Patient-specific Monte Carlo-based organ dose estimates in spiral CT via optical 3D body scanning and adaptation of a voxelized phantom dataset: proof-of-principle

被引:1
作者
Maddaloni, Francesca Saveria [1 ,2 ]
Sarno, Antonio [1 ,2 ]
Mettivier, Giovanni [1 ,2 ]
Clemente, Stefania [3 ]
Oliviero, Caterina [3 ]
Ricciardi, Roberta [2 ,4 ]
Varallo, Antonio [2 ,4 ]
Russo, Paolo [1 ,2 ]
机构
[1] Univ Napoli Federico II, Dipartimento Fis Ettore Pancini, I-80126 Naples, Italy
[2] INFN Sez Napoli, I-80126 Naples, Italy
[3] Azienda Osped Univ Federico II, I-80131 Naples, Italy
[4] Univ Napoli Federico II, Scuola Specializzaz Fis Med, I-80131 Naples, Italy
关键词
CT dosimetry; voxelized phantoms; Monte Carlo simulations; organ doses; Geant4; 3D scanner; INTERNAL DOSIMETRY; RADIOCHROMIC FILM; SIZE; VALIDATION; TOMOSYNTHESIS; FEASIBILITY; MEDICINE; CHEST; SCANS;
D O I
10.1088/1361-6560/acc37d
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. We present a method for personalized organ dose estimates obtained before the computed tomography (CT) exam, via 3D optical body scanning and Monte Carlo (MC) simulations. Approach. A voxelized phantom is derived by adapting a reference phantom to the body size and shape measured with a portable 3D optical scanner, which returns the 3D silhouette of the patient. This was used as an external rigid envelope for incorporating a tailored version of the internal body anatomy derived from a phantom dataset (National Cancer Institute, NIH, USA) matched for gender, age, weight, and height. The proof-of-principle was conducted on adult head phantoms. The Geant4 MC code provided estimates of the organ doses from 3D absorbed dose maps in the voxelized body phantom. Main results. We applied this approach for head CT scanning using an anthropomorphic voxelized head phantom derived from 3D optical scans of manikins. We compared the estimates of head organ doses with those provided by the NCICT 3.0 software (NCI, NIH, USA). Head organ doses differed up to 38% using the proposed personalized estimate and MC code, with respect to corresponding estimates calculated for the standard (non-personalized) reference head phantom. Preliminary application of the MC code to chest CT scans is shown. Real-time pre-exam personalized CT dosimetry is envisaged with adoption of a Graphics Processing Unit-based fast MC code. Significance. The developed procedure for personalized organ dose estimates before the CT exam, introduces a new approach for realistic description of size and shape of patients via voxelized phantoms specific for each patient.
引用
收藏
页数:17
相关论文
共 47 条
  • [1] AAPM, 2011, 204 AAPM, DOI [10.37206/143, DOI 10.37206/143]
  • [2] AAPM, 2019, 293 AAPM, DOI [10.37206/185, DOI 10.37206/185]
  • [3] Estimation of size-specific dose estimates (SSDE) for paediatric and adults patients based on a single slice
    Abuhaimed, Abdullah
    Martin, Colin J.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 74 : 30 - 39
  • [4] American Association of Physicists in Medicine, 2019, a joint report of AAPM task group 246 and the European Federation of organizations for medical physics (EFOMP) Report No: 246, DOI [10.37206/190, DOI 10.37206/190]
  • [5] [Anonymous], 2019, 184 NAT COUNC RAD PR
  • [6] Increasing organ dose accuracy through voxel phantom organ matching with individual patient anatomy
    Borbinha, Jorge
    Di Maria, Salvatore
    Madeira, Paula
    Belchior, Ana
    Baptista, Mariana
    Vaz, Pedro
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2019, 159 : 35 - 46
  • [7] Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task
    Brock, Kristy K.
    Mutic, Sasa
    McNutt, Todd R.
    Li, Hua
    Kessler, Marc L.
    [J]. MEDICAL PHYSICS, 2017, 44 (07) : E43 - E76
  • [8] CT Dosimetry What Has Been Achieved and What Remains to Be Done
    Damilakis, John
    [J]. INVESTIGATIVE RADIOLOGY, 2021, 56 (01) : 62 - 68
  • [9] Patient organ and effective dose estimation in CT: comparison of four software applications
    De Mattia, Cristina
    Campanaro, Federica
    Rottoli, Federica
    Colombo, Paola Enrica
    Pola, Andrea
    Vanzulli, Angelo
    Torresin, Alberto
    [J]. EUROPEAN RADIOLOGY EXPERIMENTAL, 2020, 4 (01)
  • [10] GEANT4 Monte Carlo simulations for virtual clinical trials in breast X-ray imaging: Proof of concept
    di Franco, F.
    Sarno, A.
    Mettivier, G.
    Hernandez, A. M.
    Bliznakova, K.
    Boone, J. M.
    Russo, P.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 74 : 133 - 142