Modified design of x-ray protective clothing to enhance radiation protection for interventional radiologists

被引:1
作者
Choi, Tae Won [1 ]
Chung, Jin Wook [2 ,3 ,4 ,5 ]
Kwon, Yohan [1 ]
机构
[1] Ajou Univ, Dept Radiol, Sch Med, Suwon, South Korea
[2] Seoul Natl Univ Hosp, Dept Radiol, Seoul, South Korea
[3] Seoul Natl Univ, Dept Radiol, Coll Med, Seoul, South Korea
[4] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul, South Korea
[5] Seoul Natl Univ Hosp, Dept Radiol, 101 Daehak ro,Jongno gu, Seoul 03080, South Korea
关键词
effective dose; interventional radiology; radiation protection; EXPOSURE; OPERATOR; ORGAN;
D O I
10.1002/mp.16309
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundIn interventional radiology procedures, the operator typically stands on the right side of the patient's right thigh to manipulate devices through the femoral sheath. Because the standard x-ray protective clothing is designed as sleeveless and scatter radiations from the patient are mainly incident from the left-anterior direction to the operator, the arm hole of the clothing may be a significant unprotected area, contributing to an increase in the operator's organ doses and effective dose. PurposeThis study aimed to compare the organ doses and effective dose received by the interventional radiologist when wearing the standard x-ray protective clothing and when wearing the modified clothing with an additional shoulder guard. MethodsThe experimental setup aimed to simulate actual clinical practice in interventional radiology. The patient phantom was located at the beam center to generate scatter radiation. An adult female anthropomorphic phantom loaded with 126 nanoDots (Landauer Inc., Glenwood, IL) was used to measure organ and effective doses to the operator. The standard wrap-around type x-ray protective clothing offered 0.25-mm lead-equivalent protection, and the frontal overlap area offered 0.50-mm lead-equivalent protection. The shoulder guard was custom-made with a material providing x-ray protection equivalent to lead of 0.50 mm thickness. The organ and effective doses were compared between the operator wearing the standard protective clothing and the one wearing the modified clothing with a shoulder guard. ResultsAfter adding the shoulder guard, doses to the lungs, bone marrow, and esophagus decreased by 81.9%, 58.6%, and 58.7%, respectively, and the effective dose to the operator decreased by 47.7%. ConclusionsWidespread use of modified x-ray protective clothing with shoulder guards can significantly decrease the overall occupational radiation risk in interventional radiology.
引用
收藏
页码:3825 / 3832
页数:8
相关论文
共 31 条
[1]   Characteristics of an OSLD in the diagnostic energy range [J].
Al-Senan, Rani M. ;
Hatab, Mustapha R. .
MEDICAL PHYSICS, 2011, 38 (07) :4396-4405
[2]  
[Anonymous], 1995, Ann ICRP, V25, P1
[3]  
[Anonymous], 2007, Ann ICRP, V37, P1, DOI DOI 10.1016/J.ICRP.2007.10.003
[4]   Get Protected! Recommendations for Staff in IR [J].
Bartal, Gabriel ;
Vano, Eliseo ;
Paulo, Graciano .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2021, 44 (06) :871-876
[5]   Contemporary Interventional Radiology Dosimetry: Analysis of 4,784 Discrete Procedures at a Single Institution [J].
Bundy, Jacob J. ;
Chick, Jeffrey Forris Beecham ;
Hage, Anthony N. ;
Gemmete, Joseph J. ;
Srinivasa, Rajiv N. ;
Johnson, Evan J. ;
Christodoulou, Emmanuel ;
Srinivasa, Ravi N. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2018, 15 (09) :1214-1221
[6]  
Choi TW., 2020, J RADIOL PROT, V40
[7]   Use of effective dose to assess x-ray protective clothing [J].
Eder, Heinrich ;
Schlattl, Helmut .
JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) :R140-R151
[8]   Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures: a phantom study [J].
Fetterly, Kenneth A. ;
Schueler, Beth A. ;
Grams, Michael P. ;
Sturchio, Glenn M. .
JOURNAL OF RADIOLOGICAL PROTECTION, 2017, 37 (01) :43-58
[9]   Comparative evaluation of organ and effective doses for paediatric patients with those for adults in chest and abdominal CT examinations [J].
Fujii, K. ;
Aoyama, T. ;
Koyama, S. ;
Kawaura, C. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (956) :657-667
[10]  
Giretzlehner M., 2021, MED KAUNAS, V57