Optimization of limit of detection in Taylor dispersion analysis: Application to the size determination of vaccine antigens

被引:1
作者
Malburet, Camille [1 ,2 ]
Martin, Michel [3 ]
Leclercq, Laurent [1 ]
Cotte, Jean-Francois [2 ]
Thiebaud, Jerome [2 ]
Biron, Jean-Philippe [1 ]
Chamieh, Joseph [1 ]
Cottet, Herve [1 ]
机构
[1] Univ Montpellier, IBMM, CNRS, ENSCM, Montpellier, France
[2] Sanofi, Analyt Sci, 1541 Ave Marcel, F-69280 Marcy Letoile, France
[3] Univ Paris Cite, Sorbonne Univ, PMMH, CNRS, F-75005 Paris, France
来源
TALANTA OPEN | 2023年 / 7卷
关键词
Taylor dispersion analysis; Proteins; Vaccines; Limit of detection; Limit of quantification; Quantitative analysis; PROTEINS; DISEASE;
D O I
10.1016/j.talo.2023.100209
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a new vaccine requires the precise characterization of all the physicochemical parameters of the vaccine antigens, which are the molecules that induce the immune response. Taylor dispersion analysis (TDA) is a promising alternative technique for the determination of diffusion coefficients and hydrodynamic radii of proteins, macromolecules and nanoparticles. In this work, TDA was used to determine the hydrodynamic radius distribution and its average value of four antigens: diphtheria toxoid (DT), tetanus toxoid (TT), hepatitis B surface antigen (HBsAg) and polyribosyl-ribitol phosphate conjugated to tetanus toxoid (PRP-T). The robustness of the results obtained was investigated on bare fused silica capillary and hydroxypropylcellulose coated capillary. The impact of operational parameters on the limit of detection (LOD) and limit of quantification (LOQ) were studied from both theoretical and experimental points of view. The influence of the diameter and the length of the capillary on the LOD and LOQ were studied as well as the impact of the mobilization pressure. General guidelines for the choice of the initial operating conditions are given for the development of future TDA methods.
引用
收藏
页数:9
相关论文
共 30 条
  • [1] Akinin V. V., 2017, DOKL EARTH SCI, V476, P1173, DOI [10.1021/jf402198r, DOI 10.18334/CE.16.2.114225, 10.1134/s1028334x17100129, DOI 10.1134/S1028334X17100129]
  • [2] Vaccination greatly reduces disease, disability, death and inequity worldwide
    Andre, F. E.
    Booy, R.
    Bock, H. L.
    Clemens, J.
    Datta, S. K.
    John, T. J.
    Lee, B. W.
    Lolekha, S.
    Peltola, H.
    Ruff, T. A.
    Santosham, M.
    Schmitt, H. J.
    [J]. BULLETIN OF THE WORLD HEALTH ORGANIZATION, 2008, 86 (02) : 140 - 146
  • [3] ON THE DISPERSION OF A SOLUTE IN A FLUID FLOWING THROUGH A TUBE
    ARIS, R
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1956, 235 (1200): : 67 - 77
  • [4] Size-characterization of natural and synthetic polyisoprenes by Taylor dispersion analysis
    Biron, Jean-Philippe
    Bonfils, Frederic
    Cipelletti, Luca
    Cottet, Herve
    [J]. POLYMER TESTING, 2018, 66 : 244 - 250
  • [5] Boqué R, 2009, LC GC EUR, V22, P82
  • [6] Size characterization of lipid-based self-emulsifying pharmaceutical excipients during lipolysis using Taylor dispersion analysis with fluorescence detection
    Chamieh, Joseph
    Merdassi, Habib
    Rossi, Jean-Christophe
    Jannin, Vincent
    Demarne, Frederic
    Cottet, Herve
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2018, 537 (1-2) : 94 - 101
  • [7] Limits in Size of Taylor Dispersion Analysis: Representation of the Different Hydrodynamic Regimes and Application to the Size-Characterization of Cubosomes
    Chamieh, Joseph
    Leclercq, Laurent
    Martin, Michel
    Slaoui, Sofia
    Jensen, Henrik
    Ostergaard, Jesper
    Cottet, Herve
    [J]. ANALYTICAL CHEMISTRY, 2017, 89 (24) : 13487 - 13493
  • [8] Hydrodynamic size characterization of a self-emulsifying lipid pharmaceutical excipient by Taylor dispersion analysis with fluorescent detection
    Chamieh, Joseph
    Jannin, Vincent
    Demarne, Frederic
    Cottet, Herve
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2016, 513 (1-2) : 262 - 269
  • [9] Comparison of single and double detection points Taylor Dispersion Analysis for monodisperse and polydisperse samples
    Chamieh, Joseph
    Cottet, Herve
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2012, 1241 : 123 - 127
  • [10] Measuring Arbitrary Diffusion Coefficient Distributions of Nano-Objects by Taylor Dispersion Analysis
    Cipelletti, Luca
    Biron, Jean-Philippe
    Martin, Michel
    Cottet, Herve
    [J]. ANALYTICAL CHEMISTRY, 2015, 87 (16) : 8489 - 8496