Accuracy to detect neonatal hyperglycaemia using real-time continuous glucose monitoring during postoperative period

被引:1
作者
Wasiq, Mohammed Abdul [1 ]
Behura, Sushree Smita [1 ]
Sahoo, Swaranjika [1 ]
Panda, Santosh Kumar [1 ]
机构
[1] Kalinga Inst Med Sci, Dept Pediat, Bhubaneswar 751024, Orissa, India
关键词
Blood glucose; Continuous glucose monitoring; Hyperglycaemia; Neonates; CONSENSUS;
D O I
10.1007/s00431-022-04777-1
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Identifying hyperglycaemia during postoperative period is essential for neonates. The objective of the study was to analyse the accuracy and reliability of continuous glucose monitoring (CGM) device for detecting hyperglycaemia during postoperative period in neonates. In this prospective study, hourly glucose recordings by CGM device and six hourly by glucometer glucose (GG-reference test for patient management) were collected in ten surgical neonates during first three postoperative days. Mean absolute relative difference (MARD) and proportion of CGM values within +/- 15%/15mg/dL, +/- 20%/20 mg/dL, and +/- 30%/30 mg/dL of GG, were analysed from matched pair CGM and GG recordings. The diagnostic performance of CGM for neonatal hyperglycaemia (> 150 mg/dL) was expressed as sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV); and the clinical reliability was assessed in Clarke Error Grid Analysis. A total of 720 CGM glucose readings, 120 matched-paired glucosereadings by CGM --GG, and 37 episodes were hyperglycaemia by GG. The MARD was 10.76 %; the per-centages of glucose readings within 15%/15 mg/dL, 20%/20 mg/dL, and 30%/30 mg/dL were 94.6%, 97.3% and 100% during the hyperglycaemia period. The sensitivity, specificity, PPV and NPV to detect hyperglycaaemia by CGM device were 100%, 93.9%, 88% and 100 % respectively. In Clarke Error Grid Analysis, 97.3 % points were in zone A and B during the hyperglycaemia period. Conclusion: CGM device can be a clinically reliable tool for hyperglycaemia management during postoperative period in neonates.
引用
收藏
页码:1083 / 1087
页数:5
相关论文
共 17 条
[1]   ASPEN Clinical Guidelines: Hyperglycemia and Hypoglycemia in the Neonate Receiving Parenteral Nutrition [J].
Arsenault, Danielle ;
Brenn, Megan ;
Kim, Sendia ;
Gura, Kathleen ;
Compher, Charlene ;
Simpser, Edwin ;
Puder, Mark .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2012, 36 (01) :81-95
[2]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[3]   Feasibility of automated insulin delivery guided by continuous glucose monitoring in preterm infants [J].
Beardsall, Kathryn ;
Thomson, Lynn ;
Elleri, Daniela ;
Dunger, David B. ;
Hovorka, Roman .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2020, 105 (03) :F279-F284
[4]  
Bischoff Adrianne R, 2021, J Neonatal Perinatal Med, V14, P183, DOI 10.3233/NPM-200535
[5]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[6]   International Consensus on Use of Continuous Glucose Monitoring [J].
Danne, Thomas ;
Nimri, Revital ;
Battelino, Tadej ;
Bergenstal, Richard M. ;
Close, Kelly L. ;
DeVries, J. Hans ;
Garg, Satish ;
Heinemann, Lutz ;
Hirsch, Irl ;
Amiel, Stephanie A. ;
Beck, Roy ;
Bosi, Emanuele ;
Buckingham, Bruce ;
Cobelli, Claudio ;
Dassau, Eyal ;
Doyle, Francis J., III ;
Heller, Simon ;
Hovorka, Roman ;
Jia, Weiping ;
Jones, Tim ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Maahs, David ;
Murphy, Helen R. ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Saboo, Banshi ;
Scharf, Mauro ;
Tamborlane, William V. ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2017, 40 (12) :1631-1640
[7]  
Funtanilla Vienica D, 2019, P T, V44, P550
[8]   Continuous glucose monitoring for the prevention of morbidity and mortality in preterm infants [J].
Galderisi, Alfonso ;
Bruschettini, Matteo ;
Russo, Chiara ;
Hall, Rebecka ;
Trevisanuto, Daniele .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2020, (12)
[9]  
Mondal, 2021, CLARKE ERROR GRID AN, DOI [10.6084/m9.figshare.17166620.v1, DOI 10.6084/M9.FIGSHARE.17166620.V1]
[10]   Accuracy of continuous glucose monitoring in preterm infants: a systematic review and meta-analysis [J].
Nava, Chiara ;
Hedenmalm, Astrid Modiano ;
Borys, Franciszek ;
Hooft, Lotty ;
Bruschettini, Matteo ;
Jenniskens, Kevin .
BMJ OPEN, 2020, 10 (12)