Optimization of Electrolytes with Redox Reagents to Improve the Impedimetric Signal for Use with a Low-Cost Analyzer

被引:0
作者
Cheng, Yu-Hsuan [1 ]
Chande, Charmi [1 ]
Li, Zhenglong [1 ]
Haridas Menon, Niranjan [1 ]
Kaaliveetil, Sreerag [1 ]
Basuray, Sagnik [1 ]
机构
[1] New Jersey Inst Technol, Otto H York Dept Chem & Mat Engn, 323 Dr Martin Luther King Jr Blvd, Newark, NJ 07102 USA
来源
BIOSENSORS-BASEL | 2023年 / 13卷 / 12期
关键词
impededance spectroscopy; electrochemical detection; redox probes; low-cost analyzer; IMPEDANCE SPECTROSCOPY; BIOSENSOR; PROTEIN; SENSITIVITY; PLATFORM; SENSORS; LABEL;
D O I
10.3390/bios13120999
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The most well-known criterion for POC devices is ASSURED, and affordability, i.e., using low-cost instrumentation, is the most challenging one. This manuscript provides a pathway for transitioning ESSENCE, an impedance-based biosensor platform, from using an expensive benchtop analyzer-KeySight 4294A (similar to$50k)-to using a significantly portable and cheaper USB oscilloscope-Analog Discovery 2 (similar to$200)-with similar sensitivity (around 100 times price difference). To achieve this, we carried out a fundamental study of the interplay between an electrolyte like potassium chloride (KCl), and an electrolyte buffer like phosphate buffered saline (PBS) in the presence and absence of a redox buffer like ferro/ferricyanide system and ([Ru(bpy)3](2+)). Redox molecules in the electrolyte caused a significant change in the Nyquist curve of the impedance depending on the redox molecule type. The redox species and the background electrolyte have their own RC semicircles in the Nyquist curve, whose overlap depends on the redox concentration and electrolyte ionic strength. We found that by increasing the electrolyte ionic strength or the redox concentration, the RC semicircle moves to higher frequencies and vice versa. Importantly, the use of the buffer electrolyte, instead of KCl, led to a lower standard deviation and overall signal (lesser sensitivity). However, to achieve the best results from the biorecognition signal, we chose a buffered electrolyte like PBS with high ionic strength and lowered the redox probe concentrations to minimize the standard deviation and reduce any noise from migrating to the low-cost analyzer. Comparing the two analyzers shows similar results, with a lowered detection limit from the low-cost analyzer.
引用
收藏
页数:14
相关论文
共 44 条
  • [1] Developments in biosensors for CoV detection and future trends
    Antiochia, Riccarda
    [J]. BIOSENSORS & BIOELECTRONICS, 2021, 173
  • [2] The electrochemical impedance spectrum of asymmetric electrolytes across low to moderate frequencies
    Balu, Bhavya
    Khair, Aditya S. S.
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2022, 911
  • [3] Generation of polystyrene-specific antibodies for developing immunoassays to analyze microplastics and nanoplastics
    Cao, Liu
    Zeng, Ximin
    Lin, Jun
    [J]. CHEMICAL ENGINEERING JOURNAL, 2023, 465
  • [4] Electrochemical biosensors for pathogen detection
    Cesewski, Ellen
    Johnson, Blake N.
    [J]. BIOSENSORS & BIOELECTRONICS, 2020, 159
  • [5] Improving the sensitivity of electrochemical sensors through a complementary luminescent mode: A new spectroelectrochemical approach
    Chatterjee, Sayandev
    Fujimoto, Meghan S.
    Cheng, Yu Hsuan
    Kargupta, Roli
    Soltis, Jennifer A.
    Motkuri, Radha Kishan
    Basuray, Sagnik
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2019, 284 : 663 - 674
  • [6] Metal-Organic Framework-Based Microfluidic Impedance Sensor Platform for Ultrasensitive Detection of Perfluorooctanesulfonate
    Cheng, Yu H.
    Barpaga, Dushyant
    Soltis, Jennifer A.
    Shutthanandan, V.
    Kargupta, Roli
    Han, Kee Sung
    McGrail, B. Peter
    Motkuri, Radha Kishan
    Basuray, Sagnik
    Chatterjee, Sayandev
    [J]. ACS APPLIED MATERIALS & INTERFACES, 2020, 12 (09) : 10503 - 10514
  • [7] ESSENCE-A rapid, shear-enhanced, flow-through, capacitive electrochemical platform for rapid detection of biomolecules
    Cheng, Yu-Hsuan
    Kargupta, Roli
    Ghoshal, Debjit
    Li, Zhenglong
    Chande, Charmi
    Feng, Lixin
    Chatterjee, Sayandev
    Koratkar, Nikhil
    Motkuri, Radha Kishan
    Basuray, Sagnik
    [J]. BIOSENSORS & BIOELECTRONICS, 2021, 182 (182)
  • [8] Effect of electrode configuration on the sensitivity of nucleic acid detection in a non-planar, flow-through, porous interdigitated electrode
    Cheng, Yu-Hsuan
    Moura, Pedro Antonio Reis
    Li Zhenglong
    Feng, Lixin
    Arokiam, Siril
    Yang, Juliana
    Hariharan, Mahima
    Basuray, Sagnik
    [J]. BIOMICROFLUIDICS, 2019, 13 (06):
  • [9] Continuously perfusable, customisable, and matrix-free vasculature on a chip platform
    Chesnais, Francois
    Joel, Jordan
    Hue, Jonas
    Shakib, Sima
    Di Silvio, Lucy
    Grigoriadis, Agamemnon E.
    Coward, Trevor
    Veschini, Lorenzo
    [J]. LAB ON A CHIP, 2023, 23 (04) : 761 - 772
  • [10] A dual-responsive electrochemical biosensor based on artificial protein imprinted polymers and natural hyaluronic acid for sensitive recognition towards biomarker CD44
    Cui, Min
    Sun, Xiaofeng
    Liu, Rui
    Du, Meigui
    Song, Xiaoming
    Wang, Shuai
    Hu, Wei
    Luo, Xiliang
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2022, 371