Design of Unilateral Complete Presurgical Nasoalveolar Molding (PNAM) Corrector Based on Feature Points Extraction of Complex 3D Surface

被引:1
作者
Li, Li [1 ]
Liu, Tao [1 ]
Fang, Dongshen [1 ]
机构
[1] Wuhan Univ, Sch Elect Informat, Wuhan 430072, Peoples R China
来源
APPLIED SCIENCES-BASEL | 2023年 / 13卷 / 09期
关键词
unilateral complete cleft lip and palate; computer-aided design; PNAM correction; feature points extraction; BILATERAL CLEFT-LIP; PALATE; INFANTS; TECHNOLOGY;
D O I
10.3390/app13095251
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Cleft lip and palate is a congenital maxillofacial deformity. Unilateral complete cleft lip and palate is one of the most common clinical types. Nasal alveolar molding (PNAM) is a recognized strategy for the treatment of cleft lip and palate. However, the current design of PNAM devices mainly relies on the subjective experience of doctors. The purpose of this paper is to describe the design and manufacture of a new computer-aided design appliance, which can be applied to the presurgical nasoalveolar molding of unilateral complete cleft lip and palate, eliminate individual differences, and improve production efficiency. In this paper, seven feature points on the healthy side and the affected side are extracted by the method of Gaussian curvature and ridge line extraction, and the healthy side rotation and built-in model are designed by using these seven feature points, which can quickly generate eight treatment stages of PNAM. The correction effects of the PNAM appliance designed in this paper were compared with the original maxillary model and the clinical PNAM appliance (hand-made by subjective experience) from the aspects of alveolar fissure width and symmetry. The PNAM appliance designed in this paper can effectively improve the symmetry of patients with unilateral complete cleft lip and palate (morphological similarity: t = 3.250, p = 0.01; length similarity: t = 1.559, p = 0.150) and reduce the width of alveolar cleft (t = 8.330, p < 0.01). This can fully achieve the therapeutic effect of PNAM appliances prepared by experienced doctors and is more efficient. The method based on complex 3D surface feature point extraction can provide the basis for the design and evaluation of a unilateral complete PNAM correction model, improve the design and production efficiency of unilateral complete cleft lip and palate appliance, eliminate the design problems caused by individual differences, and reduce the burden of doctors.
引用
收藏
页数:15
相关论文
共 27 条
  • [1] The Effects of Presurgical Nasoalveolar Molding on the Midface Symmetry of Children with Unilateral Cleft Lip and Palate: A Long-term Follow-up Study
    AlHayyan, Wasmiya A.
    Pani, Sharat Chandra
    AlJohar, Aziza J.
    AlQatami, Fawzi M.
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY-GLOBAL OPEN, 2018, 6 (07)
  • [2] RapidNAM: generative manufacturing approach of nasoalveolar molding devices for presurgical cleft lip and palate treatment
    Bauer, Franz Xaver
    Schoenberger, Markus
    Gattinger, Johannes
    Eblenkamp, Markus
    Wintermantel, Erich
    Rau, Andrea
    Guell, Florian Dieter
    Wolff, Klaus-Dietrich
    Loeffelbein, Denys J.
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2017, 62 (04): : 407 - 414
  • [3] Braumann B, 2002, CLEFT PALATE-CRAN J, V39, P1, DOI 10.1597/1545-1569(2002)039<0001:TDAOMC>2.0.CO
  • [4] 2
  • [5] Effects of pre-surgical nasoalveolar moulding on maxillary arch and nasal form in unilateral cleft lip and palate before lip surgery
    Fuchigami, T.
    Kimura, N.
    Kibe, T.
    Tezuka, M.
    Amir, M. S.
    Suga, H.
    Takemoto, Y.
    Hashiguchi, M.
    Maeda-Iino, A.
    Nakamura, N.
    [J]. ORTHODONTICS & CRANIOFACIAL RESEARCH, 2017, 20 (04) : 209 - 215
  • [6] Correction of maxillary deformity in infants with bilateral cleft lip and palate using computer-assisted design
    Gong, Xin
    Yu, Quan
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2012, 114 (05): : S74 - S78
  • [7] Vector of device placement and trajectory of mandibular distraction
    Grayson, BH
    McCormick, S
    Santiago, PE
    McCarthy, JG
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 1997, 8 (06) : 473 - 480
  • [8] Grayson BH, 1999, CLEFT PALATE-CRAN J, V36, P486, DOI 10.1597/1545-1569_1999_036_0486_pnmiiw_2.3.co_2
  • [9] GRAYSON BH, 1993, PLAST RECONSTR SURG, V92, P1422
  • [10] A semi-automated virtual workflow solution for the design and production of intraoral molding plates using additive manufacturing: the first clinical results of a pilot-study
    Grill, Florian D.
    Ritschl, Lucas M.
    Bauer, Franz X.
    Rau, Andrea
    Gau, Dominik
    Roth, Maximilian
    Eblenkamp, Markus
    Wolff, Klaus-Dietrich
    Loeffelbein, Denys J.
    [J]. SCIENTIFIC REPORTS, 2018, 8