Online geometry calibration for retrofit computed tomography from a mouse rotation system and a small-animal imager

被引:1
作者
Zhou, Huanyi [1 ]
Reeves, Stanley [1 ]
Chou, Cheng-Ying [2 ]
Brannen, Andrew [3 ]
Panizzi, Peter [3 ]
机构
[1] Auburn Univ, Elect & Comp Engn Dept, Auburn, AL 36849 USA
[2] Natl Taiwan Univ, Dept Biomechatron Engn, 1,Sec 4,Roosevelt Rd, Taipei 10617, Taiwan
[3] Auburn Univ, Harrison Coll Pharm, Drug Discovery & Dev, Auburn, AL USA
基金
美国国家卫生研究院;
关键词
CT preprocessing; online geometry calibration; preclinical CBCT; MICRO-CT; BEAM;
D O I
10.1002/mp.15953
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Computed tomography (CT) generates a three-dimensional rendering that can be used to interrogate a given region or desired structure from any orientation. However, in preclinical research, its deployment remains limited due to relatively high upfront costs. Existing integrated imaging systems that provide merged planar X-ray also dwarfs CT popularity in small laboratories due to their added versatility. Purpose In this paper, we sought to generate CT-like data using an existing small-animal X-ray imager with a specialized specimen rotation system, or MiSpinner. This setup conforms to the cone-beam CT (CBCT) geometry, which demands high spatial calibration accuracy. Therefore, a simple but robust geometry calibration algorithm is necessary to ensure that the entire imaging system works properly and accurately. Methods Because the rotation system is not permanently affixed, we propose a structure tensor-based two-step online (ST-TSO) geometry calibration algorithm. Specifically, two datasets are needed, namely, calibration and actual measurements. A calibration measurement detects the background of the system forward X-ray projections. A study on the background image reveals the characteristics of the X-ray photon distribution, and thus, provides a reliable estimate of the imaging geometry origin. Actual measurements consisted of an X-ray of the intended object, including possible geometry errors. A comprehensive image processing technique helps to detect spatial misalignment information. Accordingly, the first processing step employs a modified projection matrix-based calibration algorithm to estimate the relevant geometric parameters. Predicted parameters are then fine-tuned in a second processing step by an iterative strategy based on the symmetry property of the sum of projections. Virtual projections calculated from the parameters after two-step processing compensate for the scanning errors and are used for CT reconstruction. Experiments on phantom and mouse imaging data were performed to validate the calibration algorithm. Results Once system correction was conducted, CBCT of a CT bar phantom and a cohort of euthanized mice were analyzed. No obvious structure error or spatial artifacts were observed, validating the accuracy of the proposed geometry calibration method. Digital phantom simulation indicated that compared with the preset spatial values, errors in the final estimated parameters could be reduced to 0.05 degrees difference in dominant angle and 0.5-pixel difference in dominant axis bias. The in-plane resolution view of the CT-bar phantom revealed that the resolution approaches 150 mu$\umu$m. Conclusions A constrained two-step online geometry calibration algorithm has been developed to calibrate an integrated X-ray imaging system, defined by a first-step analytical estimation and a second-step iterative fine-tuning. Test results have validated its accuracy in system correction, thus demonstrating the potential of the described system to be modified and adapted for preclinical research.
引用
收藏
页码:192 / 208
页数:17
相关论文
共 35 条
  • [31] Abnormal pixel detection using sum-of-projections symmetry in cone beam computed tomography
    Yang, Xiaoquan
    Meng, Yuanzheng
    Gong, Hui
    Deng, Yong
    Luo, Qingming
    [J]. OPTICS EXPRESS, 2012, 20 (10): : 11014 - 11030
  • [32] Data consistency based rigid motion artifact reduction in fan-beam CT
    Yu, Hengyong
    Wang, Ge
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2007, 26 (02) : 249 - 260
  • [33] Cone-beam imaging with tilted rotation axis: Method and performance evaluation
    Zhao, Chumin
    Herbst, Magdalena
    Vogt, Sebastian
    Ritschl, Ludwig
    Kappler, Steffen
    Siewerdsen, Jeffrey H.
    Zbijewski, Wojciech
    [J]. MEDICAL PHYSICS, 2020, 47 (08) : 3305 - 3320
  • [34] Estimating the Center of Rotation of Tomographic Imaging Systems with a Limited Number of Projections
    Zhou, Huanyi
    Reeves, Stanley J.
    Panizzi, Peter R.
    [J]. 2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 3157 - 3160
  • [35] Automated Motion Correction for In Vivo Optical Projection Tomography
    Zhu, Shouping
    Dong, Di
    Birk, Udo Jochen
    Rieckher, Matthias
    Tavernarakis, Nektarios
    Qu, Xiaochao
    Liang, Jimin
    Tian, Jie
    Ripoll, Jorge
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (07) : 1358 - 1371