Preliminary study on dosimetry characteristics of a novel cylindrical dose verification system

被引:1
作者
He, Long [1 ]
Zhu, Jinhan [2 ]
Wang, Xuetao [3 ]
Zhang, Bailin [3 ]
Hu, Qiang [4 ]
Chen, Lixin [2 ,5 ]
Liu, Xiaowei [1 ,6 ]
机构
[1] Sun Yat Sen Univ, Sch Phys, Guangzhou, Peoples R China
[2] Sun Yat sen Univ, Canc Ctr, Guangzhou, Peoples R China
[3] Guangzhou Univ Chinese Med, Radiat Oncol Dept, Affiliated Hosp 2, Guangzhou, Peoples R China
[4] Guangzhou Raydose Med Technol Co Ltd, Guangzhou, Peoples R China
[5] Sun Yat Sen Univ, Canc Ctr, 651 Dong Feng Rd East, Guangzhou, Peoples R China
[6] Sun Yat Sen Univ, Sch Phys, 135 Xin Gang Rd West, Guangzhou, Chin, Myanmar
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2023年 / 24卷 / 10期
基金
中国国家自然科学基金;
关键词
detector array; dose verification; dosimetry; ionization chamber; DIODE-ARRAY; IMRT; VMAT; CALIBRATION; TOMOTHERAPY; DEPENDENCE; MATRIXX; DEVICE;
D O I
10.1002/acm2.14138
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo develop a novel ionization chamber array dosimetry system, study its dosimetry characteristics, and perform preliminary tests for plan dose verification.MethodsThe dosimetry characteristics of this new array were tested, including short-term and long-term reproducibility, dose linearity, dose rate dependence, field size dependence, and angular dependence. The open field and MLC field plans were designed for dose testing. Randomly select 30 patient treatment plans (10 intensity-modulated radiation therapy [IMRT] plans and 20 volumetric modulated arc therapy [VMAT] plans) that have undergone dose verification using Portal Dosimetry to perform verification measurement and evaluate dose verification test results.ResultsThe dosimetry characteristics of the arrays all performed well. The gamma passing rates (3%/2 mm) were more than 96% for the combined open field and MLC field plans. The average gamma pass rates were (99.54 & PLUSMN; 0.58)% and (96.70 & PLUSMN; 3.41)% for the 10 IMRT plans and (99.32 & PLUSMN; 0.89)% and (94.91 & PLUSMN; 6.01)% for the 20 VMAT plans at the 3%/2 mm and 2%/2 mm criteria, respectively, which is similar to the Portal Dosimetry's measurement results.ConclusionsThis novel ionization chamber array demonstrates good dosimetry characteristics and is suitable for clinical IMRT and VMAT plan verifications.
引用
收藏
页数:12
相关论文
共 22 条
  • [1] A novel approach to EPID-based 3D volumetric dosimetry for IMRT and VMAT QA
    Alhazmi, Abdulaziz
    Gianoli, Chiara
    Neppl, Sebastian
    Martins, Juliana
    Veloza, Stella
    Podesta, Mark
    Verhaegen, Frank
    Reiner, Michael
    Belka, Claus
    Parodi, Katia
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (11)
  • [2] ROC analysis in patient specific quality assurance
    Carlone, Marco
    Cruje, Charmainne
    Rangel, Alejandra
    McCabe, Ryan
    Nielsen, Michelle
    MacPherson, Miller
    [J]. MEDICAL PHYSICS, 2013, 40 (04)
  • [3] Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA
    Chaswal, Vibha
    Weldon, Michael
    Gupta, Nilendu
    Chakravarti, Arnab
    Rong, Yi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04): : 212 - 225
  • [4] Sensitivity evaluation of two commercial dosimeters in detecting Helical TomoTherapy treatment delivery errors
    Deshpande, Shrikant
    Geurts, Mark
    Vial, Philip
    Metcalfe, Peter
    Holloway, Lois
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 37 : 68 - 74
  • [5] Treatment plan verification: A review on the comparison of dose distributions
    Diamantopoulos, Stefanos
    Platoni, Kalliopi
    Patatoukas, Georgios
    Karaiskos, Pantelis
    Kouloulias, Vassilis
    Efstathopoulos, Efstathios
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 67 : 107 - 115
  • [6] Evaluation of a new VMAT QA device, or the "X" and "O" array geometries
    Feygelman, Vladimir
    Zhang, Geoffrey
    Stevens, Craig
    Nelms, Benjamin E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 146 - 168
  • [7] Cross verification of independent dose recalculation, log files based, and phantom measurement-based pretreatment quality assurance for volumetric modulated arc therapy
    Han, Ce
    Yi, Jinling
    Zhu, Kecheng
    Zhou, Yongqiang
    Ai, Yao
    Zheng, Xiaomin
    Xie, Congying
    Jin, Xiance
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 98 - 104
  • [8] Response of the ArcCHECK® device at 6 MV and 15 MV for VMAT and IMRT quality control
    Jimenez-Melguizo, Miguel
    Espinosa, Miguel
    Montes, Joaquin
    Guirado, Damian
    Lallena, Antonio M.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 80 : 373 - 382
  • [9] A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery
    Jursinic, PA
    Nelms, BE
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 870 - 879
  • [10] MapCHECK used for rotational IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc
    Jursinic, Paul A.
    Sharma, Renu
    Reuter, Jim
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2837 - 2846