Deep learning-based protoacoustic signal denoising for proton range verification

被引:3
|
作者
Wang, Jing [1 ,2 ]
Sohn, James J. [3 ]
Lei, Yang [1 ,2 ]
Nie, Wei [4 ]
Zhou, Jun [1 ,2 ]
Avery, Stephen [5 ]
Liu, Tian [6 ]
Yang, Xiaofeng [1 ,2 ]
机构
[1] Emory Univ, Dept Radiat Oncol, Atlanta, GA 30322 USA
[2] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
[3] Northwestern Univ, Dept Radiat Oncol, Chicago, IL USA
[4] Inova Schar Canc Inst, Radiat Oncol Div, Fairfax, VA USA
[5] Univ Penn, Dept Radiat Oncol, Philadelphia, PA USA
[6] Mt Sinai Med Ctr, Dept Radiat Oncol, New York, NY 10029 USA
基金
美国国家卫生研究院;
关键词
protoacoustic; signal denoising; Bragg peak; deep learning; stack auto-encoder; POSITRON-EMISSION-TOMOGRAPHY; BEAM; THERAPY; AUTOENCODERS; DELIVERY;
D O I
10.1088/2057-1976/acd257
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Proton therapy is a type of radiation therapy that can provide better dose distribution compared to photon therapy by delivering most of the energy at the end of range, which is called the Bragg peak (BP). The protoacoustic technique was developed to determine the BP locations in vivo, but it requires a large dose delivery to the tissue to obtain a high number of signal averaging (NSA) to achieve a sufficient signal-to-noise ratio (SNR), which is not suitable for clinical use. A novel deep learning-based technique has been proposed to denoise acoustic signals and reduce BP range uncertainty with much lower doses. Three accelerometers were placed on the distal surface of a cylindrical polyethylene (PE) phantom to collect protoacoustic signals. In total, 512 raw signals were collected at each device. Device-specific stack autoencoder (SAE) denoising models were trained to denoise the noise-containing input signals, which were generated by averaging only 1, 2, 4, 8, 16, or 24 raw signals (low NSA signals), while the clean signals were obtained by averaging 192 raw signals (high NSA). Both supervised and unsupervised training strategies were employed, and the evaluation of the models was based on mean squared error (MSE), SNR, and BP range uncertainty. Overall, the supervised SAEs outperformed the unsupervised SAEs in BP range verification. For the high accuracy detector, it achieved a BP range uncertainty of 0.20 +/- 3.44 mm by averaging over 8 raw signals, while for the other two low accuracy detectors, they achieved the BP uncertainty of 1.44 +/- 6.45 mm and -0.23 +/- 4.88 mm by averaging 16 raw signals, respectively. This deep learning-based denoising method has shown promising results in enhancing the SNR of protoacoustic measurements and improving the accuracy in BP range verification. It greatly reduces the dose and time for potential clinical applications.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Wavelet-based Protoacoustic Signal Denoising for Proton Range Verification
    Sohn, James
    Nie, Wei
    Avery, Stephen M.
    He, Xiuxiu
    Zhou, Jun
    Yang, Xiaofeng
    Liu, Tian
    MEDICAL IMAGING 2020: ULTRASONIC IMAGING AND TOMOGRAPHY, 2020, 11319
  • [2] Wavelet-Based Signal Processing Approach for Protoacoustic Measurement in Proton Range Verification
    Sohn, J.
    Nie, W.
    Yang, X.
    He, X.
    Zhou, J.
    Avery, S.
    Liu, T.
    MEDICAL PHYSICS, 2019, 46 (06) : E602 - E602
  • [3] Deep Learning-Based Prompt-Gamma Imaging for Range Verification in Proton Radiotherapy
    Jiang, Z.
    Polf, J.
    Ren, L.
    MEDICAL PHYSICS, 2022, 49 (06) : E210 - E210
  • [4] CT-Based Protoacoustic Image Reconstruction for Range Verification in Proton Therapy
    Yu, Y.
    Qi, P.
    Peng, H.
    Xing, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E484 - E484
  • [5] Dictionary-based protoacoustic dose map imaging for proton range verification
    Freijo, Clara
    Herraiz, Joaquin L.
    Sanchez-Parcerisa, Daniel
    Manuel Udias, Jose
    PHOTOACOUSTICS, 2021, 21 (21)
  • [6] Time Reversal-Based Protoacoustic Reconstruction for Range and Dose Verification in Proton Therapy
    Yu, Y.
    Zhang, D.
    Xing, L.
    Peng, H.
    MEDICAL PHYSICS, 2019, 46 (06) : E366 - E366
  • [7] A Plan Verification Platform for Online Adaptive Proton Therapy Using Deep Learning-Based Monte-Carlo Denoising
    Zhang, G.
    Chen, X.
    Dai, J.
    Men, K.
    MEDICAL PHYSICS, 2022, 49 (06) : E792 - E792
  • [8] A plan verification platform for online adaptive proton therapy using deep learning-based Monte-Carlo denoising
    Zhang, Guoliang
    Chen, Xinyuan
    Dai, Jianrong
    Men, Kuo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 103 : 18 - 25
  • [9] A deep learning-based framework For ECG signal denoising based on stacked cardiac cycle tensor
    Rasti-Meymandi, Arash
    Ghaffari, Aboozar
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2022, 71
  • [10] Simulation studies of time reversal-based protoacoustic reconstruction for range and dose verification in proton therapy
    Yu, Yajun
    Li, Zhongxing
    Zhang, Dong
    Xing, Lei
    Peng, Hao
    MEDICAL PHYSICS, 2019, 46 (08) : 3649 - 3662