Image quality improvement of a one-step spectral CT reconstruction on a prototype photon-counting scanner

被引:1
|
作者
Rodesch, Pierre-Antoine [1 ]
Si-Mohamed, Salim A. [1 ,2 ]
Lesaint, Jerome [1 ]
Douek, Philippe C. [1 ,2 ]
Rit, Simon [1 ]
机构
[1] Univ Lyon, INSA Lyon, UJM St Etienne, CNRS,Inserm,CREATIS UMR5220,U1294,UCBLyon 1, F-69373 Lyon, France
[2] Louis Pradel Hosp, Dept Radiol, Hosp Civils Lyon, Bron, France
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 01期
基金
欧盟地平线“2020”;
关键词
computed tomography; spectral CT; photon-counting detector; material decomposition; image quality; one-step; COMPUTED-TOMOGRAPHY; NOISE-REDUCTION; ALGORITHM; DECOMPOSITION; PERFORMANCE; SOCIETY;
D O I
10.1088/1361-6560/ad11a3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. X-ray spectral computed tomography (CT) allows for material decomposition (MD). This study compared a one-step material decomposition MD algorithm with a two-step reconstruction MD algorithm using acquisitions of a prototype CT scanner with a photon-counting detector (PCD). Approach. MD and CT reconstruction may be done in two successive steps, i.e. decompose the data in material sinograms which are then reconstructed in material CT images, or jointly in a one-step algorithm. The one-step algorithm reconstructed material CT images by maximizing their Poisson log-likelihood in the projection domain with a spatial regularization in the image domain. The two-step algorithm maximized first the Poisson log-likelihood without regularization to decompose the data in material sinograms. These sinograms were then reconstructed into material CT images by least squares minimization, with the same spatial regularization as the one step algorithm. A phantom simulating the CT angiography clinical task was scanned and the data used to measure noise and spatial resolution properties. Low dose carotid CT angiographies of 4 patients were also reconstructed with both algorithms and analyzed by a radiologist. The image quality and diagnostic clinical task were evaluated with a clinical score. Main results. The phantom data processing demonstrated that the one-step algorithm had a better spatial resolution at the same noise level or a decreased noise value at matching spatial resolution. Regularization parameters leading to a fair comparison were selected for the patient data reconstruction. On the patient images, the one-step images received higher scores compared to the two-step algorithm for image quality and diagnostic. Significance. Both phantom and patient data demonstrated how a one-step algorithm improves spectral CT image quality over the implemented two-step algorithm but requires a longer computation time. At a low radiation dose, the one-step algorithm presented good to excellent clinical scores for all the spectral CT images.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Addressing CT metal artifacts using photon-counting detectors and one-step spectral CT image reconstruction
    Schmidt, Taly Gilat
    Sammut, Barbara A.
    Barber, Rina Foygel
    Pan, Xiaochuan
    Sidky, Emil Y.
    MEDICAL PHYSICS, 2022, 49 (05) : 3021 - 3040
  • [2] Spectral CT reconstruction with an explicit photon-counting detector model: a "one-step" approach
    Rodesch, Pierre-Antoine
    Rebuffel, V.
    Fournier, C.
    Forbes, F.
    Verger, L.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [3] Constrained one-step material decomposition reconstruction of head CT data from a silicon photon-counting prototype
    Schmidt, Taly Gilat
    Sidky, Emil Y.
    Pan, Xiaochuan
    Barber, Rina Foygel
    Gronberg, Fredrik
    Sjolin, Martin
    Danielsson, Mats
    MEDICAL PHYSICS, 2023, 50 (10) : 6008 - 6021
  • [4] A deep learning one-step solution to material image reconstruction in photon counting spectral CT
    Eguizabal, Alma
    Oktema, Ozan
    Persson, Mats
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [5] Implementation of a One-Step Iterative Reconstruction Approach for Spectral Photon-Counting Computed Tomography
    Beaulieu, W.
    Simard, M.
    Bouchard, H.
    MEDICAL PHYSICS, 2024, 51 (09) : 6555 - 6556
  • [6] One-step iterative reconstruction approach based on eigentissue decomposition for spectral photon-counting computed tomography
    Simard, Mikael
    Bouchard, Hugo
    JOURNAL OF MEDICAL IMAGING, 2022, 9 (04)
  • [7] Spectral photon-counting CT system: Toward improved image quality performance in conventional and spectral CT imaging
    Greffier, Joel
    Frandon, Julien
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2021, 102 (05) : 271 - 272
  • [8] Quantum Iterative Reconstruction for Abdominal Photon-counting Detector CT Improves Image Quality
    Sartoretti, Thomas
    Landsmann, Anna
    Nakhostin, Dominik
    Eberhard, Matthias
    Roeren, Christian
    Mergen, Victor
    Higashigaito, Kai
    Raupach, Rainer
    Alkadhi, Hatem
    Euler, Andre
    RADIOLOGY, 2022, 303 (02)
  • [9] Improved Image Reconstruction Using Multi-Energy Information in Spectral Photon-Counting CT
    Niu, Pei
    Wang, Lihui
    Xie, Bingqing
    Robini, Marc
    Boussel, Loic
    Douek, Phillippe
    Zhu, Yuemin
    Yang, Feng
    IEEE ACCESS, 2021, 9 : 97981 - 97989
  • [10] Spectral photon-counting CT in cardiovascular imaging
    Sandfort, Veit
    Persson, Mats
    Pourmorteza, Amir
    Noel, Peter B.
    Fleischmann, Dominik
    Willemink, Martin J.
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2021, 15 (03) : 218 - 225