Relative efficiency of radiophotoluminescent glass detectors in low energy ion beams

被引:6
作者
Majer, Marija [1 ,3 ]
Pasaricek, Luka [1 ]
Knezovic, Zeljka [1 ]
Bokulic, Tomislav [2 ]
Provatas, Georgios [1 ]
Mihalic, Iva Bozicevic [1 ]
机构
[1] Rudjer Boskovic Inst, Zagreb 10000, Croatia
[2] Univ Zagreb, Fac Sci, Zagreb 10000, Croatia
[3] Rudjer Boskovic Inst, Bijenicka C 54, Zagreb 10000, Croatia
关键词
Radiophotoluminescent detectors; Ion beams; Relative efficiency; Dose response; HEAVY CHARGED-PARTICLES; TRACK STRUCTURE-THEORY; THERMO-LUMINESCENCE; ALPHA-PARTICLES; TL EFFICIENCY; MONTE-CARLO; DOSIMETRY; DISTRIBUTIONS; RADIATION; SYSTEM;
D O I
10.1016/j.radmeas.2023.106973
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In heavy charged particle dosimetry, the luminescence efficiency strongly depends on linear energy transfer (LET) and characterization of the detectors' efficiency as a function of the particle LET is very important. Available literature data for the efficiency of radiophotoluminescent (RPL) glass detectors are scarce and especially for low energy range no data was found. In this study, detectors made of the most widely known RPL material (silver activated phosphate glass) were exposed to various low energy ion beams (1H, 7Li, and 12C). For every ion beam, the linear dose (fluence) response of the RPL glass detectors was confirmed for the investigated dose (fluence) range and the RPL efficiency relative to 60Co gamma rays was evaluated. The relative efficiency decreased as LET increased but not as a unique function of LET, where the relative efficiency (expressed for doses in RPL glass) decreased from 0.85 & PLUSMN; 0.06 for the 5 MeV 1H ion beam (LETw = 8.1 keV/& mu;m) to 0.017 & PLUSMN; 0.001 for the 15 MeV 12C ion beam (LETw = 673.8 keV/& mu;m). More experimental data are still needed but results obtained in this study may motivate the application and testing of microdosimetric models for RPL glass detectors.
引用
收藏
页数:8
相关论文
共 50 条
[1]  
[Anonymous], 2008, 1002008 JCGM BUR INT
[2]  
ATGC, 2007, Explanation Material of RPL Glass Dosemeter: Small Element System
[3]  
Benbow S., 2020, GRAPH GRABBER 202
[4]   Space radiation dosimetry in low-Earth orbit and beyond [J].
Benton, ER ;
Benton, E .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 184 (1-2) :255-294
[5]   Dosimetry of densely ionising radiation with three LiF phosphors for space applications [J].
Bilski, P. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :397-400
[6]   THERMOLUMINESCENCE EFFICIENCY OF LIFMG,CU,P (MCP-N) DETECTORS TO PHOTONS, BETA-ELECTRONS, ALPHA-PARTICLES AND THERMAL-NEUTRONS [J].
BILSKI, P ;
OLKO, P ;
BURGKHARDT, B ;
PIESCH, E ;
WALIGORSKI, MPR .
RADIATION PROTECTION DOSIMETRY, 1994, 55 (01) :31-38
[7]   Self-attenuation of TL light from LiF - a comparison of different experimental methods [J].
Bilski, P ;
Budzanowski, M .
RADIATION MEASUREMENTS, 2001, 33 (05) :679-685
[8]  
Birks J.B., 1967, The Theory and Practice of Scintillation Counting: International Series of Monographs on Electronics and Instrumentation
[9]   SCINTILLATIONS FROM ORGANIC CRYSTALS - SPECIFIC FLUORESCENCE AND RELATIVE RESPONSE TO DIFFERENT RADIATIONS [J].
BIRKS, JB .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION A, 1951, 64 (382) :874-877
[10]   Correction of stopping power and LET quenching for radiophotoluminescent glass dosimetry in a therapeutic proton beam [J].
Chang, Weishan ;
Koba, Yusuke ;
Katayose, Tetsurou ;
Yasui, Keisuke ;
Omachi, Chihiro ;
Hariu, Masatsugu ;
Saitoh, Hidetoshi .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (23) :8869-8881