A 3D Mathematical Breast Texture Model With Parameters Automatically Inferred From Clinical Breast CT Images

被引:4
作者
Li, Zhijin [1 ]
Carton, Ann-Katherine [1 ]
Muller, Serge [1 ]
Almecija, Thomas [2 ]
de Carvalho, Pablo Milioni [1 ]
Desolneux, Agnes [3 ]
机构
[1] GE Healthcare France, F-78530 Buc, France
[2] GE Renewable Energy France, F-92100 Boulogne Billancourt, France
[3] Ecole Normale Super Paris Saclay, Ctr Borelli, F-91190 Gif Sur Yvette, France
关键词
Breast; Three-dimensional displays; Mathematical models; Computed tomography; Ellipsoids; Solid modeling; Phantoms; X-ray breast imaging; virtual clinical trials; 3D breast texture model; stochastic geometry; statistical parameter inference; SOFTWARE PHANTOM; SIMULATION; TOMOSYNTHESIS; OPTIMIZATION; STATISTICS;
D O I
10.1109/TMI.2022.3224223
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A numerical realistic 3D anthropomorphic breast model is useful for evaluating breast imaging applications. A method is proposed to model small and medium-scale fibroglandular and intra-glandular adipose tissues observed in the center part of clinical breast CT images. The method builds upon a previously proposed model formulated as stochastic geometric processes with mathematically tractable parameters. In this work, the medium-scale parameters were automatically and objectively inferred from breast CT images. We hypothesized that a set of random ellipsoids exhibiting cluster interaction is representative to model the medium-scale intra-glandular adipose compartments. The ellipsoids were reconstructed using a multiple birth, death and shift algorithm. Then, a Matern cluster process was used to fit the reconstructed ellipsoid centers. Finally, distributions of the ellipsoid shapes and orientations were estimated using maximum likelihood estimators. Feasibility was demonstrated on 16 volumes of interests (VOI). To assess the realism of the 3D breast texture model, beta and LFE metrics computed in simulated projection images of simulated texture realizations and clinical images were compared. Visual realism was illustrated. For 12 out of 16 VOIs, our hypothesis on clustering interaction process is confirmed. The average beta-values from simulated texture images (3.7 to 4.2) of the 12 different VOIs are higher than the average beta value from 2D clinical images (2.87). LFE of simulated texture images and clinical mammograms are similar. Compared to our previous model, whereby simulation parameters were based upon empirical observations, our inference method substantially augments the ability to generate textures with higher visual realism and larger morphological variety.
引用
收藏
页码:1107 / 1120
页数:14
相关论文
共 62 条
[41]   Image browsing in slow medical liquid crystal displays [J].
Liang, Hongye ;
Park, Subok ;
Gallas, Brandon D. ;
Myers, Kyle J. ;
Badano, Aldo .
ACADEMIC RADIOLOGY, 2008, 15 (03) :370-382
[42]   Dedicated breast CT: Initial clinical experience [J].
Lindfors, Karen K. ;
Boone, John M. ;
Nelson, Thomas R. ;
Yang, Kai ;
Kwan, Alexander L. C. ;
Miller, DeWitt F. .
RADIOLOGY, 2008, 246 (03) :725-733
[43]   Ellipsoids and matrix-valued valuations [J].
Ludwig, M .
DUKE MATHEMATICAL JOURNAL, 2003, 119 (01) :159-188
[44]  
Lutwak E, 2000, DUKE MATH J, V104, P375
[45]   Three-Dimensional In Silico Breast Phantoms for Multimodal Image Simulations [J].
Mahr, David M. ;
Bhargava, Rohit ;
Insana, Michael F. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) :689-697
[46]   Full-size anthropomorphic phantom for 2D and 3D breast x-ray imaging [J].
Mainprize, James G. ;
Mawdsley, Gordon E. ;
Carton, Ann-Katherine ;
Li, Zhijin ;
Klausz, Remy ;
Muller, Serge ;
Yaffe, Martin J. .
15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
[47]   The relationship between anatomic noise and volumetric breast density for digital mammography [J].
Mainprize, James G. ;
Tyson, Albert H. ;
Yaffe, Martin J. .
MEDICAL PHYSICS, 2012, 39 (08) :4660-4668
[48]   Evaluation of the visual realism of breast texture phantoms in digital mammography [J].
Marinov, Stoyko ;
Carton, Ann-Katherine ;
Cockmartin, Lesley ;
Marshall, Nicholas ;
Van Ongeval, Chantal ;
De la Rosa, Ruben Sanchez ;
Muller, Serge ;
Bosmans, Hilde .
15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
[49]   THE KOLMOGOROV-SMIRNOV TEST FOR GOODNESS OF FIT [J].
MASSEY, FJ .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1951, 46 (253) :68-78
[50]   Characterizing anatomical variability in breast CT images [J].
Metheany, Kathrine G. ;
Abbey, Craig K. ;
Packard, Nathan ;
Boone, John M. .
MEDICAL PHYSICS, 2008, 35 (10) :4685-4694