Stability Study of an Interventional Surgery Robot Based on Active Disturbance Rejection Control

被引:2
作者
Ma, Xu [1 ]
Wen, Quan [1 ]
机构
[1] Tianjin Univ Technol, Sch Elect Engn & Automat, Tianjin Key Lab New Energy Power Convers Transmiss, Tianjin 300384, Peoples R China
关键词
vascular interventional surgery; surgery training; bio-inspired control; human-machine interaction; active disturbance rejection control; fuzzy reasoning; artificial neural networks; SYSTEM; DESIGN;
D O I
10.3390/electronics12092115
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Interventional surgery robots are essential in cardiovascular surgery as remote medical devices. By performing remote surgery, surgeons can reduce surgical fatigue and after-effects from heavy surgical instruments and radiation, ensuring that cardiovascular surgery is performed in a safe and reliable manner. To enhance stability during interventional procedures and reduce the impact of surgical risk due to factors where the robotic guidewire section from the end is vulnerable to mechanical jitter or blockage by blood flow, lipids, and thrombus inside the vessel, a new control method is proposed. The active disturbance rejection controller (ADRC) combined with intelligence algorithms is used to improve the performance of the controller by introducing the fuzzy inference algorithm and RBF neural network algorithm to self-adjust the parameters of the controller so that it has a greater ability to compensate for the disturbance factors appearing in the system. In numerical simulation experiments, the advantages and disadvantages of the ADRC combined with intelligence algorithms and the control performance of the conventional control strategy are analyzed in terms of the following: disturbance suppression performance and flexibility performance, respectively. Finally, different types of working conditions have been designed in the experimental platform to simulate the operation flow of in vivo vascular surgery. Experimental results show that the controller proposed in this paper meets the high accuracy, fast response, and low deviation required by interventional vascular surgery robots in complex surgical environments and can provide a more reliable guarantee for the stability of interventional surgery robots.
引用
收藏
页数:18
相关论文
共 33 条
[1]  
Au Stephanie, 2014, Asian Cardiovasc Thorac Ann, V22, P110, DOI 10.1177/0218492313484736
[2]   Operation evaluation in-human of a novel remote-controlled vascular interventional robot [J].
Bao, Xianqiang ;
Guo, Shuxiang ;
Xiao, Nan ;
Li, Youxiang ;
Yang, Cheng ;
Shen, Rui ;
Cui, Jinxin ;
Jiang, Yuhua ;
Liu, Xinke ;
Liu, Keyun .
BIOMEDICAL MICRODEVICES, 2018, 20 (02)
[3]  
Duan H.D., 2012, THESIS JILIN U JILIN
[4]   On the foundation of active disturbance rejection control [J].
Gao, Z.-Q. (z.gao@ieee.org), 1600, South China University of Technology (30)
[5]  
Gao ZQ, 2003, P AMER CONTR CONF, P4989
[6]   Design and performance evaluation of a novel robotic catheter system for vascular interventional surgery [J].
Guo, Jian ;
Guo, Shuxiang ;
Shao, Lin ;
Wang, Peng ;
Gao, Qiang .
MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2016, 22 (09) :2167-2176
[7]   Design and characteristics evaluation of a novel teleoperated robotic catheterization system with force feedback for vascular interventional surgery [J].
Guo, Jian ;
Guo, Shuxiang ;
Yu, Yang .
BIOMEDICAL MICRODEVICES, 2016, 18 (05)
[8]   Study on real-time force feedback for a master-slave interventional surgical robotic system [J].
Guo, Shuxiang ;
Wang, Yuan ;
Xiao, Nan ;
Li, Youxiang ;
Jiang, Yuhua .
BIOMEDICAL MICRODEVICES, 2018, 20 (02)
[9]  
Han J., 1998, Control and Decision, V13, P19, DOI DOI 10.13195/J.CD.1998.01.19.HANJQ.004
[10]  
Han J., 2008, Active disturbance rejection control: a technique for estimating and compensating the uncertainties