共 23 条
Biosensing by Polymer-Coated Etched Long-Period Fiber Gratings Working near Mode Transition and Turn-around Point
被引:3
|作者:
Dey, Tanoy Kumar
[1
]
Trono, Cosimo
[2
]
Biswas, Palas
[1
]
Giannetti, Ambra
[2
]
Basumallick, Nandini
[1
]
Baldini, Francesco
[2
]
Bandyopadhyay, Somnath
[1
]
Tombelli, Sara
[2
]
机构:
[1] CSIR, Cent Glass & Ceram Res Inst, CGCRI, 196 Raja S C Mullick Rd, Kolkata 700032, India
[2] CNR, Ist Fis Applicata Nello Carrara, IFAC, Via Madonna Piano 10, I-50019 Sesto Fiorentino, Italy
来源:
BIOSENSORS-BASEL
|
2023年
/
13卷
/
07期
关键词:
long-period fiber grating;
label-free biosensing;
etching;
cladding mode;
sensitivity enhancement;
CLADDING-MODE;
SENSITIVITY;
OPTIMIZATION;
SENSOR;
D O I:
10.3390/bios13070731
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A methodology to enhance the sensitivity of long-period fiber gratings (LPFGs) based on the combination of three different enhancement approaches is presented; the methods here adopted are the working near mode transition (MT) of a cladding mode (CM), working near the turn-around point of a CM and the enhancement of the evanescent field of CMs by reducing the cladding diameter or by increasing the order number of CMs. In order to combine these enhancement methodologies, an electrostatic self-assembly (ESA) process was used to deposit a polymeric overlay, with a chosen thickness, onto the etched fiber. The add-layer sensitivity of the sensor was theoretically calculated, and the demonstration of the real applicability of the developed LPFG as a biosensor was performed by means of an IgG/anti-IgG immunoassay in human serum in a thermostated microfluidic system. The limits of detection (LODs) calculated by following different procedures (three times the standard deviation of the blank and the mean value of the residuals) were 6.9 x 10(-8) & mu;g/mL and 4.5 x 10(-6) & mu;g/mL, respectively. The calculated LODs demonstrate the effectiveness of the applied methodology for sensitivity enhancement.
引用
收藏
页数:15
相关论文