A Novel Muscle Innervation Zone Estimation Method Using Monopolar High Density Surface Electromyography

被引:1
|
作者
Huang, Chengjun [1 ]
Chen, Maoqi [2 ]
Zhang, Yingchun [3 ]
Li, Sheng [4 ]
Klein, Cliff S. [5 ]
Zhou, Ping [2 ]
机构
[1] Baylor Coll Med, Dept Neurosci, Houston, TX 77030 USA
[2] Univ Hlth & Rehabil Sci, Sch Rehabil Sci & Engn, Qingdao 200027, Shandong, Peoples R China
[3] Univ Houston, Dept Biomed Engn, Houston, TX 77024 USA
[4] Univ Texas Hlth Sci Ctr Houston, TIRR Mem Hermann Hosp, Dept Phys Med & Rehabil, Houston, TX 77030 USA
[5] Guangdong Work Injury Rehabil Ctr, Guangzhou 510645, Guangdong, Peoples R China
关键词
High density surface electromyography (EMG); innervation zone (IZ); principal component analysis (PCA); the 2nd principal component; PRINCIPAL COMPONENT ANALYSIS; FIBER CONDUCTION-VELOCITY; MOTOR UNITS; EMG; POSITION; LOCALIZATION; ORGANIZATION; ACTIVATION; SIGNALS; MATRIX;
D O I
10.1109/TNSRE.2022.3215612
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study presents a novel method to estimate a muscle's innervation zone (IZ) location from monopolar high density surface electromyography (EMG) signals. Based on the fact that 2nd principal component coefficients derived from principal component analysis (PCA) are linearly related with the time delay of different channels, the channels located near the IZ should have the shortest time delays. Accordingly, we applied a novel method to estimate a muscle's IZ based on PCA. The performance of the developed method was evaluated by both simulation and experimental approaches. The method based on 2nd principal component of monopolar high density surface EMG achieved a comparable performance to cross-correlation analysis of bipolar signals when noise was simulated to be independently distributed across all channels. However, in simulated conditions of specific channel contamination, the PCA based method achieved superior performance than the cross-correlation method. Experimental high density surface EMG was recorded from the bicepsbrachii of 9 healthysubjects during maximum voluntary contractions. The PCA and cross-correlation based methods achieved high agreement, with a difference in IZ location of 0.47 +/- 0.4 IED (inter-electrode distance = 8 mm). The results indicate that analysis of 2nd principal component coefficients provides a useful approach for IZ estimation using monopolar high density surface EMG.
引用
收藏
页码:22 / 30
页数:9
相关论文
共 50 条
  • [21] Applying muscle synergy analysis to forearm high-density electromyography of healthy people
    Geng, Yanjuan
    Chen, Ziyin
    Zhao, Yang
    Cheung, Vincent C. K.
    Li, Guanglin
    FRONTIERS IN NEUROSCIENCE, 2022, 16
  • [22] One-Channel Surface Electromyography Decomposition for Muscle Force Estimation
    Sun, Wentao
    Zhu, Jinying
    Jiang, Yinlai
    Yokoi, Hiroshi
    Huang, Qiang
    FRONTIERS IN NEUROROBOTICS, 2018, 12
  • [23] Development of a Matlab-Based Graphical User Interface for Analysis of High-Density Surface Electromyography Signals
    Oliveira, I. S.
    Favretto, M. A.
    Cossul, S.
    Marques, J. L. B.
    XXVII BRAZILIAN CONGRESS ON BIOMEDICAL ENGINEERING, CBEB 2020, 2022, : 1829 - 1835
  • [24] Comparison of Facial Muscle Activation Patterns Between Healthy and Bell's Palsy Subjects Using High-Density Surface Electromyography
    Cui, Han
    Zhong, Weizheng
    Yang, Zhuoxin
    Cao, Xuemei
    Dai, Shuangyan
    Huang, Xingxian
    Hu, Liyu
    Lan, Kai
    Li, Guanglin
    Yu, Haibo
    FRONTIERS IN HUMAN NEUROSCIENCE, 2021, 14
  • [25] Innervation zone of the vastus medialis muscle: position and effect on surface EMG variables
    Gallina, A.
    Merletti, R.
    Gazzoni, M.
    PHYSIOLOGICAL MEASUREMENT, 2013, 34 (11) : 1411 - 1422
  • [26] NECK MOVEMENT CLASSIFICATION USING SURFACE ELECTROMYOGRAPHY: A NOVEL LINEAR ALGORITHM BASED ON MUSCLE COORDINATION
    Zhu, Yi
    Mahnan, Arash
    Konczak, Jurgen
    PROCEEDINGS OF THE 2021 DESIGN OF MEDICAL DEVICES CONFERENCE (DMD2021), 2021,
  • [27] Analysis of the suprahyoid muscles during tongue elevation: High-density surface electromyography as a novel tool for swallowing-related muscle assessment
    Yoshikawa, Kohei
    Nakamori, Masahiro
    Ushio, Kai
    Toko, Megumi
    Yamada, Hidetada
    Nishikawa, Yuichi
    Fukuoka, Tatsuyuki
    Maruyama, Hirofumi
    Mikami, Yukio
    JOURNAL OF ORAL REHABILITATION, 2024, 51 (09) : 1872 - 1880
  • [28] Advancing Muscle-Computer Interfaces with High-Density Electromyography
    Amma, Christoph
    Krings, Thomas
    Boer, Jonas
    Schultz, Tanja
    CHI 2015: PROCEEDINGS OF THE 33RD ANNUAL CHI CONFERENCE ON HUMAN FACTORS IN COMPUTING SYSTEMS, 2015, : 929 - 938
  • [29] High-Density surface Electromyography applications & reliability vs. muscle fatigue-a short review
    Sa-Ngiamsak, T.
    Castela Torres Costa, J.
    Santos Baptista, J.
    OCCUPATIONAL SAFETY AND HYGIENE II, 2014, : 265 - 269
  • [30] New insights into pain-related changes in muscle activation revealed by high-density surface electromyography
    Falla, Deborah
    Gallina, Alessio
    JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2020, 52