A hybrid FDTD/MoM algorithm with a non-uniform grid for MRI RF coil design

被引:3
作者
Liu, Yang [1 ,2 ]
Wang, Qiuliang [1 ,2 ]
Liu, Feng [3 ]
机构
[1] Chinese Acad Sci, Inst Elect Engn, Key Lab Appl Superconduct, Beijing 100190, Peoples R China
[2] Univ Chinese Acad Sci, Sch Elect Elect & Commun Engn, Beijing 100049, Peoples R China
[3] Univ Queensland, Sch Informat Technol & Elect Engn, Brisbane, Qld 4072, Australia
基金
中国国家自然科学基金;
关键词
RF coil; Hybrid method; FDTD; MoM; Numerical analysis; ELECTROMAGNETIC SCATTERING; HUMAN HEAD; FIELD;
D O I
10.1016/j.mri.2022.10.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In ultra-high-field (UHF) magnetic resonance imaging (MRI) applications, the design and analysis of highfrequency radio frequency (RF) coils requires full-wave electromagnetic (EM) methods that can handle complex field-tissue interactions. Using a Huygens' equivalent surface, the Method of Moments (MoM) and the FiniteDifference Time-Domain (FDTD) algorithm can be combined to accurately model the high-frequency RF coils. In previous research, a uniform FDTD mesh structure was considered, providing a compromised solution for coiltissue interactions. This paper proposes a hybrid FDTD/MoM algorithm with non-uniform meshes. The fine mesh domain is set at the Huygens' surface, and the other domain uses coarse meshes. The proposed algorithms are strictly validated, and their computational performance is compared against conventional methods. Results show that the new algorithm can improve the calculation efficiency without losing accuracy. Specifically, compared with the uniform FDTD method, the numerical difference between both hybrid methods remains at 3.2%. Still, the calculation time of the non-uniform grid algorithm is reduced by 64.2%, demonstrating the effectiveness of the new algorithm for modeling RF coils for UHF-MRI applications.
引用
收藏
页码:75 / 84
页数:10
相关论文
共 41 条
[31]   Design and Implementation of Split-Leg Type Elliptical Whole-Body Birdcage RF Coil at 1.5 T MRI [J].
Kumar, Suchit ;
Chung, Han-Jae ;
Jeong, You-Jin ;
Lee, Heung-Kyu ;
Oh, Chang-Hyun .
APPLIED SCIENCES-BASEL, 2021, 11 (16)
[32]   Versatile coil design and positioning of transverse-field RF surface coils for clinical 1.5-T MRI applications [J].
Alfonsetti, M ;
Clementi, V ;
Iotti, S ;
Placidi, G ;
Lodi, R ;
Barbiroli, B ;
Sotgiu, A ;
Alecei, M .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2005, 18 (02) :69-75
[33]   Integration of Electro-textile RF Coil Array with Magnetic Resonance Imaging (MRI) System: Design Strategies and Characterization Methods [J].
Zhang, Daisong ;
Rahmat-Samii, Yahya .
2018 IEEE INTERNATIONAL WORKSHOP ON ANTENNA TECHNOLOGY (IWAT), 2018,
[34]   Impact of MRI RF coil design on the RF-induced heating of medical implants: fixed B1+ rms exposure versus normal operating mode [J].
Yao, Aiping ;
Li, Zihan ;
Ma, Mingjuan .
PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (05)
[35]   Design and test of a double-nuclear RF coil for 1H MRI and 13C MRSI at 7 T [J].
Rutledge, Omar ;
Kwak, Tiffany ;
Cao, Peng ;
Zhang, Xiaoliang .
JOURNAL OF MAGNETIC RESONANCE, 2016, 267 :15-21
[36]   CFD-Guided Design of Non-Uniform Flow Channels in PEMFCs for Waste Heat Utilization in District Heating Networks [J].
Cui, Dai ;
Liu, Dong ;
Yu, Peng ;
Li, Jiayi ;
Zhou, Zhi ;
Zhang, Meishan ;
Chen, Qun ;
Yuan, Fang .
ENERGIES, 2025, 18 (08)
[37]   Design and application of a multimodality-compatible 1Tx/6Rx RF coil for monkey brain MRI at 7T [J].
Qu, Shuxian ;
Shi, Sunhang ;
Quan, Zhiyan ;
Gao, Yang ;
Wang, Minmin ;
Wang, Yueming ;
Pan, Gang ;
Lai, Hsin-Yi ;
Roe, Anna Wang ;
Zhang, Xiaotong .
NEUROIMAGE, 2023, 276
[38]   Optimum design of a horizontal shell-and-tube latent heat thermal energy storage system via non-uniform upper-and-lower cascade PCMs [J].
Zhang, Zhongbin ;
Zhu, Zhehao .
JOURNAL OF ENERGY STORAGE, 2024, 79
[39]   Entropy Generation and Thermal Radiation Impact on Magneto-Convective Flow of Heat-Generating Hybrid Nano-Liquid in a Non-Darcy Porous Medium with Non-Uniform Heat Flux [J].
Albqmi, Nora M. ;
Sivanandam, Sivasankaran .
COMPUTATION, 2024, 12 (03)
[40]   Marangoni convection with variable thermal conductivity and impact of inertial drag on radiative tri-hybrid nanofluid flow over a Riga plate with non-uniform heat emission/release [J].
Baithalu, Rupa ;
Mishra, S. R. ;
Panda, Subhajit .