Accuracy of respiratory gas variables, substrate, and energy use from 15 CPET systems during simulated and human exercise

被引:62
作者
Van Hooren, Bas [1 ]
Souren, Tjeu
Bongers, Bart C. [1 ,2 ]
机构
[1] Maastricht Univ, NUTRIM Sch Nutr & Translat Res Metab, Dept Nutr & Movement Sci, Univ Singel 50, NL-6229 ER Maastricht, Netherlands
[2] Maastricht Univ, NUTRIM Sch Nutr & Translat Res Metab, Dept Surg, Maastricht, Netherlands
关键词
graded exercise testing; metabolic cart; precision; reliability; simulation; validity; OPEN-CIRCUIT RESPIROMETRY; BREATH-BY-BREATH; METABOLIC SYSTEM; VALIDITY; RELIABILITY; VALIDATION; PATTERN; FITNESS; STABILITY; THRESHOLD;
D O I
10.1111/sms.14490
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Purpose: Various systems are available for cardiopulmonary exercise testing (CPET), but their accuracy remains largely unexplored. We evaluate the accuracy of 15 popular CPET systems to assess respiratory variables, substrate use, and energy expenditure during simulated exercise. Cross-comparisons were also performed during human cycling experiments (i.e., verification of simulation findings), and between-session reliability was assessed for a subset of systems.Methods: A metabolic simulator was used to simulate breath-by-breath gas exchange, and the values measured by each system (minute ventilation [V?E], breathing frequency [BF], oxygen uptake [V?O-2], carbon dioxide production [V?CO2], respiratory exchange ratio [RER], energy from carbs and fats, and total energy expenditure) were compared to the simulated values to assess the accuracy. The following manufacturers (system) were assessed: COSMED (Quark CPET, K5), Cortex (MetaLyzer 3B, MetaMax 3B), Vyaire (Vyntus CPX, Oxycon Pro), Maastricht Instruments (Omnical), MGC Diagnostics (Ergocard Clinical, Ergocard Pro, Ultima), Ganshorn/Schiller (PowerCube Ergo), Geratherm (Ergostik), VO2master (VO2masterPro), PNOE (PNOE), and Calibre Biometrics (Calibre).Results: Absolute percentage errors during the simulations ranged from 1.15%-50.3% for V?E, 1.05-3.79% for BF, 1.10%-17.5% for V?O-2, 1.07%-18.3% for V?CO2, 0.62%-14.8% for RER, 5.52%-99.0% for Kcal from carbs, 5.13%-133% for Kcal from fats, and 0.59%-12.1% for total energy expenditure. Between-session variation ranged from 0.86%-22.4% for V?O-2 and 1.14%-20.2% for V?CO2, respectively.Conclusion: The error of respiratory gas variables, substrate, and energy use differed substantially between systems, with only a few systems demonstrating a consistent acceptable error. We extensively discuss the implications of our findings for clinicians, researchers and other CPET users.
引用
收藏
页数:21
相关论文
共 67 条
[1]  
Akkermans MA, 2012, J SPORT SCI MED, V11, P182
[2]   Validity of four commercially available metabolic carts for assessing resting metabolic rate and respiratory exchange ratio in non-ventilated humans [J].
Alcantara, J. M. A. ;
Galgani, J. E. ;
Jurado-Fasoli, L. ;
Dote-Montero, M. ;
Merchan-Ramirez, E. ;
Ravussin, E. ;
Ruiz, J. R. ;
Sanchez-Delgado, G. .
CLINICAL NUTRITION, 2022, 41 (03) :746-754
[3]   Expired air temperature during prolonged exercise in cool- and hot-humid environments [J].
Atkins, KJ ;
Thompson, MW ;
Ward, JJ ;
Kelly, PT .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1997, 76 (04) :352-355
[4]  
Baldari C., 2015, Sport Sci. Heal, V11, P58
[5]  
Ballal M A, 1982, Clin Phys Physiol Meas, V3, P57, DOI 10.1088/0143-0815/3/1/005
[6]   Accuracy and precision of CPET equipment: A comparison of breath-by-breath and mixing chamber systems [J].
Beijst, Casper ;
Schep, Goof ;
Breda, Eric Van ;
Wijn, Pieter F. F. ;
Pul, Carola Van .
Journal of Medical Engineering and Technology, 2013, 37 (01) :35-42
[7]   NORMAL VALUES AND RANGES FOR VENTILATION AND BREATHING PATTERN AT MAXIMAL EXERCISE [J].
BLACKIE, SP ;
FAIRBARN, MS ;
MCELVANEY, NG ;
WILCOX, PG ;
MORRISON, NJ ;
PARDY, RL .
CHEST, 1991, 100 (01) :136-142
[8]  
Bongers B C., 2014, Pediatric norms for cardiopulmonary exercise testing in relation to sex and age, V2nd
[9]   Validation of the portable VmaxST system for oxygen-uptake measurement [J].
Brehm, MA ;
Harlaar, J ;
Groepenhof, H .
GAIT & POSTURE, 2004, 20 (01) :67-73
[10]   Power Output Manipulation from Below to Above the Gas Exchange Threshold Results in Exacerbated Performance Fatigability [J].
Brownstein, Callum G. ;
Pastor, Frederic Sabater ;
Mira, Jose ;
Murias, Juan M. ;
Millet, Guillaume Y. .
MEDICINE & SCIENCE IN SPORTS & EXERCISE, 2022, 54 (11) :1947-1960