Simulation study of a planar quadrupole mass filter for MEMS mass spectrometer

被引:1
作者
Cheng, Yu -Peng [1 ,2 ]
Shen, Chen [1 ,2 ]
Wu, Zhang-Xu [1 ,2 ]
Li, Shan [2 ]
Liu, You-Jiang [2 ]
Wang, Han [1 ,2 ]
Chen, Chi-Lai [2 ]
机构
[1] Univ Sci & Technol China, Hefei 230026, Peoples R China
[2] Chinese Acad Sci, Hefei Inst Phys Sci, Hefei 230031, Peoples R China
基金
国家重点研发计划;
关键词
Stacked-layer structure; Mass spectrometer chip; Ion trajectory simulation; High-pressure tolerance; Stability diagram; GAS-ANALYSIS; PERFORMANCE; FIELD;
D O I
10.1016/j.cjac.2024.100364
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Micro-electro-mechanical system (MEMS) based mass spectrometry, MEMS mass spectrometry, emerges a new technique branch of miniature mass spectrometry. One of the main challenges hindering the development of MEMS mass spectrometry is the poor compatibility between the complex electrode structure of traditional ion optics and MEMS processes. This paper proposed a planar quadrupole mass filter utilizing a highly simplified stacked-layer structure and film electrodes, which is easy to fabricate with high precision by MEMS surface micromachining. The ion trajectory simulation model of the MEMS mass spectrometry chip involving all necessary ion optics was elaborated for investigating the preliminary performance of the planar quadrupole mass filter. By optimizing the crucial parameters such as the frequency of radio frequency voltage, the ratio of direct current voltage and radio frequency voltage, pre -pole length, main -pole length, and injection kinetic energy, the preliminary simulation result of planar quadrupole mass filter achieved a 1-2 Da peak width (at 50% of peak height) and 20-80% ion transmission efficiency. Importantly, the planar quadrupole mass filter demonstrated a maximum working pressure of 0.1 Pa (using air as the buffer gas), 100 times higher than that of conventional quadrupole mass filters, while maintaining a 0.7-1.8 Da peak width and similar to 5% ion transmission efficiency. Additionally, to modify the theoretical formulas for the non -ideal quadrupolar field of planar quadrupole mass filter, a novel approach based on stability diagrams obtained by simulation was also introduced. The pretty good linearity ( R2 = 1) between ion masses in the range of 10 to 100 Da and the corresponding radio frequency voltages was verified through the simulated mass spectra. Notably, the proportional coefficient of ion mass and radio frequency voltage exhibited a minimal 4% deviation between the simulated value (4.7132 V0P/Da) and the calculated value (4.9036 V0P/Da) through the modified formula. In conclusion, the preliminary simulation results demonstrated that the planar quadrupole mass filter possessed moderate performance, high -pressure tolerance, and excellent theoretical consistency, which proved it a promising technology of MEMS mass spectrometry. It can find applications in the fields of gas detection, including the rapid response to hazardous gas emissions and in situ detection of dissolved gases in the deep sea.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Development of Quadrupole Mass Spectrometers Using Rapid Prototyping Technology
    Brkic, Boris
    France, Neil
    Clare, Adam T.
    Sutcliffe, Chris J.
    Chalker, Paul R.
    Taylor, Stephen
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2009, 20 (07) : 1359 - 1365
  • [32] Method of analyzing 4He in a deuterium atmosphere using a high-resolution quadrupole mass spectrometer
    Isobe, Y
    Yabuta, K
    Ueda, S
    Takahashi, A
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2002, 41 (2A): : 866 - 870
  • [33] Rapid Mass Analysis of Reserpine and Arginine by Ion Trap Mass Spectrometer
    Xu Fu-Xing
    Dang Qian-Kun
    Chen Yin-Juan
    Yang Kai
    Wang Qiang
    Chen Bin
    Wang Yuan-Yuan
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (06) : 949 - 953
  • [34] Macromolecular ion accelerator mass spectrometer
    Hsu, Yun-Fei
    Lin, Jung-Lee
    Chu, Ming-Lee
    Wang, Yi-Sheng
    Chen, Chung-Hsuan
    ANALYST, 2013, 138 (24) : 7384 - 7391
  • [35] Combining in-situ quadrupole mass spectrometer analyses: Study on the structure-activity relationship and lubrication mechanisms of ammonium phosphate ionic liquids
    Qu, Guanai
    Fang, Hongling
    Li, Yi
    Zhang, Songwei
    Hu, Litian
    JOURNAL OF MOLECULAR LIQUIDS, 2024, 413
  • [36] Hyperboloid mass spectrometer with a truncated trap
    E. V. Mamontov
    D. A. Ivlev
    Technical Physics Letters, 1999, 25
  • [37] Three-dimensional simulation of heat and mass transport phenomena in planar SOFCs
    Mauro, A.
    Arpino, F.
    Massarotti, N.
    INTERNATIONAL JOURNAL OF HYDROGEN ENERGY, 2011, 36 (16) : 10288 - 10301
  • [38] Hyperboloid mass spectrometer with a truncated trap
    Mamontov, EV
    Ivlev, DA
    TECHNICAL PHYSICS LETTERS, 1999, 25 (05) : 402 - 404
  • [39] Linear and Nonlinear Resonance Ejection of High Mass Ions with charge detection rectilinear ion trap mass spectrometer
    Patil, Avinash A.
    Lee, Chen-Wei
    Chou, Szu-Wei
    Cheng, Chun-Yen
    Peng, Wen-Ping
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2020, 450
  • [40] Following the Ions through a Mass Spectrometer with Atmospheric Pressure Interface: Simulation of Complete Ion Trajectories from Ion Source to Mass Analyzer
    Zhou, Xiaoyu
    Ouyang, Zheng