Evaluation of fibrin, cyanoacrylate, and polyurethane-based tissue adhesives in sutureless vascular anastomosis: a comparative mechanical ex vivo study

被引:4
|
作者
Heitzer, M. [1 ]
Kilic, K. [1 ]
Merfort, R. [2 ]
Emonts, C. [3 ]
Winnand, P. [1 ]
Kniha, K. [1 ]
Hoelzle, F. [1 ]
Modabber, A. [1 ]
机构
[1] Univ Hosp RWTH Aachen, Dept Oral & Maxillofacial Surg, Pauwelsstr 13, D-52074 Aachen, Germany
[2] Univ Hosp RWTH Aachen, Dept Orthopaed Trauma & Reconstruct Surg, Aachen, Germany
[3] Rhein Westfal TH Aachen, Inst Textiltech, Aachen, Germany
关键词
Tissue adhesives; Microsurgery; Tensile strength; Surgical anastomosis; Sutureless surgical procedures; MICROVASCULAR ANASTOMOSIS; RECONSTRUCTION; STENTS; GLUE;
D O I
10.1016/j.ijom.2023.03.011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The stability of a microvascular anastomosis is an important prerequisite for successful tissue transfer. Advances in tissue adhesives are potentially opening new avenues for their use in sutureless microsurgical anastomosis, however they have not yet gained clinical acceptance. In this ex vivo study, a novel polyurethane-based adhesive (PA) was used in sutureless anastomoses and its stability compared with that of sutureless anastomoses performed with fibrin glue (FG) and a cyanoacrylate (CA). Stability was assessed using hydrostatic (15 per group) and mechanical tests (13 per group). A total of 84 chicken femoral arteries were used in this study. The time taken to create the PA and CA anastomoses was significantly faster when compared to the FG anastomoses (P < 0.001): 1.55 +/- 0.14 min and 1.39 +/- 0.06 min, respectively, compared to 2.03 +/- 0.35 min. Both sustained significantly higher pressures (289.3 mmHg and 292.7 mmHg, respectively) than anastomoses using FG (137.3 mmHg) (P < 0.001). CA anastomoses (0.99 N; P < 0.001) and PA anastomoses (0.38 N; P = 0.009) could both withstand significantly higher longitudinal tensile forces compared to FG anastomoses (0.10 N). Considering the background of an in vitro study, the PA and CA anastomosis techniques were shown to be similar to each other and superior to FG, due to their stability and faster handling. These findings need to be validated and confirmed in further in vivo studies.
引用
收藏
页码:1137 / 1144
页数:8
相关论文
共 2 条
  • [1] Mechanical Strength and Rheological Properties of Tissue Adhesives With Regard to Colorectal Anastomosis An Ex Vivo Study
    Vakalopoulos, Konstantinos Aristotelis
    Wu, Zhouqiao
    Kroese, Leonard
    Kleinrensink, Gert-Jan
    Jeekel, Johannes
    Vendamme, Richard
    Dodou, Dimitra
    Lange, Johan Frederik
    ANNALS OF SURGERY, 2015, 261 (02) : 323 - 331
  • [2] Biodegradation and Immunological Parameters of Polyurethane-Based Tissue Adhesive in Arterial Microvascular Anastomoses-A Long-Term In Vivo Study
    Modabber, Ali
    Winnand, Philipp
    Goloborodko, Evgeny
    Moehlhenrich, Stephan Christian
    Kniha, Kristian
    Tolba, Rene
    Jockenhoevel, Stefan
    Hermanns-Sachweh, Benita
    Hoelzle, Frank
    Heitzer, Marius
    MACROMOLECULAR BIOSCIENCE, 2022, 22 (04)