An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of carbamazepine in human serum and plasma

被引:1
作者
Schierscher, Tobias [2 ]
Salzmann, Linda [2 ]
Singh, Neeraj [3 ]
Bachmann, Martina [3 ]
Bauland, Friederike [4 ]
Geistanger, Andrea [3 ]
Risch, Lorenz [2 ]
Geletneky, Christian [3 ]
Seger, Christoph [2 ]
Taibon, Judith [1 ]
机构
[1] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
[2] Dr Risch Ostschweiz AG, Buchs, Switzerland
[3] Roche Diagnost GmbH, Penzberg, Germany
[4] Chrestos Concept GmbH & Co KG, Essen, Germany
关键词
carbamazepine; ID-LC-MS/MS; qNMR; reference measurement procedure; traceability; standardization; 10 ANTIEPILEPTIC DRUGS; NMR QNMR SPECTROSCOPY; ABSOLUTE CONTENT; CARBAMAZEPINE-10,11-EPOXIDE; MS; METABOLITES; SAMPLES;
D O I
10.1515/cclm-2023-0943
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: An isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) was developed and validated to accurately measure serum and plasma concentrations of carbamazepine. Methods: Quantitative nuclear magnetic resonance (qNMR) spectroscopy was used to determine the absolute content of the reference material, ensuring its traceability to SI units. The separation of carbamazepine from potential interferences, whether known or unknown, was achieved using a C18 column. A protein precipitation protocol followed by a high dilution step was established for sample preparation. Assay validation and determination of measurement uncertainty were performed in accordance with the guidelines of the Clinical and Laboratory Standards Institute, the International Conference on Harmonization (ICH), and the Guide to the Expression of Uncertainty in Measurement (GUM). In order to demonstrate equivalence to the already existing RMP a method comparison study was performed. Results: The RMP was proven to be highly selective and specific with no evidence of a matrix effect, allowing for quantification of carbamazepine within the range of 0.800-18.0 mu g/mL. Intermediate precision and repeatability (n=60 measurements) was found to be <1.6 % and <1.3 % over all concentration levels and independent from the matrix. The relative mean bias ranged from -0.1 to 0.6 % for native serum and from -0.3 to -0.1 % for Li-heparin plasma levels. The measurement uncertainties for single measurements and target value assignment were found to be <1.8 % and <1.3 %, respectively. Method comparison showed a good agreement between the Joint Committee of Traceability in Laboratory Medicine (JCTLM) listed RMP and the candidate RMP resulting in a Passing-Bablok regression equation with a slope of 1.01 and an intercept of -0.01. The bias in the patient cohort was found to be 0.9 %. Conclusions: We present a novel LC-MS/MS-based candidate RMP for carbamazepine in human serum and plasma which provides a traceable and reliable platform for the standardization of routine assays and evaluation of clinically relevant samples.
引用
收藏
页码:1301 / 1313
页数:13
相关论文
共 50 条
[41]   Quantification of estrogens in infant formulas by isotope dilution liquid chromatography-tandem mass spectrometry [J].
Liu, Shuan ;
Li, Xiaomin ;
Zhao, Yingchen ;
Yin, Xiong ;
Zhang, Qinghe ;
Li, Xiuqin ;
Tang, Guangshi ;
Li, Hongmei .
ANALYTICAL METHODS, 2018, 10 (32) :3968-3975
[42]   Isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedures for the quantification of 24(R),25-dihydroxyvitamin D2 and 24(R),25-dihydroxyvitamin D3 in human serum and plasma [J].
Kandler, Kerstin ;
Stadlmeier, Michael ;
Singh, Neeraj ;
Bauland, Friederike ;
Geistanger, Andrea ;
Geletneky, Christian ;
Taibon, Judith .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025,
[43]   Simultaneous quantification of 18 different phytocannabinoids in serum using a highly sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method [J].
Scheunemann, A. ;
Elsner, K. ;
Germerott, T. ;
Hess, C. ;
Roehrich, J. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2021, 1173
[44]   Development of an equilibrium dialysis ID-UPLC-MS/MS candidate reference measurement procedure for free thyroxine in human serum [J].
Ribera, Ashley ;
Zhang, Li ;
Dabbs-Brown, Amonae ;
Sugahara, Otoe ;
Poynter, Krista ;
van Uytfanghe, Katleen ;
Shimizu, Eri ;
Herwaarden, Antonius E. van ;
Botelho, Julianne C. ;
Danilenko, Uliana ;
Vesper, Hubert W. .
CLINICAL BIOCHEMISTRY, 2023, 116 :42-51
[45]   An isotope dilution LC-MS/MS based candidate reference method for the quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human whole blood [J].
Taibon, Judith ;
van Rooij, Milou ;
Schmid, Rupert ;
Singh, Neeraj ;
Albrecht, Eva ;
Wright, Jo Anne ;
Geletneky, Christian ;
Schuster, Carina ;
Moerlein, Sophie ;
Vogeser, Michael ;
Seger, Christoph ;
Pongratz, Stephan ;
Kobold, Uwe .
CLINICAL BIOCHEMISTRY, 2020, 82 :73-84
[46]   A reference measurement procedure for amino acids in blood using isotope dilution ultra-performance liquid chromatography-tandem mass spectrometry [J].
Kim, Juok ;
Tran, Thi Thanh Huong ;
Hong, Seon-Pyo ;
Jeong, Ji-Seon .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2017, 1055 :72-80
[47]   Measurement of serum progesterone by isotope dilution liquid chromatography tandem mass spectrometry: a candidate reference method and its application to evaluating immunoassays [J].
Yufei Wang ;
Tianjiao Zhang ;
Haijian Zhao ;
Weiyan Zhou ;
Jie Zeng ;
Jiangtao Zhang ;
Chuanbao Zhang ;
Wenxiang Chen .
Analytical and Bioanalytical Chemistry, 2019, 411 :2363-2371
[48]   Measurement of serum progesterone by isotope dilution liquid chromatography tandem mass spectrometry: a candidate reference method and its application to evaluating immunoassays [J].
Wang, Yufei ;
Zhang, Tianjiao ;
Zhao, Haijian ;
Zhou, Weiyan ;
Zeng, Jie ;
Zhang, Jiangtao ;
Zhang, Chuanbao ;
Chen, Wenxiang .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (11) :2363-2371
[49]   Quantification of cholesterol in human serum by isotope dilution liquid chromatography/mass spectrometry [J].
Shin, Hyesun ;
Lee, Hwashim ;
Lee, Gaeho .
ANALYTICAL SCIENCE AND TECHNOLOGY, 2008, 21 (06) :502-509
[50]   Evaluation of an Isotope Dilution HPLC Tandem Mass Spectrometry Candidate Reference Measurement Procedure for Total 17-β Estradiol in Human Serum [J].
Botelho, Julianne Cook ;
Ribera, Ashley ;
Cooper, Hans C. ;
Vesper, Hubert W. .
ANALYTICAL CHEMISTRY, 2016, 88 (22) :11123-11129