Albuterol Delivery During Invasive Mechanical Ventilation via In-Line Intrapulmonary Percussive Ventilation in a Pediatric Lung Model

被引:3
作者
Berlinski, Ariel [1 ,2 ,3 ,4 ]
Willis, L. Denise [3 ]
机构
[1] Univ Arkansas Med Sci, Coll Med, Dept Pediat, Little Rock, AR USA
[2] Arkansas Childrens Res Inst, Pediat Aerosol Res Lab, Little Rock, AR USA
[3] Arkansas Childrens Hosp, Resp Care Serv, Little Rock, AR USA
[4] Univ Arkansas Med Sci, 1 Childrens Way,Slot 512-17, Little Rock, AR 72202 USA
关键词
intrapul monary percussive ventilation; airway clearance; aerosol; lung deposition; in-line adapter; lung model; pediatric; albuterol; nebulizer; AIRWAY CLEARANCE TECHNIQUES; THERAPEUTIC AEROSOLS; NEBULIZATION; DEPOSITION;
D O I
10.4187/respcare.11311
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Patients receiving mechanical ventilation often require airway clearance and inhaled therapies. Intrapulmonary percussive ventilation (IPV) combines a high -frequency percussive ventilator with a jet nebulizer. Data on aerosol delivery efficiency of the device are scarce. We evaluated albuterol delivery efficiency while using an IPV in -line adapter under different conditions. METHODS: A pediatric lung model of invasive mechanical ventilation was used. The following independent variables were evaluated: lung condition (normal vs ARDS), ventilator mode (adaptive pressure ventilation vs pressure control), percent opening of adapter (0% vs 25% vs 50%), IPV driving pressure (25 psi vs 40 psi), IPV percussion setting (easy vs hard), and endotracheal tube (ETT) size (3.5 mm vs 5.5 mm). Albuterol delivery efficiency (mass captured in the filter expressed as percentage of loading dose) was selected as the dependent variable. Albuterol was captured on a filter at the tip of the ETT and quantified via spectrophotometry (276 nm). RESULTS: Albuterol delivery efficiency ranged from 0-2.89%. Median (interquartile range) and 95% CI around the median were 0.54% (0.37- 1.00) and 0.50-0.63%, respectively. The coefficient of determination (R2) for the model including all variables was 0.363. The 2 main contributors were percent of adapter opening (R2 0.30) and IPV setting (R2 0.039). CONCLUSIONS: Albuterol delivery during invasive mechanical ventilation via in -line IPV in a pediatric lung model was inefficient. Alternative methods of delivering bronchodilators and other inhaled medications should be considered when IPV is used.
引用
收藏
页码:61 / 67
页数:7
相关论文
共 20 条
  • [1] THE DELIVERY OF THERAPEUTIC AEROSOLS THROUGH ENDOTRACHEAL-TUBES
    AHRENS, RC
    RIES, RA
    POPENDORF, W
    WIESE, JA
    [J]. PEDIATRIC PULMONOLOGY, 1986, 2 (01) : 19 - 26
  • [2] Effect of Tidal Volume and Nebulizer Type and Position on Albuterol Delivery in a Pediatric Model of Mechanical Ventilation
    Berlinski, Ariel
    Willis, J. Randy
    [J]. RESPIRATORY CARE, 2015, 60 (10) : 1424 - 1430
  • [3] Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model
    Berlinski, Ariel
    Willis, J. Randy
    [J]. RESPIRATORY CARE, 2013, 58 (07) : 1124 - 1133
  • [4] Berlinski A, 2010, RESP CARE, V55, P1699
  • [5] Branson RD, 2007, RESP CARE, V52, P1328
  • [6] Pediatric ARDS
    Cheifetz, Ira M.
    [J]. RESPIRATORY CARE, 2017, 62 (06) : 718 - 731
  • [7] Deakins Kathleen, 2002, Respir Care, V47, P1162
  • [8] Aerosol delivery with two ventilation modes during mechanical ventilation: a randomized study
    Dugernier, Jonathan
    Reychler, Gregory
    Wittebole, Xavier
    Roeseler, Jean
    Depoortere, Virginie
    Sottiaux, Thierry
    Michotte, Jean-Bernard
    Vanbever, Rita
    Dugernier, Thierry
    Goffette, Pierre
    Docquier, Marie-Agnes
    Raftopoulos, Christian
    Hantson, Philippe
    Jamar, Francois
    Laterre, Pierre-Francois
    [J]. ANNALS OF INTENSIVE CARE, 2016, 6
  • [9] Getinge, User's manual Servo-i Ventilator System V8.0
  • [10] Adjunct Therapies During Mechanical Ventilation: Airway Clearance Techniques, Therapeutic Aerosols, and Gases
    Kallet, Richard H.
    [J]. RESPIRATORY CARE, 2013, 58 (06) : 1053 - 1071