Personalized asymmetric multiple PID to automatize the procedure of intravenous general anesthesia

被引:1
作者
Castaneda, Carlos E. [1 ]
Orozco-Lopez, Onofre [1 ]
Abad-Gurumeta, A. [2 ]
Hernando, M. Elena [3 ,4 ]
Rodriguez-Herrero, Agustin [3 ,4 ]
机构
[1] Univ Guadalajara, Univ Ctr Lagos, Guadalajara, Mexico
[2] Infanta Leonor Hosp, Dept Anesthesiol, Madrid, Spain
[3] Univ Politecn Madrid, Biomed Technol Ctr CTB, Madrid, Spain
[4] Inst Salud Carlos III, CIBER Bioingn Biomat & Nanomed, Madrid, Spain
关键词
PID; Multiple input multiple output; Asymmetric; Intravenous anesthesia; Propofol; Remifentanil; Synergy; CONTROL SCHEME; PROPOFOL; HYPNOSIS; DEPTH; MAINTENANCE; SYSTEMS;
D O I
10.1016/j.jprocont.2023.103019
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
General anesthesia has improved considerably over the years due to the automation of drug perfusion through closed-loop control which provided numerous advantages regarding efficiency and patient safety. This work describes the implementation and Validation of a control algorithm based on a multiple-input and multiple-output proportional-integral-derivative (MIMO-PID) controller. The main objective is to keep patients closer to the desired range of monitor values into satisfactory anesthetic state.The MIMO-PID decides the drug perfusion of propofol, remifentanil, and rocuronium based on the bispectral index, noxious stimulation response, and neuromuscular blockade monitorization. The main MIMO-PID variant is based on a feed-forward target which is used to define the asymmetric errors. The asymmetry allows different errors to be defined in each branch of each single-input and single output proportional-integral-derivative. The MIMO-PID is tuned using the criterion of integral absolute error and their performance is evaluated through a range of experiments applying different metrics to evaluate monitors and drug perfusions. The satisfactory anesthetic state is achieved during the induction phase of anesthesia; however, we also had situations with a low relative deviation from the registered median values. We obtained good results by using a strict disturbance rejection.The controller was able to mitigate the surgery influence on the anesthetics effects when there were less than 20, the patient remained close to the desired anesthetic state. In the presence of significant or more numerous stimuli, the controller was able to achieve a satisfactory anesthetic state within a short period of typically less than 7 min.& COPY; 2023 Elsevier Ltd. All rights reserved.
引用
收藏
页数:16
相关论文
共 33 条
[1]  
Billard Valerie, 2015, F1000Res, V4, DOI 10.12688/f1000research.6601.1
[2]  
De Keyser R, 2017, ASIA CONTROL CONF AS, P1164, DOI 10.1109/ASCC.2017.8287335
[3]   Evaluation of a novel closed-loop total intravenous anaesthesia drug delivery system: a randomized controlled trial [J].
Hemmerling, T. M. ;
Arbeid, E. ;
Wehbe, M. ;
Cyr, S. ;
Taddei, R. ;
Zaouter, C. .
BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) :1031-1039
[4]   An Adaptive Multi-critic Neuro-fuzzy Control Framework for Intravenous Anesthesia Administration [J].
Khodaei, Mohammad Javad ;
Balaghi, M. Hadi, I ;
Mehrvarz, Amin ;
Jalili, Nader .
IFAC PAPERSONLINE, 2019, 51 (34) :202-207
[5]   The Three Laws of Autonomous and Closed-Loop Systems in Anesthesia [J].
Kuck, Kai ;
Johnson, Ken B. .
ANESTHESIA AND ANALGESIA, 2017, 124 (02) :377-380
[6]   Noxious Stimulation Response Index A Novel Anesthetic State Index Based on Hypnotic-Opioid Interaction [J].
Luginbuehl, Martin ;
Schumacher, Peter M. ;
Vuilleumier, Pascal ;
Vereecke, Hugo ;
Heyse, Bjoern ;
Bouillon, Thomas W. ;
Struys, Michel M. R. F. .
ANESTHESIOLOGY, 2010, 112 (04) :872-880
[7]   Different perspectives for monitoring nociception during general anesthesia [J].
Martinez-Vazquez, Pablo ;
Weber Jensen, Erik .
KOREAN JOURNAL OF ANESTHESIOLOGY, 2022, 75 (02) :112-123
[8]  
Merigo L., 2017, 2017 3 INT C EV BAS, P1
[9]   Event-based control tuning of propofol and remifentanil coadministration for general anaesthesia [J].
Merigo, Luca ;
Padula, Fabrizio ;
Latronico, Nicola ;
Paltenghi, Massimiliano ;
Visioli, Antonio .
IET CONTROL THEORY AND APPLICATIONS, 2020, 14 (19) :2995-3008
[10]   Optimized PID control of propofol and remifentanil coadministration for general anesthesia [J].
Merigo, Luca ;
Padula, Fabrizio ;
Latronico, Nicola ;
Paltenghi, Massimiliano ;
Visioli, Antonio .
COMMUNICATIONS IN NONLINEAR SCIENCE AND NUMERICAL SIMULATION, 2019, 72 :194-212