Video analysis of ex vivo beating hearts during preservation on the TransMedics® organ care system

被引:4
作者
Mendiola Pla, Michelle [1 ]
Berrettoni, Silvia [2 ]
Lee, Franklin H. H. [1 ]
Rozzi, Giacomo [2 ]
Marrano, Federica [2 ]
Gross, Ryan T. T. [1 ]
Evans, Amy [3 ]
Wendell, David C. C. [4 ]
Lezberg, Paul [5 ]
Burattini, Margherita [2 ,6 ]
lo Muzio, Francesco Paolo [2 ]
Fassina, Lorenzo [7 ]
Milano, Carmelo A. A. [1 ]
Bang, Marie-Louise [8 ,9 ]
Bowles, Dawn E. E. [1 ]
Miragoli, Michele [2 ,9 ]
机构
[1] Duke Univ, Med Ctr, Dept Surg, Durham, NC USA
[2] Univ Parma, Dept Med & Surg, Parma, Italy
[3] Duke Univ, Med Ctr, Perfus Serv, Durham, NC USA
[4] Duke Univ, Duke Cardiovasc Magnet Resonance Ctr, Med Ctr, Durham, NC USA
[5] TransMed Inc, Andover, MA USA
[6] Univ Verona, Dept Surg Sci Dent & Matern, Verona, Italy
[7] Univ Pavia, Dept Elect Comp & Biomed Engn, Pavia, Italy
[8] CNR, Inst Genet & Biomed Res IRGB, Milan Unit, Milan, Italy
[9] IRCCS Humanitas Res Hosp, Milan, Italy
基金
美国国家卫生研究院;
关键词
ex vivo perfusion; normothermic; video; kinematics; biomarker; cardiac transplantation; TRANSPLANTATION;
D O I
10.3389/fcvm.2023.1216917
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundReliable biomarkers for assessing the viability of the donor hearts undergoing ex vivo perfusion remain elusive. A unique feature of normothermic ex vivo perfusion on the TransMedics & REG; Organ Care System (OCS & TRADE;) is that the donor heart is maintained in a beating state throughout the preservation period. We applied a video algorithm for an in vivo assessment of cardiac kinematics, video kinematic evaluation (Vi.Ki.E.), to the donor hearts undergoing ex vivo perfusion on the OCS & TRADE; to assess the feasibility of applying this algorithm in this setting. MethodsHealthy donor porcine hearts (n = 6) were procured from Yucatan pigs and underwent 2 h of normothermic ex vivo perfusion on the OCS & TRADE; device. During the preservation period, serial high-resolution videos were captured at 30 frames per second. Using Vi.Ki.E., we assessed the force, energy, contractility, and trajectory parameters of each heart. ResultsThere were no significant changes in any of the measured parameters of the heart on the OCS & TRADE; device over time as judged by linear regression analysis. Importantly, there were no significant changes in contractility during the duration of the preservation period (time 0-30 min, 918 & PLUSMN; 430 px/s; time 31-60 min, 1,386 & PLUSMN; 603 px/s; time 61-90 min, 1,299 & PLUSMN; 617 px/s; time 91-120 min, 1,535 & PLUSMN; 728 px/s). Similarly, there were no significant changes in the force, energy, or trajectory parameters. Post-transplantation echocardiograms demonstrated robust contractility of each allograft. ConclusionVi.Ki.E. assessment of the donor hearts undergoing ex vivo perfusion is feasible on the TransMedics OCS & TRADE;, and we observed that the donor hearts maintain steady kinematic measurements throughout the duration.
引用
收藏
页数:7
相关论文
共 13 条
[1]   Lactate during ex-situ heart perfusion does not predict the requirement for mechanical circulatory support following donation after circulatory death (DCD) heart transplants [J].
Cernic, Sendi ;
Page, Aravinda ;
Messer, Simon ;
Bhagra, Sai ;
Pettit, Stephen ;
Dawson, Sarah N. ;
Mckie, Mikel Alexander ;
Osman, Mohamed ;
Nachum, Eyal ;
White, Daniel ;
Ali, Jason ;
Fingelton, Lorraine ;
Howlett, Thomas ;
Tsui, Steven ;
Baxter, Jen ;
Falter, Florian ;
Jenkins, David P. ;
Kaul, Pradeep ;
Berman, Marius ;
Catarino, Pedro ;
Large, Stephen R. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2022, 41 (09) :1294-1302
[2]   Cardiac kinematic parameters computed from video of in situ beating heart [J].
Fassina, Lorenzo ;
Rozzi, Giacomo ;
Rossi, Stefano ;
Scacchi, Simone ;
Galetti, Maricla ;
Lo Muzio, Francesco Paolo ;
Del Bianco, Fabrizio ;
Franzone, Piero Colli ;
Petrilli, Giuseppe ;
Faggian, Giuseppe ;
Miragoli, Michele .
SCIENTIFIC REPORTS, 2017, 7
[3]   "Resuscitation" of marginal liver allografts for transplantation with machine perfusion technology [J].
Graham, Jay A. ;
Guarrera, James V. .
JOURNAL OF HEPATOLOGY, 2014, 61 (02) :418-431
[4]   Primary graft dysfunction in heart transplantation: How to recognize it, when to institute extracorporeal membrane oxygenation, and outcomes [J].
Hull, Travis D. ;
Crowley, Jerome C. ;
Villavicencio, Mauricio A. ;
D'Alessandro, David A. .
JTCVS OPEN, 2021, 8 :128-133
[5]   Normothermic Ex Vivo Perfusion Provides Superior Organ Preservation and Enables Viability Assessment of Hearts From DCD Donors [J].
Iyer, A. ;
Gao, L. ;
Doyle, A. ;
Rao, P. ;
Cropper, J. R. ;
Soto, C. ;
Dinale, A. ;
Kumarasinghe, G. ;
Jabbour, A. ;
Hicks, M. ;
Jansz, P. C. ;
Feneley, M. P. ;
Harvey, R. P. ;
Graham, R. M. ;
Dhital, K. K. ;
MacDonald, P. S. .
AMERICAN JOURNAL OF TRANSPLANTATION, 2015, 15 (02) :371-380
[6]   Normothermic Ex Situ Heart Perfusion With the Organ Care System for Cardiac Transplantation: A Meta-analysis [J].
Langmuur, Sanne J. J. ;
Amesz, Jorik H. ;
Veen, Kevin M. ;
Bogers, Ad J. J. C. ;
Manintveld, Olivier C. ;
Taverne, Yannick J. H. J. .
TRANSPLANTATION, 2022, 106 (09) :1745-1753
[7]   Artificial Intelligence Supports Decision Making during Open-Chest Surgery of Rare Congenital Heart Defects [J].
Lo Muzio, Francesco Paolo ;
Rozzi, Giacomo ;
Rossi, Stefano ;
Luciani, Giovanni Battista ;
Foresti, Ruben ;
Cabassi, Aderville ;
Fassina, Lorenzo ;
Miragoli, Michele .
JOURNAL OF CLINICAL MEDICINE, 2021, 10 (22)
[8]   A Porcine Heterotopic Heart Transplantation Protocol for Delivery of Therapeutics to a Cardiac Allograft [J].
Pla, Michelle Mendiola ;
Evans, Amy ;
Lee, Franklin H. ;
Chiang, Yuting ;
Bishawi, Muath ;
Vekstein, Andrew ;
Kang, Lillian ;
Zapata, Diego ;
Gross, Ryan ;
Carnes, Alexis ;
Gault, Lynden E. ;
Balko, Julie A. ;
Bonadonna, Desiree ;
Ho, Sam ;
Lezberg, Paul ;
Bryner, Benjamin S. ;
Schroder, Jacob N. ;
Milano, Carmelo A. ;
Bowles, Dawn E. .
JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2022, (180)
[9]   Machine Perfusion for Human Heart Preservation: A Systematic Review [J].
Qin, Guangqi ;
Jernryd, Victoria ;
Sjoberg, Trygve ;
Steen, Stig ;
Nilsson, Johan .
TRANSPLANT INTERNATIONAL, 2022, 35
[10]   Right ventricular functional recovery depends on timing of pulmonary valve replacement in tetralogy of Fallot: a video kinematic study [J].
Rozzi, Giacomo ;
Lo Muzio, Francesco Paolo ;
Fassina, Lorenzo ;
Rossi, Stefano ;
Statello, Rosario ;
Sandrini, Camilla ;
Laricchiuta, Maira ;
Faggian, Giuseppe ;
Miragoli, Michele ;
Luciani, Giovanni Battista .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2021, 59 (06) :1329-1336