Ultrasonic needle hydrophone calibration in air by a parabolic off-axis mirror focused beam using three-transducer reciprocity

被引:6
作者
Svilainis, Linas [1 ]
Chaziachmetovas, Andrius [1 ]
Kaskonas, Paulius [1 ]
Alvarez-Arenas, Tomas E. Gomez [2 ]
机构
[1] Kaunas Univ Technol, Elect Engn Dept, LT-51368 Kaunas, Lithuania
[2] CSIC, Ultrason & Sensors Technol Dept, Madrid, Spain
关键词
Acoustic metrology; Air -coupled ultrasound; Probe characterization; Plane wave reciprocity; Sensitivity calibration; PIEZOELECTRIC TRANSDUCERS; PROFILE MEASUREMENTS; COUPLED TRANSDUCER; FIELD; SIMULATIONS; RADIATION;
D O I
10.1016/j.ultras.2023.107025
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An acoustic field distribution investigation in air requires a small receiving sensor. Needle hydrophones seem to be an attractive solution, and it has previously been demonstrated that needle hydrophones designed for use in water can be used in air. The metrology problem is that an absolute sensitivity calibration is needed, because needle hydrophones are not characterized in air, especially for frequencies below 1 MHz, which is of interest for air-coupled ultrasound. Conventional, three-transducer/microphone reciprocity calibration requires measure-ments to be done in the far field. However, when transducer diameter is large and the frequency is high, the required measurement distance becomes very large: 3 m for a 20 mm source, transmitting at 1 MHz. Large propagation distance leads to high attenuation and nonlinear effects in air propagation, and distortion and losses accumulate. Small needle hydrophones have low sensitivity, so that high excitation amplitudes would be required, which can lead to transducer heating and increase nonlinearity effects. A derivative of the three -transducer reciprocity calibration method is proposed, where a large aperture transducer is focused onto a hy-drophone, using hybrid of plane wave and spherical wave reciprocity. Use of a focused source minimizes the frequency-dependent diffraction effects, and the spherical wave approximation is valid at the focal distance, and low level excitation signals can be used. Focusing is accomplished using a parabolic off-axis mirror. Calibration is in transmission, which reduces the complexity of the electrical measurements. The corresponding equations have been derived for this setup. Calibration of the transducer and needle hydrophone absolute sensitivity is obtained.
引用
收藏
页数:11
相关论文
共 44 条
  • [1] 0 mm needle hydrophone (NH1000), TECHNICAL DATA SHEET
  • [2] 5 mm needle hydrophone (NH1000), TECHN DAT SHEET PREC
  • [3] Real-Time 3-D Imaging Using an Air-Coupled Ultrasonic Phased-Array
    Allevato, Gianni
    Hinrichs, Jan
    Rutsch, Matthias
    Adler, Jan Philipp
    Jager, Axel
    Pesavento, Marius
    Kupnik, Mario
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (03) : 796 - 806
  • [4] Acoustic impedance matching of piezoelectric transducers to the air
    Alvarez-Arenas, TEG
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (05) : 624 - 633
  • [5] [Anonymous], 2020, 6212732022 IEC
  • [6] [Anonymous], 2021, 1 8 INCH PRESSURE FI
  • [7] [Anonymous], 2020, IEC 60565-1:2020
  • [8] [Anonymous], 2020, 6109432016 IEC
  • [9] [Anonymous], 2010, Standard EN 12668-2:2010
  • [10] Beam profile measurements and simulations for ultrasonic transducers operating in air
    Benny, G
    Hayward, G
    Chapman, R
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 107 (04) : 2089 - 2100