Electromagnetic and RF pulse design simulation based optimization of an eight-channel loop array for 11.7T brain imaging

被引:8
作者
Chu, Son [1 ]
Gras, Vincent [2 ]
Mauconduit, Franck [2 ]
Massire, Aurelien [3 ]
Boulant, Nicolas [2 ]
Gunamony, Shajan [1 ,4 ]
机构
[1] Univ Glasgow, Imaging Ctr Excellence, Glasgow City, Scotland
[2] Univ Paris Saclay, CEA, CNRS, BAOBAB,NeuroSpin, Gif Sur Yvette, France
[3] Siemens Healthcare SAS, St Denis, France
[4] Univ Glasgow, Imaging Ctr Excellence, Glasgow City G51 4LB, Scotland
基金
欧盟地平线“2020”;
关键词
11.7T MRI; parallel transmit; transmit array design; UHF MRI; TO-NOISE RATIO; PARALLEL TRANSMISSION; ABSORPTION RATE; MRI; POWER; SAFETY; INHOMOGENEITY; HOMOGENEITY; CIRCUIT; SAR;
D O I
10.1002/mrm.29654
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Optimization of transmit array performance is crucial in ultra-high-field MRI scanners such as 11.7T because of the increased RF losses and RF nonuniformity. This work presents a new workflow to investigate and minimize RF coil losses, and to choose the optimum coil configuration for imaging.Methods: An 8-channel transceiver loop-array was simulated to analyze its loss mechanism at 499.415 MHz. A folded-end RF shield was developed to limit radiation loss and improve the B-1(+) efficiency. The coil element length, and the shield diameter and length were further optimized using electromagnetic (EM) simulations. The generated EM fields were used to perform RF pulse design (RFPD) simulations under realistic constraints. The chosen coil design was constructed to demonstrate performance equivalence in bench and scanner measurements.Results: The use of conventional RF shields at 11.7T resulted in significantly high radiation losses of 18.4%. Folding the ends of the RF shield combined with optimizing the shield diameter and length increased the absorbed power in bio-logical tissue and reduced the radiation loss to 2.4%. The peak B-1(+) of the optimal array was 42% more than the reference array. Phantom measurements validated the numerical simulations with a close match of within 4% of the predicted B-1(+) .Conclusion: A workflow that combines EM and RFPD simulations to numerically optimize transmit arrays was developed. Results have been validated using phantom measurements. Our findings demonstrate the need for optimizing the RF shield in conjunction with array element design to achieve efficient excitation at 11.7T.
引用
收藏
页码:770 / 783
页数:14
相关论文
共 50 条
[1]  
Allard J., 2022, 31 ANN M ISMRM LOND
[2]   Bent folded-end dipole head array for ultrahigh-field MRI turns "dielectric resonance" from an enemy to a friend [J].
Avdievich, Nikolai, I ;
Solomakha, Georgiy ;
Ruhm, Loreen ;
Bause, Jonas ;
Scheffler, Klaus ;
Henning, Anke .
MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (06) :3453-3467
[3]   Double-row 18-loop transceive 32-loop receive tight-fit array provides for whole-brain coverage, high transmit performance, and SNR improvement near the brain center at 9.4T [J].
Avdievich, Nikolai I. ;
Giapitzakis, Ioannis-Angelos ;
Bause, Jonas ;
Shajan, Gunamony ;
Scheffler, Klaus ;
Henning, Anke .
MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (05) :3392-3405
[4]  
Barberi EA, 2000, MAGNET RESON MED, V43, P284, DOI 10.1002/(SICI)1522-2594(200002)43:2<284::AID-MRM16>3.0.CO
[5]  
2-C
[6]   Strongly modulating pulses for counteracting RF inhomogeneity at high fields [J].
Boulant, N. ;
Le Bihan, D. ;
Amadon, A. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (03) :701-708
[7]   Counteracting Radio Frequency Inhomogeneity in the Human Brain at 7 Tesla Using Strongly Modulating Pulses [J].
Boulant, N. ;
Mangin, J. -F. ;
Amadon, A. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (05) :1165-1172
[8]   B1+ Interferometry for the Calibration of RF Transmitter Arrays [J].
Brunner, David O. ;
Pruessmann, Klaas P. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (06) :1480-1488
[9]   MRI and MRS of the human brain at magnetic fields of 14 T to 20 T: Technical feasibility, safety, and neuroscience horizons [J].
Budinger, Thomas F. ;
Bird, Mark D. .
NEUROIMAGE, 2018, 168 :509-531
[10]   Numerical Evaluation of Image Homogeneity, Signal-to-Noise Ratio, and Specific Absorption Rate for Human Brain Imaging at 1.5, 3, 7, 10.5, and 14T in an 8-Channel Transmit/Receive Array [J].
Cao, Zhipeng ;
Park, Joshua ;
Cho, Zang-Hee ;
Collins, Christopher M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (05) :1432-1439