Monte Carlo modelling of a 6 MV Elekta linear accelerator for in-field and out-of-field dosimetry

被引:4
|
作者
Al-Saleh, Wafa M. [1 ,2 ]
Hugtenburg, Richard P. [3 ,4 ]
机构
[1] King Saud Bin Abdulaziz Univ Hlth Sci, Mail Code 525,POB 2477, Alahsa 31982, Hofuf, Saudi Arabia
[2] King Abdullah Int Med Res Ctr, Alahsa, Saudi Arabia
[3] Swansea Univ, Sch Med, Singleton Pk, Swansea SA2 8PP, Wales
[4] Swansea Bay Univ Hlth Board, Singleton Hosp, Dept Med Phys & Clin Engn, Swansea SA2 8QA, Wales
关键词
Linear accelerators;
D O I
10.1016/j.radphyschem.2022.110584
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
In this study, a 6 MV Elekta linear accelerator was modelled and validated for in-field dosimetry against measurements by comparing the output factors, percentage depth dose, and lateral profiles. The resultant lateral profiles for the validated model were then compared with measurements for the out-of-field region up to 60 cm off-axis. It was initially found that there were discrepancies between the calculated and the measured profiles in the out-of-field region from 10 to 60 cm off-axis. Hence, two main adjustments were made to the length of jaws and materials of some of the head components in the LINAC model so as to be exactly as described by the manufacturer in order to acquire the best possible agreement with the measured values. Also, small head components such as the ion chamber, mirror, and Mylar sheet are included in the model to improve the results. Only in-line components were included in the model; the surrounding shielding components were excluded. The results after modifications of the head model showed good agreement with measurements up to 40 cm of field; however, from 40 to 60 cm there are still discrepancies of up to 80%, and the model underestimates the measured doses. Background radiation was measured inside the LINAC room and added as an error bar to measurements, which reduce the difference to 15% at distal points. This could possibly be due to the fact room scatter was not included in the model.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Automated Photon Monte Carlo Linear Accelerator Model for Calculating In-Field and Out-Of-Field Dose
    Randeniya, S.
    Mirkovic, D.
    Kry, S.
    Titt, U.
    Newhauser, W.
    Howell, R.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validation
    Bednarz, Bryan
    Xu, X. George
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04): : N43 - N57
  • [3] Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter
    Almberg, S. S.
    Frengen, J.
    Lindmo, T.
    MEDICAL PHYSICS, 2012, 39 (08) : 5194 - 5203
  • [4] A publicly available dataset of out-of-field dose profiles of a 6 MV linear accelerator
    Peet, Samuel C.
    Cassim, Naasiha
    Kairn, Tanya
    Trapp, Jamie, V
    Crowe, Scott B.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (02) : 613 - 621
  • [5] A publicly available dataset of out-of-field dose profiles of a 6 MV linear accelerator
    Samuel C. Peet
    Naasiha Cassim
    Tanya Kairn
    Jamie V. Trapp
    Scott B. Crowe
    Physical and Engineering Sciences in Medicine, 2022, 45 : 613 - 621
  • [6] Out-of-field beam characteristics of a 6 MV photon beam: Results of a Monte Carlo study
    Atarod, M.
    Shokrani, P.
    Azarnoosh, A.
    APPLIED RADIATION AND ISOTOPES, 2013, 72 : 182 - 194
  • [7] A Monte Carlo model for calculating out-of-field dose from a Varian 6 MV beam
    Kry, Stephen F.
    Titt, Uwe
    Ponisch, Falk
    Followill, David
    Vassiliev, Oleg N.
    White, R. Allen
    Mohan, Radhe
    Salehpour, Mohammad
    MEDICAL PHYSICS, 2006, 33 (11) : 4405 - 4413
  • [8] Monte Carlo Simulation of Cobalt-60 Teletherapy Unit Modeling In-Field and Out-Of-Field Doses for Applications in Computational Radiation Dosimetry
    Petroccia, H.
    Li, Z.
    Mendenhall, N.
    Bolch, W.
    MEDICAL PHYSICS, 2015, 42 (06) : 3464 - 3464
  • [9] A Monte Carlo simulation of out-of-field radiation from an 18-MV beam
    Kry, S
    Titt, U
    Poenisch, F
    Followill, D
    Vassiliev, O
    Mohan, R
    Salehpour, M
    MEDICAL PHYSICS, 2005, 32 (06) : 1889 - 1889
  • [10] MONTE CARLO STUDY OF THE OUT-OF-FIELD SPECTRA FOR 6MV AND 15MV PHOTON BEAMS WITH AND WITHOUT MLC.
    Theodorou, K.
    Paramerits, I.
    Tsiamas, P.
    Kappas, C.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S411 - S411