摘要
Heart diseases rank among the most fatal health concerns globally, with the majority being preventable through early diagnosis and effective treatment. Electrocardiogram (ECG) analysis is critical in detecting heart diseases, as it captures the heart's electrical activities. For continuous monitoring, wearable electrocardiographic devices must ensure user comfort over extended periods, typically 24 to 48 h. These devices demand specialized algorithms with low computational complexity to accommodate memory and power consumption constraints. One of the most crucial aspects of ECG signals is accurately detecting heartbeat intervals, specifically the R peaks. In this study, we introduce a novel algorithm designed for wearable devices, offering two primary attributes: robustness against noise and low computational complexity. Our algorithm entails fitting a least-squares parabola to the ECG signal and adaptively shaping it as it sweeps through the signal. Notably, our proposed algorithm eliminates the need for band-pass filters, which can inadvertently smooth the R peaks, making them more challenging to identify. We compared the algorithm's performance using two extensive databases: the meta-database QT database and the BIH-MIT database. Importantly, our method does not necessitate the precise localization of the ECG signal's isoelectric line, contributing to its low computational complexity. In the analysis of the QT database, our algorithm demonstrated a substantial advantage over the classical Pan-Tompkins algorithm and maintained competitiveness with state-of-the-art approaches. In the case of the BIH-MIT database, the performance results were more conservative; they continued to underscore the real-world utility of our algorithm in clinical contexts.
引用
收藏
页数:17
相关论文
共 50 条
  • [41] SCINTIGRAPHIC EVALUATION OF R-PEAK CHANGES IN THE STRESS ECG
    FRIDRICH, L
    GASSNER, A
    VAGNER, M
    SYKORA, J
    MOSTBECK, G
    MAROSI, L
    PICHLER, M
    ACTA MEDICA AUSTRIACA, 1984, 11 : 170 - 171
  • [42] R-peak Extraction for Wireless ECG Monitoring System
    Dave, Tejal
    Pandya, Utpal
    2018 INTERNATIONAL CONFERENCE ON ADVANCES IN COMPUTING, COMMUNICATIONS AND INFORMATICS (ICACCI), 2018, : 995 - 999
  • [43] A Dual Self-Calibrating Framework for Noninvasive Fetal ECG R-Peak Detection
    Qiao, Lihong
    Hu, Shuai
    Xiao, Bin
    Bi, Xiuli
    Li, Weisheng
    Gao, Xinbo
    IEEE INTERNET OF THINGS JOURNAL, 2023, 10 (18) : 16579 - 16593
  • [44] Design And Implementation Of A Low-Complexity R-peak Detection Algorithm
    Chen, Zirong
    Lin, Zhongmin
    Chen, Hao
    Wang, Xin-an
    Wu, Yingying
    2018 CHINA SEMICONDUCTOR TECHNOLOGY INTERNATIONAL CONFERENCE (CSTIC), 2018,
  • [45] A Novel Method for R-peak Detection in Noisy ECG Signals Using EEMD and ICA
    Safari, Amirhossein
    Hesar, Hamed Danandeh
    Mohebbi, Maryam
    Faradji, Farhad
    2016 23RD IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING AND 2016 1ST INTERNATIONAL IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING (ICBME), 2016, : 150 - 153
  • [46] R-Peak Detection Using Chaos Analysis in Standard and Real Time ECG Databases
    Gupta, V.
    Mittal, M.
    Mittal, V.
    IRBM, 2019, 40 (06) : 341 - 354
  • [47] Rapid Processor Customization for Design Optimization: A Case Study of ECG R-peak Detection
    Milosevic, Mladen
    Jovanov, Emil
    Milenkovic, Aleksandar
    2011 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2011, : 209 - 212
  • [48] Towards Reliable ECG Analysis: Addressing Validation Gaps in the Electrocardiographic R-Peak Detection
    Ali, Syed Talha Abid
    Kim, Sebin
    Kim, Young-Joon
    APPLIED SCIENCES-BASEL, 2024, 14 (21):
  • [49] REWARD: Design, Optimization, and Evaluation of a Real-Time Relative-Energy Wearable R-Peak Detection Algorithm
    Orlandic, Lara
    de Giovanni, Elisabetta
    Arza, Adriana
    Yazdani, Sasan
    Vesin, Jean-Marc
    Atienza, David
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 3341 - 3347
  • [50] Analysis and Detection R-Peak Detection using Modified Pan-Tompkins Algorithm
    Sathyapriya, L.
    Murali, L.
    Manigandan, T.
    2014 INTERNATIONAL CONFERENCE ON ADVANCED COMMUNICATION CONTROL AND COMPUTING TECHNOLOGIES (ICACCCT), 2014, : 483 - 487