Performance Characteristics of a New-Generation Digital Bismuth Germanium Oxide PET/CT System, Omni Legend 32, According to NEMA NU 2-2018 Standards

被引:19
作者
Yamagishi, Shin [1 ]
Miwa, Kenta [2 ]
Kamitaki, Shun [1 ]
Anraku, Kouichi [1 ]
Sato, Shun [1 ]
Yamao, Tensho [2 ]
Kubo, Hitoshi [2 ]
Miyaji, Noriaki [2 ]
Oguchi, Kazuhiro [1 ]
机构
[1] Aizawa Hosp, Ctr Radiol & Diagnost Imaging, Nagano, Japan
[2] Fukushima Med Univ, Sch Hlth Sci, Dept Radiol Sci, Fukushima, Japan
关键词
PET/CT; NEMA; performance evaluation; Omni Legend; SiPM; RECONSTRUCTION;
D O I
10.2967/jnumed.123.266140
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Omni Legend 32 PET/CT system features silicon photomultiplier (SiPM)-based detectors with bismuth germanium oxide crystals and a 32-cm axial field of view (FOV). The present study aimed to determine the performance characteristics of the Omni Legend 32 PET/CT system according to National Electrical Manufacturers Association (NEMA) NU 2-2018 standards. Methods: The PET component of this system comprises 22 detector modules; each module contains 24 detector blocks with 72 bismuth germanium oxide crystals with a volume of 4.1 3 4.1 3 30 mm coupled to 18 SiPM devices with a 6 3 6 mm area, resulting in an axial FOV of 32 cm. The spatial resolution, sensitivity, count rate performance, and image quality delivered by PET were evaluated using the NEMA NU 2-2018 standard. PET images of 2 patients were evaluated to get a visual first impression of the Omni Legend 32 PET/CT system together with Precision DL. Results: The average spatial resolution at 1, 10, and 20 cm from the central axis was 4.3, 5.3, and 6.2 mm, respectively, for filtered backprojection and 3.7, 4.3, and 5.1 mm, respectively, for ordered-subset expectation maximization. The NEMA sensitivity was 47.30 and 47.05 cps/kBq at the axial center of the FOV and at a 10-cm radial offset, respectively. The scatter fraction, count rate accuracy, and peak noiseequivalent count rates were 35.4%, 1.7%, and 501.7 kcps, respectively, at 15.7 kBq/mL. Contrast recovery for the NEMA body phantom from the smallest to the largest sphere ranged from 61.3% to 93.0%, with a background variability of 5.4%-11.7% and a lung error of 5.1% for Q.Clear (beta-value, 50). Good patient image quality was obtained with the Omni Legend 32. Conclusion: The Omni Legend 32 has class-leading sensitivity and count rates within the category of whole-body PET systems while maintaining spatial resolution broadly comparable to that of other current SiPM-based PET/CT systems. This combination of properties results in a very good image quality.
引用
收藏
页码:1990 / 1997
页数:8
相关论文
共 33 条
[21]   Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard [J].
Rausch, Ivo ;
Ruiz, Agustin ;
Valverde-Pascual, Itziar ;
Cal-Gonzalez, Jacobo ;
Beyer, Thomas ;
Carrio, Ignasi .
JOURNAL OF NUCLEAR MEDICINE, 2019, 60 (04) :561-567
[22]   Performance Characteristics of the Whole-Body Discovery IQ PET/CT System [J].
Reynes-Llompart, Gabriel ;
Gamez-Cenzano, Cristina ;
Romero-Zayas, Inmaculada ;
Rodriguez-Bel, Laura ;
Vercher-Conejero, Jose L. ;
Marti-Climent, Josep M. .
JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (07) :1155-1161
[23]   Digital PET/CT: a new intriguing chance for clinical nuclear medicine and personalized molecular imaging [J].
Schillaci, Orazio ;
Urbano, Nicoletta .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (06) :1222-1225
[24]   Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View [J].
Spencer, Benjamin A. ;
Berg, Eric ;
Schmall, Jeffrey P. ;
Omidvari, Negar ;
Leung, Edwin K. ;
Abdelhafez, Yasser G. ;
Tang, Songsong ;
Deng, Zilin ;
Dong, Yun ;
Lv, Yang ;
Bao, Jun ;
Liu, Weiping ;
Li, Hongdi ;
Jones, Terry ;
Badawi, Ramsey D. ;
Cherry, Simon R. .
JOURNAL OF NUCLEAR MEDICINE, 2021, 62 (06) :861-870
[25]   Phantom and Clinical Evaluation of the Bayesian Penalized Likelihood Reconstruction Algorithm Q.Clear on an LYSO PET/CT System [J].
Teoh, Eugene J. ;
McGowan, Daniel R. ;
Macpherson, Ruth E. ;
Bradley, Kevin M. ;
Gleeson, Fergus V. .
JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (09) :1447-1452
[26]   Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET [J].
van der Vos, Charlotte S. ;
Koopman, Danielle ;
Rijnsdorp, Sjoerd ;
Arends, Albert J. ;
Boellaard, Ronald ;
van Dalen, Jorn A. ;
Lubberink, Mark ;
Willemsen, Antoon T. M. ;
Visser, Eric P. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 :S4-S16
[27]   Performance Characteristics of the Digital Biograph Vision PET/CT System [J].
van Sluis, Joyce ;
de Jong, Johan ;
Schaar, Jenny ;
Noordzij, Walter ;
van Snick, Paul ;
Dierckx, Rudi ;
Borra, Ronald ;
Willemsen, Antoon ;
Boellaard, Ronald .
JOURNAL OF NUCLEAR MEDICINE, 2019, 60 (07) :1031-1036
[28]   Determination of optimal regularization factor in Bayesian penalized likelihood reconstruction of brain PET images using [18F]FDG and [11C]PiB [J].
Wagatsuma, Kei ;
Miwa, Kenta ;
Kamitaka, Yuto ;
Koike, Emiya ;
Yamao, Tensho ;
Yoshii, Tokiya ;
Kobayashi, Rinya ;
Nezu, Shogo ;
Sugamata, Yuta ;
Miyaji, Noriaki ;
Imabayashi, Etsuko ;
Ishibashi, Kenji ;
Toyohara, Jun ;
Ishii, Kenji .
MEDICAL PHYSICS, 2022, 49 (05) :2995-3005
[29]   Direct comparison of brain [18F]FDG images acquired by SiPM-based and PMT-based PET/CT: phantom and clinical studies [J].
Wagatsuma, Kei ;
Sakata, Muneyuki ;
Ishibashi, Kenji ;
Hirayama, Akira ;
Kawakami, Hirofumi ;
Miwa, Kenta ;
Suzuki, Yukihisa ;
Ishii, Kenji .
EJNMMI PHYSICS, 2020, 7 (01)
[30]   Comparison between new-generation SiPM-based and conventional PMT-based TOF-PET/CT [J].
Wagatsuma, Kei ;
Miwa, Kenta ;
Sakata, Muneyuki ;
Oda, Keiichi ;
Ono, Haruka ;
Kameyama, Masashi ;
Toyohara, Jun ;
Ishii, Kenji .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 42 :203-210