Performance Characteristics of a New-Generation Digital Bismuth Germanium Oxide PET/CT System, Omni Legend 32, According to NEMA NU 2-2018 Standards

被引:19
作者
Yamagishi, Shin [1 ]
Miwa, Kenta [2 ]
Kamitaki, Shun [1 ]
Anraku, Kouichi [1 ]
Sato, Shun [1 ]
Yamao, Tensho [2 ]
Kubo, Hitoshi [2 ]
Miyaji, Noriaki [2 ]
Oguchi, Kazuhiro [1 ]
机构
[1] Aizawa Hosp, Ctr Radiol & Diagnost Imaging, Nagano, Japan
[2] Fukushima Med Univ, Sch Hlth Sci, Dept Radiol Sci, Fukushima, Japan
关键词
PET/CT; NEMA; performance evaluation; Omni Legend; SiPM; RECONSTRUCTION;
D O I
10.2967/jnumed.123.266140
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Omni Legend 32 PET/CT system features silicon photomultiplier (SiPM)-based detectors with bismuth germanium oxide crystals and a 32-cm axial field of view (FOV). The present study aimed to determine the performance characteristics of the Omni Legend 32 PET/CT system according to National Electrical Manufacturers Association (NEMA) NU 2-2018 standards. Methods: The PET component of this system comprises 22 detector modules; each module contains 24 detector blocks with 72 bismuth germanium oxide crystals with a volume of 4.1 3 4.1 3 30 mm coupled to 18 SiPM devices with a 6 3 6 mm area, resulting in an axial FOV of 32 cm. The spatial resolution, sensitivity, count rate performance, and image quality delivered by PET were evaluated using the NEMA NU 2-2018 standard. PET images of 2 patients were evaluated to get a visual first impression of the Omni Legend 32 PET/CT system together with Precision DL. Results: The average spatial resolution at 1, 10, and 20 cm from the central axis was 4.3, 5.3, and 6.2 mm, respectively, for filtered backprojection and 3.7, 4.3, and 5.1 mm, respectively, for ordered-subset expectation maximization. The NEMA sensitivity was 47.30 and 47.05 cps/kBq at the axial center of the FOV and at a 10-cm radial offset, respectively. The scatter fraction, count rate accuracy, and peak noiseequivalent count rates were 35.4%, 1.7%, and 501.7 kcps, respectively, at 15.7 kBq/mL. Contrast recovery for the NEMA body phantom from the smallest to the largest sphere ranged from 61.3% to 93.0%, with a background variability of 5.4%-11.7% and a lung error of 5.1% for Q.Clear (beta-value, 50). Good patient image quality was obtained with the Omni Legend 32. Conclusion: The Omni Legend 32 has class-leading sensitivity and count rates within the category of whole-body PET systems while maintaining spatial resolution broadly comparable to that of other current SiPM-based PET/CT systems. This combination of properties results in a very good image quality.
引用
收藏
页码:1990 / 1997
页数:8
相关论文
共 33 条
[1]   New PET technologies - embracing progress and pushing the limits [J].
Aide, Nicolas ;
Lasnon, Charline ;
Kesner, Adam ;
Levin, Craig S. ;
Buvat, Irene ;
Iagaru, Andrei ;
Hermann, Ken ;
Badawi, Ramsey D. ;
Cherry, Simon R. ;
Bradley, Kevin M. ;
McGowan, Daniel R. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (09) :2711-2726
[2]   Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [18F]-FDG PET/CT [J].
Alberts, Ian ;
Sachpekidis, Christos ;
Prenosil, George ;
Viscione, Marco ;
Bohn, Karl Peter ;
Mingels, Clemens ;
Shi, Kuangyu ;
Ashar-Oromieh, Ali ;
Rominger, Axel .
ANNALS OF NUCLEAR MEDICINE, 2021, 35 (04) :485-492
[3]   The effect of camera geometry on singles flux, scatter fraction and trues and randoms sensitivity for cylindrical 3D PET - A simulation study. [J].
Badawi, RD ;
Kohlmyer, SG ;
Harrison, RL ;
Vannoy, SD ;
Lewellen, TK .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) :1228-1232
[4]   Performance comparison of two commercial BGO-based PET/CT scanners using NEMA NU 2-2001 [J].
Bolard, Gregory ;
Prior, John O. ;
Modolo, Luca ;
Delaloye, Angelika Bischof ;
Kosinski, Marek ;
Wastiel, Claude ;
Malterre, Jerome ;
Bulling, Shelley ;
Bochud, Francois ;
Verdun, Francis R. .
MEDICAL PHYSICS, 2007, 34 (07) :2708-2717
[5]   Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard [J].
Chen, Shuguang ;
Hu, Pengcheng ;
Gu, Yushen ;
Yu, Haojun ;
Shi, Hongcheng .
EJNMMI PHYSICS, 2020, 7 (01)
[6]   Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards [J].
De Ponti, E. ;
Morzenti, S. ;
Guerra, L. ;
Pasquali, C. ;
Arosio, M. ;
Bettinardi, V. ;
Crespi, A. ;
Gilardi, M. C. ;
Messa, C. .
MEDICAL PHYSICS, 2011, 38 (02) :968-974
[7]   Phantom-based image quality assessment of clinical 18F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CT [J].
Gnesin, Silvano ;
Kieffer, Christine ;
Zeimpekis, Konstantinos ;
Papazyan, Jean-Pierre ;
Guignard, Renaud ;
Prior, John O. ;
Verdun, Francis R. ;
Lima, Thiago V. M. .
EJNMMI PHYSICS, 2020, 7 (01)
[8]   Studies of a Next-Generation Silicon-Photomultiplier-Based Time-of-Flight PET/CT System [J].
Hsu, David F. C. ;
Ilan, Ezgi ;
Peterson, William T. ;
Uribe, Jorge ;
Lubberink, Mark ;
Levin, Craig S. .
JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (09) :1511-1518
[9]   Quantitative comparison of data-driven gating and external hardware gating for 18F-FDG PET-MRI in patients with esophageal tumors [J].
Kvernby, Sofia ;
Hult, Nafsika Korsavidou ;
Lindstrom, Elin ;
Sigfridsson, Jonathan ;
Linder, Gustav ;
Hedberg, Jakob ;
Ahlstrom, Hakan ;
Bjerner, Tomas ;
Lubberink, Mark .
EUROPEAN JOURNAL OF HYBRID IMAGING, 2021, 5 (01)
[10]   Novel detector technology for clinical PET [J].
Lecomte, Roger .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 :69-85