A Monte Carlo study to evaluate and optimise the angular dependence of the Octa-A 2D silicon array detector used for dosimetry in stereotactic radiotherapy

被引:3
|
作者
Filipev, Ilia [1 ,4 ]
Biasi, Giordano [1 ,2 ]
Poder, Joel [1 ,3 ]
Lerch, Michael [1 ]
Petasecca, Marco [1 ]
Guatelli, Susanna [1 ]
Kron, Tomas [1 ,2 ]
Rosenfeld, Anatoly [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[2] Peter MacCallum Canc Ctr, Melbourne, Vic, Australia
[3] St George Hosp, Canc Care Ctr, Kogarah, NSW, Australia
[4] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
关键词
Stereotactic ablative body radiotherapy; Smallfield dosimetry; 2D monolithic silicon array detector; Angular dependence; Geant4; Monte Carlo simulation; BODY RADIATION-THERAPY; SPINAL METASTASES; SIMULATION; TRANSPORT; FIELDS; DIODE;
D O I
10.1016/j.radmeas.2023.106979
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Purpose: A detector assembly based on a 2D silicon array sensor, Octa, was previously demonstrated to be accurate for small field relative dosimetry. Experimental studies showed angular dependence of the response that has also variations with the field size. A Monte Carlo simulation model was developed and used in this work to study how the detector assembly's components contribute to the observed angular dependence.Methods: Geant4 was used as the Monte Carlo code. A detailed model of the Octa detector and its packaging was placed in the centre of a solid water phantom with the central sensitive volumes (SVs) located at the isocentre and irradiated at polar angles from 0 to 180 degrees (the beam was normal to the detector plane at 0 degrees). Phase-space files from the IAEA database were used to model a CyberKnife X-ray beam with collimator diameters of 5, 10, and 15 mm. Reference simulations were compared with previous experimental results. To investigate the impact of the assembly's materials, simulations were conducted by iteratively substituting the non-water-equivalent materials of certain components with solid water. In terms of geometry, the influence of the dimensions of an air gap on top of the SVs and the silicon substrate thickness were investigated.Results: Presence of the silicon surrounding the detector's SVs caused the highest decrease in energy deposition - up to (27 & PLUSMN; 1)%. Thickness of the air gap placed on top of the SVs had no effect at angles below 90 degrees, but for 90-180 degrees, it showed the largest impact on the variation in angular dependence between fields of different sizes - from (6 & PLUSMN; 2)% for 1 mm gap to (24 & PLUSMN; 2)% for 2 mm gap. The variation in the silicon substrate thickness and the air gap's lateral length, as well as presence of the PCB layer did not show significant influence on the angular response.Conclusion: Further development of the detector assembly's design can be driven by mechanical robustness to provide reliability for clinical routine without sacrificing angular dependence. Current design is recommended for use at angles 0-90 degrees: it has shown independence of the angular response from small field size variation and it was previously optimised to be correction-free in output factor measurements. The air gap dimensions are recommended as the main variable parameters for optimisation for further studies when small X-ray beams are measured by the detector positioned in axial plane.
引用
收藏
页数:9
相关论文
共 6 条
  • [1] Technical Note: Angular dependence of a 2D monolithic silicon diode array for small field dosimetry
    Stansook, Nauljun
    Utitsarn, Kananan
    Petasecca, Marco
    Newall, Matthew K.
    Duncan, Mitchell
    Nitschke, Kym
    Carolan, Martin
    Metcalfe, Peter
    Lerch, Michael L. F.
    Perevertaylo, Vladimir L.
    Tome, Wolfgang A.
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (08) : 4313 - 4321
  • [2] A 2D silicon detector array for quality assurance in small field dosimetry: DUO
    Al Shukaili, Khalsa
    Petasecca, Marco
    Newall, Matthew
    Espinoza, Anthony
    Perevertaylo, Vladimir L.
    Corde, Stephanie
    Lerch, Michael
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (02) : 628 - 636
  • [3] Dosimetric characteristics of a 2D silicon diode array for stereotactic radiotherapy end-to-end patient-specific QA
    Bai, Long
    Xiao, Qing
    Wang, Qiang
    Zhao, Jianling
    Li, Guangjun
    Bai, Sen
    RADIATION PHYSICS AND CHEMISTRY, 2020, 173
  • [4] HDR brachytherapy invivo source position verification using a 2D diode array: A Monte Carlo study
    Poder, Joel
    Cutajar, Dean
    Guatelli, Susanna
    Petasecca, Marco
    Howie, Andrew
    Bucci, Joseph
    Rosenfeld, Anatoly
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 163 - 172
  • [5] A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams
    Biasi, G.
    Petasecca, M.
    Guatelli, S.
    Hardcastle, N.
    Carolan, M.
    Perevertaylo, V.
    Kron, T.
    Rosenfeld, A. B.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 45 : 117 - 126
  • [6] Monte Carlo simulation of a 2D dynamic multileaf collimator to improve the plan quality in radiotherapy plan: a proof-of-concept study
    Park, Hyojun
    Choi, Hyun-Joon
    Min, Chul Hee
    Kim, Jung-in
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24)