Feasibility of delivered dose reconstruction for MR-guided SBRT of pancreatic tumors with fast, real-time 3D cine MRI

被引:7
|
作者
Grimbergen, Guus [1 ]
Pötgens, Giulia G. [2 ]
Eijkelenkamp, Hidde [1 ]
Raaymakers, Bas W. [1 ]
Intven, Martijn P. W. [1 ]
Meijer, Gert J. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiat Oncol, Utrecht, Netherlands
[2] Eindhoven Univ Technol, Dept Biomed Engn, Eindhoven, Netherlands
关键词
Pancreatic cancer; MR -guided SBRT; Intrafraction motion; Dose reconstruction; 3D cine MRI; DEFORMABLE IMAGE REGISTRATION; RESPIRATORY MOTION; RADIOTHERAPY; UNCERTAINTIES; ACCUMULATION; SYSTEM; FIELDS; HEAD; CONSISTENCY; VALIDATION;
D O I
10.1016/j.radonc.2023.109506
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: In MR-guided SBRT of pancreatic cancer, intrafraction motion is typically monitored with (interleaved) 2D cine MRI. However, tumor surroundings are often not fully captured in these images, and motion might be distorted by through-plane movement. In this study, the feasibility of highly accelerated 3D cine MRI to reconstruct the delivered dose during MR-guided SBRT was assessed. Materials and methods: A 3D cine MRI sequence was developed for fast, time-resolved 4D imaging, featuring a low spatial resolution that allows for rapid volumetric imaging at 430 ms. The 3D cines were acquired during the entire beam-on time of 23 fractions of online adaptive MR-guided SBRT for pancreatic tumors on a 1.5 T MR-Linac. A 3D deformation vector field (DVF) was extracted for every cine dynamic using deformable image registration. Next, these DVFs were used to warp the partial dose delivered in the time interval between consecutive cine acquisitions. The warped dose plans were summed to obtain a total delivered dose. The delivered dose was also calculated under various motion correction strategies. Key DVH parameters of the GTV, duodenum, small bowel and stomach were extracted from the delivered dose and compared to the planned dose. The uncertainty of the calculated DVFs was determined with the inverse consistency error (ICE) in the high-dose regions.Results: The mean (SD) relative (ratio delivered/planned) D99% of the GTV was 0.94 (0.06), and the mean (SD) relative D0.5cc of the duodenum, small bowel, and stomach were respectively 0.98 (0.04), 1.00 (0.07), and 0.98 (0.06). In the fractions with the lowest delivered tumor coverage, it was found that significant lateral drifts had occurred. The DVFs used for dose warping had a low uncertainty with a mean (SD) ICE of 0.65 (0.07) mm.Conclusion: We employed a fast, real-time 3D cine MRI sequence for dose reconstruction in the upper abdomen, and demonstrated that accurate DVFs, acquired directly from these images, can be used for dose warping. The reconstructed delivered dose showed only a modest degradation of tumor coverage, mostly attainable to baseline drifts. This emphasizes the need for motion monitoring and development of intrafraction treatment adaptation solutions, such as baseline drift corrections.(c) 2023 The Author(s). Published by Elsevier B.V. Radiotherapy and Oncology 182 (2023) 1-7 This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页数:7
相关论文
共 40 条
  • [1] Feasibility of real-time motion tracking using cine MRI during MR-guided radiation therapy for abdominal targets
    Keiper, Timothy D.
    Tai, An
    Chen, Xinfeng
    Paulson, Eric
    Lathuiliere, Fabienne
    Beriault, Silvain
    Hebert, Francois
    Cooper, David T.
    Lachaine, Martin
    Li, X. Allen
    MEDICAL PHYSICS, 2020, 47 (08) : 3554 - 3566
  • [2] Real-time motion monitoring using orthogonal cine MRI during MR-guided adaptive radiation therapy for abdominal tumors on 1.5T MR-Linac
    Jassar, Hassan
    Tai, An
    Chen, Xinfeng
    Keiper, Timothy D. D.
    Paulson, Eric
    Lathuiliere, Fabienne
    Beriault, Silvain
    Hebert, Francois
    Savard, Laurence
    Cooper, David T. T.
    Cloake, Sneha
    Li, X. Allen
    MEDICAL PHYSICS, 2023, 50 (05) : 3103 - 3116
  • [3] Gaussian Processes for real-time 3D motion and uncertainty estimation during MR-guided radiotherapy
    Huttinga, Niek R. F.
    Bruijnen, Tom
    van den Berg, Cornelis A. T.
    Sbrizzi, Alessandro
    MEDICAL IMAGE ANALYSIS, 2023, 88
  • [4] Feasibility study on 3D image reconstruction from 2D orthogonal cine-MRI for MRI-guided radiotherapy
    Paganelli, Chiara
    Lee, Danny
    Kipritidis, John
    Whelan, Brendan
    Greer, Peter B.
    Baroni, Guido
    Riboldi, Marco
    Keall, Paul
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2018, 62 (03) : 389 - 400
  • [5] Soft-tissue prostate intrafraction motion tracking in 3D cine-MR for MR-guided radiotherapy
    Keizer, D. M. de Muinck
    Kerkmeijer, L. G. W.
    Maspero, M.
    Andreychenko, A.
    van Zyp, J. R. N. van der Voort
    van den Berg, C. A. T.
    Raaymakers, B. W.
    Lagendijk, J. J. W.
    de Boer, J. C. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (23)
  • [6] The impact of 2D cine MR imaging parameters on automated tumor and organ localization for MR-guided real-time adaptive radiotherapy
    Menten, Martin J.
    Fast, Martin F.
    Wetscherek, Andreas
    Rank, Christopher M.
    Kachelriess, Marc
    Collins, David J.
    Nill, Simeon
    Oelfke, Uwe
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23)
  • [7] Real-Time Non-Rigid 3D Respiratory Motion Estimation for MR-Guided Radiotherapy Using MR-MOTUS
    Huttinga, Niek R. F.
    Bruijnen, Tom
    Van den Berg, Cornelis A. T.
    Sbrizzi, Alessandro
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2022, 41 (02) : 332 - 346
  • [8] Accelerating volumetric cine MRI (VC-MRI) using undersampling for real-time 3D target localization/tracking in radiation therapy: a feasibility study
    Harris, Wendy
    Yin, Fang-Fang
    Wang, Chunhao
    Zhang, You
    Cai, Jing
    Ren, Lei
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (01)
  • [9] Automated Real-time Needle-Guide Tracking for Fast 3-T MR-guided Transrectal Prostate Biopsy: A Feasibility Study
    Zamecnik, Patrik
    Schouten, Martijn G.
    Krafft, Axel J.
    Maier, Florian
    Schlemmer, Heinz-Peter
    Barentsz, Jelle O.
    Bock, Michael
    Fuetterer, Jurgen J.
    RADIOLOGY, 2014, 273 (03) : 879 - 886
  • [10] Real-time 4D dose reconstruction for tracked dynamic MLC deliveries for lung SBRT
    Kamerling, Cornelis Ph.
    Fast, Martin F.
    Ziegenhein, Peter
    Menten, Martin J.
    Nill, Simeon
    Oelfke, Uwe
    MEDICAL PHYSICS, 2016, 43 (11) : 6072 - 6081