9]   Effects of Electrode-Tissue Proximity on Cardiac Lesion Formation Using Pulsed Field Ablation [J].
Howard, Brian ;
Verma, Atul ;
Tzou, Wendy S. ;
Mattison, Lars ;
Kos, Bor ;
Miklavcic, Damijan ;
Onal, Birce ;
Stewart, Mark T. ;
Sigg, Daniel C. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2022, 15 (10) :706-713
[10]   Characterization of Phrenic Nerve Response to Pulsed Field Ablation [J].
Howard, Brian ;
Haines, David E. ;
Verma, Atul ;
Kirchhof, Nicole ;
Barka, Noah ;
Onal, Birce ;
Stewart, Mark T. ;
Sigg, Daniel C. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2022, 15 (06) :E010127