Reducing PTV margins for prostate SBRT with motion compensation and gating techniques

被引:6
作者
Goddard, Lee [1 ,2 ]
Jeong, Kyoungkeun [1 ,2 ]
Tang, Justin [1 ,2 ]
Garg, Madhur [1 ,2 ]
Tome, Wolfgang A. A. [1 ,2 ,3 ]
机构
[1] Montefiore Med Ctr, Dept Radiat Oncol, Bronx, NY 10461 USA
[2] Albert Einstein Coll Med, 1300 Morris Pk Ave,Block Bldg Room 106, Bronx, NY 10461 USA
[3] Montefiore Med Ctr, Div Med Phys, 1300 Morris Pk Ave,Block Bldg Room 106, Bronx, NY 10461 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2023年 / 24卷 / 04期
关键词
Prostate SBRT intrafraction motion; STEREOTACTIC BODY RADIOTHERAPY; DOSIMETRIC IMPACT; INTRAFRACTION MOTION; ROTATIONS;
D O I
10.1002/acm2.13861
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate the dosimetric accuracy of prostate SBRT when motion is considered. To account for target movement, motion compensation and gating techniques were investigated with PTV margins reduced to 2 mm. To allow for dosimetric measurements a Delta4 phantom, Gafchromic film, and Hexamotion motion platform were utilized. Four motion files were utilized that represent a range of motions. Analysis of measured prostate motions for fifteen patients was performed to ensure detected motions were similar to those previously reported and motion files utilized were suitable. Five patient plans were utilized to allow for the effects of MLC and target motion interplay to be investigated. For both motion compensation and gating techniques, plans were delivered to the stationary phantom and for each of four motion types with/without compensation/gating enabled. Using a 3%, 2 mm and 80% threshold gamma criteria, film measurements had an average pass rate of 80.5% for uncorrected deliveries versus 96.0% for motion compensated deliveries. For gated techniques average pass rates increased from 89.9% for uncorrected to 94.8% with gating enabled. Measurements with the Delta4 arrays were analyzed with a 3%, 2 mm and 10% threshold dose. An average pass rate of 83.8% was measured for uncorrected motions versus 94.8% with motion compensation. For the gated technique an average pass rate of 87.2% was found for uncorrected motions versus 96.9% with gating enabled. These results show that very high gamma pass rates are achievable when motion compensation or gating techniques are applied. When target motion is not accounted for shifts up to 5 mm in planned versus delivered isodose distributions were found. However, when motion compensation, or gated techniques were applied, much smaller differences between planned and delivered isodose distributions were found. With these techniques dose delivery accuracy is greatly improved, allowing for PTV margins to be reduced.
引用
收藏
页数:9
相关论文
共 31 条
  • [1] Accuray, 2022, SYNCHR REAL TIM DEL
  • [2] Amro H, 2014, MED PHYS, V41, P217, DOI 10.1118/1.4888311
  • [3] [Anonymous], 2017, NCCN CLIN PRACTICE G
  • [4] [Anonymous], 2022, PLUS INDEX PHP S IHZ
  • [5] Measurements of intrafraction motion and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers
    Aubry, JF
    Beaulieu, L
    Girouard, LM
    Aubin, S
    Tremblay, D
    Laverdière, J
    Vigneault, E
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : 30 - 39
  • [6] Brenner DJ, 2002, INT J RADIAT ONCOL, V52, P6
  • [7] Evaluating the potential benefit of reduced planning target volume margins for low and intermediate risk patients with prostate cancer using real-time electromagnetic tracking
    Chaurasia, Avinash R.
    Sun, Kelly J.
    Premo, Christopher
    Brand, Timothy
    Tinnel, Brent
    Barczak, Stacie
    Halligan, John
    Brown, Michael
    Macdonald, Dusten
    [J]. ADVANCES IN RADIATION ONCOLOGY, 2018, 3 (04) : 630 - 638
  • [8] Assessing the Dosimetric Impact of Real-Time Prostate Motion During Volumetric Modulated Arc Therapy
    Diego Azcona, Juan
    Xing, Lei
    Chen, Xin
    Bush, Karl
    Li, Ruijiang
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 88 (05): : 1167 - 1174
  • [9] The radiobiology of prostate cancer including new aspects of fractionatedradiotherapy
    Fowler, JF
    [J]. ACTA ONCOLOGICA, 2005, 44 (03) : 265 - 276
  • [10] Commissioning and routine quality assurance of the Radixact Synchrony system
    Goddard, Lee
    Jeong, Kyoungkeun
    Tome, Wolfgang A.
    [J]. MEDICAL PHYSICS, 2022, 49 (02) : 1181 - 1195