Simultaneous imaging of luminescence and prompt x-rays during irradiation with spot-scanning proton beams at clinical dose level

被引:1
作者
Yamamoto, Seiichi [1 ]
Yamashita, Tomohiro [2 ]
Kobashi, Yusuke [2 ]
Yabe, Takuya [3 ]
Akagi, Takashi [4 ]
Yamaguchi, Mitsutaka [3 ]
Kawachi, Naoki [3 ]
Kamada, Kei [5 ]
Yoshikawa, Akira [5 ]
Kataoka, Jun [1 ]
机构
[1] Waseda Univ, Fac Sci & Engn, Tokyo, Japan
[2] Kobe Proton Ctr, Kobe, Japan
[3] Natl Inst Quantum Sci & Technol QST, Takasaki Inst Adv Quantum Sci, Fdn Quantum Technol Res Directorate, Chiba, Japan
[4] Hyogo Ion Beam Med Ctr, Tatsuno, Japan
[5] Tohoku Univ, New Ind Creat Hatchery Ctr NICHe, Sendai, Japan
关键词
luminescence of water; prompt x-rays; simultaneous imaging; proton beams; POSITRON-EMISSION-TOMOGRAPHY; RANGE VERIFICATION; PET SYSTEM; GAMMA-RAYS; CAMERA; DELIVERY; WATER;
D O I
10.1088/2057-1976/acdf61
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Prompt x-ray imaging is a promising method for observing the beam shape from outside a subject. However, its distribution is different from dose distribution, and thus a comparison with the dose is required. Meanwhile, luminescence imaging of water is a possible method for imaging the dose distribution. Consequently, we performed simultaneous imaging of luminescence and prompt x-rays during irradiation by proton beams to compare the distributions between these two different imaging methods. Optical imaging of water was conducted with spot-scanning proton beams at clinical dose level during irradiation to a fluorescein (FS) water phantom set in a black box. Prompt x-ray imaging was also conducted simultaneously from outside the black box using a developed x-ray camera during proton beam irradiation to the phantom. We measured images of the luminescence of FS water and prompt x-rays for various types of proton beams, including pencil beams, spread-out Bragg peak (SOBP) beams, and clinically used therapy beams. After the imaging, ranges were estimated from FS water and prompt x-rays and compared with those calculated with a treatment planning system (TPS). We could measure the prompt x-ray and FS water images simultaneously for all types of proton beams. The ranges estimated from the FS water and those calculated with the TPS closely matched, within a difference of several mm. Similar range difference was found between the results estimated from prompt x-ray images and those calculated with the TPS. We confirmed that the simultaneous imaging of luminescence and prompt x-rays were possible during irradiation with spot-scanning proton beams at a clinical dose level. This method can be applied to range estimation as well as comparison with the dose for prompt x-ray imaging or other imaging methods used in therapy with various types of proton beams at a clinical dose level.
引用
收藏
页数:11
相关论文
共 44 条
[1]   Development of a low-energy x-ray camera for the imaging of secondary electron bremsstrahlung x-ray emitted during proton irradiation for range estimation [J].
Ando, Koki ;
Yamaguchi, Mitsutaka ;
Yamamoto, Seiichi ;
Toshito, Toshiyuki ;
Kawachi, Naoki .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (12) :5006-5020
[2]   VISUALIZATION AND TRANSPORT OF POSITRON EMISSION FROM PROTON ACTIVATION INVIVO [J].
BENNETT, GW ;
ARCHAMBEAU, JO ;
ARCHAMBEAU, BE ;
MELTZER, JI ;
WINGATE, CL .
SCIENCE, 1978, 200 (4346) :1151-1153
[3]   Spread-out Bragg peak proton FLASH irradiation using a clinical synchrocyclotron: Proof of concept and ion chamber characterization [J].
Darafsheh, Arash ;
Hao, Yao ;
Zhao, Xiandong ;
Zwart, Townsend ;
Wagner, Miles ;
Evans, Tucker ;
Reynoso, Francisco ;
Zhao, Tianyu .
MEDICAL PHYSICS, 2021, :4472-4484
[4]   Feasibility of proton FLASH irradiation using a synchrocyclotron for preclinical studies [J].
Darafsheh, Arash ;
Hao, Yao ;
Zwart, Townsend ;
Wagner, Miles ;
Catanzano, Daniel ;
Williamson, Jeffrey F. ;
Knutson, Nels ;
Sun, Baozhou ;
Mutic, Sasa ;
Zhao, Tianyu .
MEDICAL PHYSICS, 2020, 47 (09) :4348-4355
[5]   A MONTE-CARLO SIMULATION OF THE POSSIBLE USE OF POSITRON EMISSION TOMOGRAPHY IN PROTON RADIOTHERAPY [J].
DELGUERRA, A ;
DIDOMENICO, G ;
GAMBACCINI, M ;
MARZIANI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) :379-384
[6]   3D prompt gamma imaging for proton beam range verification [J].
Draeger, E. ;
Mackin, D. ;
Peterson, S. ;
Chen, H. ;
Avery, S. ;
Beddar, S. ;
Polf, J. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (03)
[7]   Usefulness of positron-emission tomographic images after proton therapy [J].
Hishikawa, Y ;
Kagawa, K ;
Murakami, M ;
Sakai, H ;
Akagi, T ;
Abe, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1388-1391
[8]   First test of the prompt gamma ray timing method with heterogeneous targets at a clinical proton therapy facility [J].
Hueso-Gonzalez, Fernando ;
Enghardt, Wolfgang ;
Fiedler, Fine ;
Golnik, Christian ;
Janssens, Guillaume ;
Petzoldt, Johannes ;
Prieels, Damien ;
Priegnitz, Marlen ;
Roemer, Katja E. ;
Smeets, Julien ;
Stappen, Francois Vander ;
Wagner, Andreas ;
Pausch, Guntram .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (16) :6247-6272
[9]   Quantitative comparison of suitability of various beams for range monitoring with induced β+ activity in hadron therapy [J].
Inaniwa, T ;
Tomitani, T ;
Kohno, T ;
Kanai, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (06) :1131-1145
[10]   Positron camera for range verification of heavy-ion radiotherapy [J].
Iseki, Y ;
Mizuno, H ;
Futami, Y ;
Tomitani, T ;
Kanai, T ;
Kanazawa, M ;
Kitagawa, A ;
Murakami, T ;
Nishio, T ;
Suda, M ;
Urakabe, E ;
Yunoki, A ;
Sakai, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 515 (03) :840-849