Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams

被引:0
|
作者
Dreindl, Ralf [1 ,6 ]
Bolsa-Ferruz, Marta [1 ]
Fayos-Sola, Rosa [1 ,2 ]
Cabal, Fatima Padilla [1 ,3 ]
Scheuchenpflug, Lukas [1 ,4 ]
Elia, Alessio [1 ]
Amico, Antonio [1 ,5 ]
Carlino, Antonio [1 ]
Stock, Markus [1 ]
Grevillot, Loic [1 ]
机构
[1] MedAustron Ion Therapy Ctr, Wiener Neustadt, Austria
[2] Hosp Univ La Princesa, Dept Med Phys & Radiat Protect, Madrid, Spain
[3] Med Univ Vienna, AKH Wien, Dept Radiat Oncol, Div Med Radiat Phys, Vienna, Austria
[4] Univ Vienna, Fac Phys, Dept Isotope Phys, Vienna, Austria
[5] IRCCS, Med Phys Dept, Veneto Inst Oncol IOV, Padua, Italy
[6] MedAustron Ion Therapy Ctr, Marie Curie Str 5, A-2700 Wiener Neustadt, Austria
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2024年 / 25卷 / 05期
关键词
beam time reduction; clinical implementation; commissioning; independent dose calculation; pencil beam scanning; proton therapy; GAMMA-EVALUATION; THERAPY; VALIDATION; INDEX;
D O I
10.1002/acm2.14328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeExperimental patient-specific QA (PSQA) is a time and resource-intensive process, with a poor sensitivity in detecting errors. Radiation therapy facilities aim to substitute it by means of independent dose calculation (IDC) in combination with a comprehensive beam delivery QA program. This paper reports on the commissioning of the IDC software tool myQA iON (IBA Dosimetry) for proton therapy and its clinical implementation at the MedAustron Ion Therapy Center.MethodsThe IDC commissioning work included the validation of the beam model, the implementation and validation of clinical CT protocols, and the evaluation of patient treatment data. Dose difference maps, gamma index distributions, and pass rates (GPR) have been reviewed. The performance of the IDC tool has been assessed and clinical workflows, simulation settings, and GPR tolerances have been defined.ResultsBeam model validation showed agreement of ranges within +/- 0.2 mm, Bragg-Peak widths within +/- 0.1 mm, and spot sizes at various air gaps within +/- 5% compared to physical measurements. Simulated dose in 2D reference fields deviated by -0.3% +/- 0.5%, while 3D dose distributions differed by 1.8% on average to measurements. Validation of the CT calibration resulted in systematic differences of 2.0% between IDC and experimental data for tissue like samples. GPRs of 99.4 +/- 0.6% were found for head, head and neck, and pediatric CT protocols on a 2%/2 mm gamma criterion. GPRs for the adult abdomen protocol were at 98.9% on average with 3%/3 mm. Root causes of GPR outliers, for example, implants were identified and evaluated.ConclusionIDC has been successfully commissioned and integrated into the MedAustron clinical workflow for protons in 2021. IDC has been stepwise and safely substituting experimental PSQA since February 2021. The initial reduction of proton experimental PSQA was about 25% and reached up to 90% after 1 year.
引用
收藏
页数:16
相关论文
共 50 条
  • [41] FRoG-A New Calculation Engine for Clinical Investigations with Proton and Carbon Ion Beams at CNAO
    Choi, KyungDon
    Mein, Stewart B.
    Kopp, Benedikt
    Magro, Giuseppe
    Molinelli, Silvia
    Ciocca, Mario
    Mairani, Andrea
    CANCERS, 2018, 10 (11):
  • [42] Feasibility of patient specific quality assurance for proton therapy based on independent dose calculation and predicted outcomes
    Meijers, Arturs
    Marmitt, Gabriel Guterres
    Siang, Kelvin Ng Wei
    van der Schaaf, Arjen
    Knopf, Antje C.
    Langendijk, Johannes A.
    Both, Stefan
    RADIOTHERAPY AND ONCOLOGY, 2020, 150 : 136 - 141
  • [43] Generalized eMC implementation for Monte Carlo dose calculation of electron beams from different machine types
    Fix, Michael K.
    Cygler, Joanna
    Frei, Daniel
    Volken, Werner
    Neuenschwander, Hans
    Born, Ernst J.
    Manser, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (09) : 2841 - 2859
  • [44] Implementation of an improved dose-per-MU model for double-scattered proton beams to address interbeamline modulation width variability
    Lin, Liyong
    Shen, JiaJien
    Ainsley, Christopher G.
    Solberg, Timothy D.
    McDonough, James E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (03): : 297 - 306
  • [45] A modular dose delivery system for treating moving targets with scanned ion beams: Performance and safety characteristics, and preliminary tests
    Lis, Michelle
    Donetti, Marco
    Newhauser, Wayne
    Durante, Marco
    Dey, Joyoni
    Weber, Ulrich
    Wolf, Moritz
    Steinsberger, Timo
    Graeff, Christian
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 307 - 316
  • [46] Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system
    Naonori Hu
    Hiroki Tanaka
    Ryo Kakino
    Syuushi Yoshikawa
    Mamoru Miyao
    Kazuhiko Akita
    Kayako Isohashi
    Teruhito Aihara
    Keiji Nihei
    Koji Ono
    Radiation Oncology, 16
  • [47] Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system
    Hu, Naonori
    Tanaka, Hiroki
    Kakino, Ryo
    Yoshikawa, Syuushi
    Miyao, Mamoru
    Akita, Kazuhiko
    Isohashi, Kayako
    Aihara, Teruhito
    Nihei, Keiji
    Ono, Koji
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [48] Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization
    Gayou, Olivier
    Miften, Moyed
    MEDICAL PHYSICS, 2007, 34 (08) : 3183 - 3192
  • [49] Implementation of the validation testing in MPPG 5.a "Commissioning and QA of treatment planning dose calculations-megavoltage photon and electron beams"
    Jacqmin, Dustin J.
    Bredfeldt, Jeremy S.
    Frigo, Sean P.
    Smilowitz, Jennifer B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (01): : 115 - 127
  • [50] Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm
    Zhang, Aizhen
    Wen, Ning
    Nurushev, Teamour
    Burmeister, Jay
    Chetty, Indrin J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 127 - 145