Characterization of Curved Piezoelectric Micromachined Ultrasound Transducers Fabricated by Chip-Scale Glass Blowing Technique

被引:3
作者
Huang, Chichen [1 ]
Khandare, Shubham P. [2 ]
Kothapalli, Sri-Rajasekhar [2 ]
Tadigadapa, Srinivas [1 ]
机构
[1] Northeastern Univ, Dept Elect & Comp Engn, Boston, MA 02115 USA
[2] Penn State Univ, Dept Biomed Engn, University Pk, PA 16802 USA
基金
美国国家科学基金会;
关键词
Mechanical sensors; ultrasound transducers; aluminum nitride; chip scale glassblowing; curved glass structures; piezoelectric micromachined ultrasound transducer (pMUT);
D O I
10.1109/LSENS.2023.3308112
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This letter presents the development of curved piezoelectric micromachined ultrasound transducers (pMUTs) using a novel chip-scale glass-blowing method on suspended glass templates. This new approach allows for controllable diaphragm curvature and high fill-factor arrays of varying sizes, making it possible to study the impact of the curvature on the performance of the pMUTs. A finite element analysis (FEA) model was built to guide the design. Thirty percent Scandium-doped Aluminum Nitride (Sc-AlN) was chosen for good piezoelectrical coefficient and biocompatibility. The 100 nm platinum/750 nm Sc-AlN/100 nm gold films are sputtered on the curved glass membrane with diameters ranging from 75 to 750 mu m. pMUT with a diameter of 150 mu m with 5.6 mu m depth of curvature had resonance frequencies of 2.2 MHz along with center displacements up to 154 nm/V in air. Impedance measurements at resonance on the pMUTs show k(eff )(2)of 1.31%. Optimal curvatures were found experimentally and matched the FEA model. Changes in mode shapes were found when curvatures were deeper than the optimal value.
引用
收藏
页数:4
相关论文
共 14 条
[1]  
Akhbari S, 2015, PROC IEEE MICR ELECT, P928, DOI 10.1109/MEMSYS.2015.7051112
[2]  
Akhbari S, 2014, 2014 IEEE 27TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), P124, DOI 10.1109/MEMSYS.2014.6765589
[3]   Novel 3D PZT thin film structure for micromechanics [J].
Calame, F. ;
Muralt, P. .
JOURNAL OF ELECTROCERAMICS, 2007, 19 (04) :399-402
[4]   DESIGN OF EFFICIENT BROAD-BAND PIEZOELECTRIC TRANSDUCERS [J].
DESILETS, CS ;
FRASER, JD ;
KINO, GS .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (03) :115-125
[5]   Fabrication of MEMS ZnO domeshaped-diaphragm transducers for high-frequency ultrasonic imaging [J].
Feng, GH ;
Sharp, CC ;
Zhou, QF ;
Pang, W ;
Kim, ES ;
Shung, KK .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2005, 15 (03) :586-590
[6]  
Guedes A., 2011, TRANSDUCERS 2011 - 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, P2062, DOI 10.1109/TRANSDUCERS.2011.5969223
[7]   Three-dimensional micro electromechanical system piezoelectric ultrasound transducer [J].
Hajati, Arman ;
Latev, Dimitre ;
Gardner, Deane ;
Hajati, Azadeh ;
Imai, Darren ;
Torrey, Marc ;
Schoeppler, Martin .
APPLIED PHYSICS LETTERS, 2012, 101 (25)
[8]  
Jiang G.-L., 2018, P IEEE INT ULTR S, P1
[9]  
Jiang XY, 2016, PROC IEEE MICR ELECT, P107
[10]   Innovation - Healing sound: the use of ultrasound in drug delivery and other therapeutic applications [J].
Mitragotri, S .
NATURE REVIEWS DRUG DISCOVERY, 2005, 4 (03) :255-260