Common carotid segmentation in 18F-sodium fluoride PET/CT scans: Head-to-head comparison of artificial intelligence-based and manual method

被引:7
作者
Piri, Reza [1 ,2 ]
Hamakan, Yaran [1 ]
Vang, Ask [1 ]
Edenbrandt, Lars [3 ,4 ]
Larsson, Mans [5 ]
Enqvist, Olof [5 ,6 ]
Gerke, Oke [1 ,2 ]
Hoilund-Carlsen, Poul Flemming [1 ,2 ]
机构
[1] Odense Univ Hosp, Dept Nucl Med, DK-5000 Odense, Denmark
[2] Univ Southern Denmark, Dept Clin Res, Odense, Denmark
[3] Univ Gothenburg, Sahlgrenska Acad, Inst Med, Dept Mol & Clin Med, Gothenburg, Sweden
[4] Sahlgrens Univ Hosp, Dept Clin Physiol, Reg Vastra Gotaland, Gothenburg, Sweden
[5] Eigenvision AB, Malmo, Sweden
[6] Chalmers Univ Technol, Dept Elect Engn, Gothenburg, Sweden
关键词
artificial intelligence; atherosclerosis; carotids; positron emission tomography; ATHEROSCLEROSIS; QUANTIFICATION; RISK; CALCIFICATION; INFLAMMATION; ARTERIES; DISEASE; LESIONS; STROKE;
D O I
10.1111/cpf.12793
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Background Carotid atherosclerosis is a major cause of stroke, traditionally diagnosed late. Positron emission tomography/computed tomography (PET/CT) with F-18-sodium fluoride (NaF) detects arterial wall micro-calcification long before macro-calcification becomes detectable by ultrasound, CT or magnetic resonance imaging. However, manual PET/CT processing is time-consuming and requires experience. We compared a convolutional neural network (CNN) approach with manual segmentation of the common carotids. Methods Segmentation in NaF-PET/CT scans of 29 healthy volunteers and 20 angina pectoris patients were compared for segmented volume (Vol) and mean, maximal, and total standardized uptake values (SUVmean, SUVmax, and SUVtotal). SUVmean was the average of SUVmeans within the VOI, SUVmax the highest SUV in all voxels in the VOI, and SUVtotal the SUVmean multiplied by the Vol of the VOI. Intra and Interobserver variability with manual segmentation was examined in 25 randomly selected scans. Results Bias for Vol, SUVmean, SUVmax, and SUVtotal were 1.33 +/- 2.06, -0.01 +/- 0.05, 0.09 +/- 0.48, and 1.18 +/- 1.99 in the left and 1.89 +/- 1.5, -0.07 +/- 0.12, 0.05 +/- 0.47, and 1.61 +/- 1.47, respectively, in the right common carotid artery. Manual segmentation lasted typically 20 min versus 1 min with the CNN-based approach. Mean Vol deviation at repeat manual segmentation was 14% and 27% in left and right common carotids. Conclusions CNN-based segmentation was much faster and provided SUVmean values virtually identical to manually obtained ones, suggesting CNN-based analysis as a promising substitute of slow and cumbersome manual processing.
引用
收藏
页码:71 / 77
页数:7
相关论文
共 37 条
[1]   Global Overview of the Epidemiology of Atherosclerotic Cardiovascular Disease [J].
Barquera, Simon ;
Pedroza-Tobias, Andrea ;
Medina, Catalina ;
Hernandez-Barrera, Lucia ;
Bibbins-Domingo, Kirsten ;
Lozano, Rafael ;
Moran, Andrew E. .
ARCHIVES OF MEDICAL RESEARCH, 2015, 46 (05) :328-338
[2]   Deep learning for segmentation of 49 selected bones in CT scans: First step in automated PET/CT-based 3D quantification of skeletal metastases [J].
Belal, Sarah Lindgren ;
Sadik, May ;
Kaboteh, Reza ;
Enqvist, Olof ;
Ulen, Johannes ;
Poulsen, Mads H. ;
Simonsen, Jane ;
Hoilund-Carlsen, Poul F. ;
Edenbrandt, Lars ;
Tragardh, Elin .
EUROPEAN JOURNAL OF RADIOLOGY, 2019, 113 :89-95
[3]   Thoracic aorta calcification but not inflammation is associated with increased cardiovascular disease risk: results of the CAMONA study [J].
Blomberg, Bjorn A. ;
de Jong, Pim A. ;
Thomassen, Anders ;
Lam, Marnix G. E. ;
Vach, Werner ;
Olsen, Michael H. ;
Mali, Willem P. T. M. ;
Narula, Jagat ;
Alavi, Abass ;
Hoilund-Carlsen, Poul F. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 (02) :249-258
[4]   Delayed 18F-fluorodeoxyglucose PET/CT imaging improves quantitation of atherosclerotic plaque inflammation: Results from the CAMONA study [J].
Blomberg, Bjorn A. ;
Thomassen, Anders ;
Takx, Richard A. P. ;
Hildebrandt, Malene G. ;
Simonsen, Jane A. ;
Buch-Olsen, Karen M. ;
Diederichsen, Axel C. P. ;
Mickley, Hans ;
Alavi, Abass ;
Hoilund-Carlsen, Poul F. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (03) :588-597
[5]   Impact of Personal Characteristics and Technical Factors on Quantification of Sodium 18F-Fluoride Uptake in Human Arteries: Prospective Evaluation of Healthy Subjects [J].
Blomberg, Bjorn Alexander ;
Thomassen, Anders ;
de Jong, Pim A. ;
Simonsen, Jane A. ;
Lam, Marnix G. E. H. ;
Nielsen, Anne L. ;
Mickley, Hans ;
Mali, Willem P. T. M. ;
Alavi, Abass ;
Hoilund-Carlsen, Poul F. .
JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (10) :1534-1540
[6]   Updating PET/CT performance standards and PET/CT interpretation criteria should go hand in hand [J].
Boellaard, Ronald ;
Sera, Terez ;
Kaalep, Andres ;
Hoekstra, Otto S. ;
Barrington, Sally F. ;
Zijlstra, Josee M. .
EJNMMI RESEARCH, 2019, 9 (01)
[7]   FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0 [J].
Boellaard, Ronald ;
Delgado-Bolton, Roberto ;
Oyen, Wim J. G. ;
Giammarile, Francesco ;
Tatsch, Klaus ;
Eschner, Wolfgang ;
Verzijlbergen, Fred J. ;
Barrington, Sally F. ;
Pike, Lucy C. ;
Weber, Wolfgang A. ;
Stroobants, Sigrid ;
Delbeke, Dominique ;
Donohoe, Kevin J. ;
Holbrook, Scott ;
Graham, Michael M. ;
Testanera, Giorgio ;
Hoekstra, Otto S. ;
Zijlstra, Josee ;
Visser, Eric ;
Hoekstra, Corneline J. ;
Pruim, Jan ;
Willemsen, Antoon ;
Arends, Bertjan ;
Kotzerke, Joerg ;
Bockisch, Andreas ;
Beyer, Thomas ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (02) :328-354
[8]   Exact Parametric Confidence Intervals for Bland-Altman Limits of Agreement [J].
Carkeet, Andrew .
OPTOMETRY AND VISION SCIENCE, 2015, 92 (03) :E71-E80
[9]  
Chambless LE, 2000, AM J EPIDEMIOL, V151, P478
[10]  
Cicek O., 2016, P INT C MED IM COMP