Assessment of Magnetic Resonance Imaging Artefacts Caused by Equine Anaesthesia Equipment: A Cadaver Study

被引:0
作者
Testa, Barbara [1 ,2 ]
Biggi, Marianna [3 ]
Byrne, Christian A. [1 ,4 ]
Bell, Andrew [1 ]
机构
[1] Univ Glasgow, One Hlth & Vet Med, Sch Biodivers, Glasgow, Scotland
[2] Univ Edinburgh, Royal Dick Sch Vet Studies, Roslin, Scotland
[3] Registered Off St Johns Innovat Ctr, Vet CT, Cambridge, England
[4] Hird & Partners Shelf Equine Hosp, Lower Giles Hill Farm, Halifax, England
关键词
Equine; Anesthesia; Low -field MRI; Artifacts; Anesthetic equipment; Tafonius; ALVEOLAR RECRUITMENT MANEUVER; ENHANCED COMPUTED-TOMOGRAPHY; POSITIVE AIRWAY PRESSURE; END-EXPIRATORY PRESSURE; CONFIDENTIAL INQUIRY; HIGH-FIELD; HORSES; RADIOGRAPHY; VENTILATION; DELIVERY;
D O I
10.1016/j.jevs.2023.104492
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Acquisition of magnetic resonance images of the equine limb is still sometimes conducted under general anaesthesia. Despite low-field systems allow the use of standard anaesthetic equipment, possible interferences of the extensive electronic componentry of advanced anaesthetic machines on image quality is unknown. This prospective, blinded, cadaver study investigated the effects of seven standardised conditions (Tafonius positioned as in clinical cases, Tafonius on the boundaries of the controlled area, anaesthetic monitoring only, Mallard anaesthetic machine, Bird ventilator, complete electronic silence in the room (negative control), source of electronic interference [positive control]) on image quality through the acquisition of 78 sequences using a 0.31T equine MRI scanner. Images were graded with a 4-point scoring system, where 1 denoted absence of artefacts and 4 major artefacts requiring repetition in a clinical setting. A lack of STIR fat suppression was commonly reported (16/26). Ordinal logistic regression showed no statistically significant differences in image quality between the negative control and either the non-Tafonius or the Tafonius groups ( P = 0.535 and P = 0.881, respectively), and with the use of Tafonius compared to the other anaesthetic machines ( P = 0.578). The only statistically significant differences in scores were observed between the positive control and the non-Tafonius ( P = 0.006) and the Tafonius groups ( P = 0.017). Our findings suggest that anaesthetic machines and monitoring do not appear to affect MRI scan quality and support the use of Tafonius during acquisition of images with a 0.31T MRI system in a clinical context.
引用
收藏
页数:9
相关论文
共 39 条
[1]   Effects of positive end-expiratory pressure titration on gas exchange, respiratory mechanics and hemodynamics in anesthetized horses [J].
Ambrosio, Aline M. ;
Ida, Keila K. ;
Souto, Maria T. M. R. ;
Oshiro, Alexandre H. ;
Fantoni, Denise T. .
VETERINARY ANAESTHESIA AND ANALGESIA, 2013, 40 (06) :564-572
[2]   Effects of 12 and 17 cmH2O positive end-expiratory pressure applied after alveolar recruitment maneuver on pulmonary gas exchange and compliance in isoflurane-anesthetized horses [J].
Andrade, Felipe S. R. M. ;
Faco, Lara L. ;
Ida, Keila K. ;
Silva, Luis C. L. C. ;
Fantoni, Denise T. ;
Ambrosio, Aline M. .
VETERINARY ANAESTHESIA AND ANALGESIA, 2019, 46 (01) :64-73
[3]   Review of hypoxaemia in anaesthetized horses: predisposing factors, consequences and management [J].
Auckburally, Adam ;
Nyman, Gorel .
VETERINARY ANAESTHESIA AND ANALGESIA, 2017, 44 (03) :397-408
[4]   A review of how magnetic resonance imaging can aid in case management of common pathological conditions of the equine foot [J].
Barrett, M. F. ;
Frisbie, D. D. ;
King, M. R. ;
Werpy, N. M. ;
Kawcak, C. E. .
EQUINE VETERINARY EDUCATION, 2017, 29 (12) :683-693
[5]   Qualitative Comparison of 0.27T, 1.5T, and 3T Magnetic Resonance Images of the Normal Equine Foot [J].
Bolen, Geraldine ;
Audigie, Fabrice ;
Spriet, Mathieu ;
Vandenberghe, Filip ;
Busoni, Valeria .
JOURNAL OF EQUINE VETERINARY SCIENCE, 2010, 30 (01) :9-20
[6]   The risk of death: the Confidential Enquiry into Perioperative Small Animal Fatalities [J].
Brodbelt, David C. ;
Blissitt, Karen J. ;
Hammond, Richard A. ;
Neath, Prue J. ;
Young, Lestey E. ;
Pfeiffer, Dirk U. ;
Wood, James L. N. .
VETERINARY ANAESTHESIA AND ANALGESIA, 2008, 35 (05) :365-373
[7]   Clinical magnetic resonance image quality of the equine foot is significantly influenced by acquisition system [J].
Byrne, Christian A. ;
Marshall, John F. ;
Voute, Lance C. .
EQUINE VETERINARY JOURNAL, 2021, 53 (03) :469-480
[8]   VALIDATION OF DELAYED GADOLINIUM-ENHANCED MAGNETIC RESONANCE IMAGING OF CARTILAGE AND T2 MAPPING FOR QUANTIFYING DISTAL METACARPUS/METATARSUS CARTILAGE THICKNESS IN THOROUGHBRED RACEHORSES [J].
Carstens, Ann ;
Kirberger, Robert M. ;
Dahlberg, Leif E. ;
Prozesky, Leon ;
Fletcher, Lizelle ;
Lammentausta, Eveliina .
VETERINARY RADIOLOGY & ULTRASOUND, 2013, 54 (02) :139-148
[9]  
Floriano D, 2020, AM J VET RES, V81, P857, DOI 10.2460/ajvr.81.11.857
[10]   Post anaesthetic myopathy/neuropathy in horses undergoing magnetic resonance imaging compared to horses undergoing surgery [J].
Franci, R. ;
Leece, E. A. ;
Brearley, J. C. .
EQUINE VETERINARY JOURNAL, 2006, 38 (06) :497-501