Creating uniform cluster dose spread-out Bragg peaks for proton and carbon beams

被引:0
作者
Ortiz, Ramon [1 ,2 ]
Faddegon, Bruce [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA USA
[2] 1600 Divisadero St, San Francisco, CA 94115 USA
基金
美国国家卫生研究院;
关键词
cluster dose; ionization detail; Monte Carlo; nanodosimetry; particle therapy; TRACK STRUCTURE; NANODOSIMETRY; THERAPY; TOPAS; IONS;
D O I
10.1002/mp.16991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundPreliminary data have shown a close association of the generalized ionization cluster size dose (in short, cluster dose) with cell survival, independent of particle type, and energy, when cluster dose is derived from an ionization detail parameter preferred for its association with cell survival. Such results suggest cluster dose has the potential to replace RBE-weighted dose in proton and ion beam radiotherapy treatment plan optimization, should a uniform cluster dose lead to comparable biological effects. However, further preclinical investigations are warranted to confirm this premise.PurposeTo present an analytical approach to create uniform cluster dose spread-out Bragg peaks (SOBP) for evaluation of the potential of cluster dose to result in uniform biological effect.MethodsWe modified the coefficients of the Bortfeld and Schlegel weight formula, an analytical method typically used for the creation of radiation dose SOBP in particle therapy, to produce uniform cluster dose SOBP of different widths (1-5 cm) at relevant clinical proton and carbon beam energies. Optimum parameters were found by minimization of the ratio between the maximum and minimum cluster dose in the SOBP region using the Nelder-Mead method.ResultsThe coefficients of the Bortfeld and Schlegel weight formula leading to uniform cluster dose SOBPs were determined for each combination of beam energy and SOBP width studied. The uniformity of the resulting cluster dose SOBPs, calculated as the relative difference between the maximum and minimum cluster dose within the SOBP, was within 0.3%-3.5% for the evaluated proton beams and 1.3%-3.4% for the evaluated carbon beams.ConclusionsThe modifications to the analytical approach to create radiation dose SOBPs resulted in uniform cluster dose proton and carbon SOBPs over a wide range of beam energies and SOBP widths. Such SOBPs should prove valuable in preclinical investigations for the selection of nanodosimetric quantities to be used in proton and ion therapy treatment planning.
引用
收藏
页码:4482 / 4488
页数:7
相关论文
共 41 条
  • [41] Determination of ion recombination and polarity effect correction factors for a plane-parallel ionization Bragg peak chamber under proton and carbon ion pencil beams
    Mirandola, A.
    Magro, G.
    Maestri, D.
    Mairani, A.
    Mastella, E.
    Molinelli, S.
    Russo, S.
    Avai, A.
    Ciocca, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (09)