A simple and portable vacuum assisted headspace solid phase microextraction device coupled to gas chromatography based on covalent organic framework/metal organic framework hybrid for simultaneous analysis of volatile and semi-volatile compounds in soil

被引:13
|
作者
Maleki, Sara [1 ]
Hashemi, Payman [1 ]
Adeli, Mohsen [1 ]
机构
[1] Lorestan Univ, Fac Chem, Dept Analyt Chem, Khorramabad, Iran
关键词
In-syringe vacuum assisted-HS-SPME; Polycyclic aromatic hydrocarbons; BTEX; covalent triazine-based; framework; metal-organic frameworks; hybrid; POLYCYCLIC AROMATIC-HYDROCARBONS; LIQUID-LIQUID MICROEXTRACTION; DEEP EUTECTIC SOLVENT; EXTRACTION; WATER; BTEX; EFFICIENT; PRESSURE; FIBER; NANOCOMPOSITE;
D O I
10.1016/j.chroma.2023.464195
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Various microextraction methods have demonstrated a positive effect when assisted by vacuum. However, working with such systems is often laborious, they often require expensive and non-portable vacuum pumps, and may even suck off some sample vapor or solid particles during the evacuation process. To address these issues, a simple, and affordable vacuum-assisted headspace solid-phase microextraction (HS-SPME) device was developed in this study. The device, named In Syringe Vacuum-assisted HS-SPME (ISV-HS-SPME), utilizes an adjustable 40 mL glass syringe as a vacuum provider and sampling vessel. A new fiber coating, made from a hybrid of covalent triazine-based frameworks and metal-organic frameworks (COF/MOF), was prepared and characterized by Fourier transform infrared spectrometry, field emission scanning electron microscopy, energy dispersive X-ray, X-ray diffraction, thermogravimetric analysis, and Brunauer-Emmett-Teller techniques for use in the ISV-HS-SPME. By optimizing parameters such as extraction temperature, extraction time, desorption temperature, desorption time, and, humidity using a simplex method, the ISV system was found to increase the extraction efficiency of polycyclic aromatic hydrocarbons (PAHs) and benzene, toluene, ethylbenzene, and xylenes (BTEX) in solid samples by up to 175%. The determinations were followed by GC-FID measurements. Compared to three commercially available fibers, the ISV-HS-SPME device with the COF/MOF (2DTP/MIL-101-Cr) fiber exhibited significantly higher peak areas for PAHs and BTEX. The linear dynamic ranges for BTEX and PAHs were 7.1- 90 0 0 ng g -1 and 0.23-90 0 0 ng g -1 , respectively, with limits of detection ranging from 2.1-5 ng g -1 for BTEX and 0.07-1.6 ng g -1 for PAHs. The relative standard deviation of the method was 2.6-7.8% for BTEX and 1.6-6.7% for PAHs. The ISV-HS-SPME was successfully used to simultaneously determine PAHs and BTEX in polluted soil samples with recoveries ranging from 80.4 to 108%.& COPY; 2023 Published by Elsevier B.V.
引用
收藏
页数:15
相关论文
共 34 条
  • [1] Optimization and validation of headspace solid-phase microextraction method coupled with gas chromatography-triple quadrupole tandem mass spectrometry for simultaneous determination of volatile and semi-volatile organic compounds in coking wastewater treatment plant
    Saber, Ayman N.
    Zhang, Haifeng
    Yang, Min
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2019, 191 (07)
  • [2] Metal organic framework MIL-101 coated fiber for headspace solid phase microextraction of volatile aromatic compounds
    Zang, Xiaohuan
    Zhang, Guijiang
    Chang, Qingyun
    Zhang, Xi
    Wang, Chun
    Wang, Zhi
    ANALYTICAL METHODS, 2015, 7 (03) : 918 - 923
  • [3] Simultaneous Extraction and Determination of Volatile Organic Compounds and Semi-volatile Organic Compounds in Indoor Air Using Multi-bed Solid Phase Extraction Device
    Ueta, Ikuo
    Takenaka, Risa
    Fujimura, Koji
    Yoshimura, Tomotaka
    Narukami, Shoji
    Mochizuki, Suguru
    Maeda, Tsuneaki
    ANALYTICAL SCIENCES, 2020, 36 (09) : 1071 - 1074
  • [4] Simple, Low-Cost and Reliable Device for Vacuum-Assisted Headspace Solid-Phase Microextraction of Volatile and Semivolatile Compounds from Complex Solid Samples
    Beiranvand, Mohammad
    Ghiasvand, Alireza
    CHROMATOGRAPHIA, 2017, 80 (12) : 1771 - 1780
  • [5] Gas-cycle-assisted headspace solid-phase microextraction coupled with gas chromatography for rapid analysis of organic pollutants
    Zhu, Wenli
    Qin, Peige
    Han, Lizhen
    Zhang, Xiaowan
    Li, Dan
    Li, Mengyuan
    Wang, Youmei
    Zhang, Xuebin
    Lu, Minghua
    Cai, Zongwei
    CHEMICAL COMMUNICATIONS, 2021, 57 (70) : 8810 - 8813
  • [6] Comparison of Volatile Organic Compounds of Thymus vulgaris Using Hydrodistillation and Headspace Solid Phase Microextraction Gas Chromatography Mass Spectrometry
    Nezhadali, Azizollah
    Akbarpour, Mina
    Shirvan, Batool Zarrabi
    Mousavi, Mehdi
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2010, 57 (01) : 40 - 43
  • [7] Analysis of volatile organic compounds in environmental water samples and soil gas by solid-phase microextraction
    Nilsson, T
    Montanarella, L
    Baglio, D
    Tilio, R
    Bidoglio, G
    Facchetti, S
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1998, 69 (03) : 217 - 226
  • [8] Analysis of semi-volatile organic compounds in indoor suspended particulate matter by thermal desorption coupled with gas chromatography/mass spectrometry
    Mercier, Fabien
    Glorennec, Philippe
    Blanchard, Olivier
    Le Bot, Barbara
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1254 : 107 - 114
  • [9] Reduced graphene-decorated covalent organic framework as a novel coating for solid-phase microextraction of phthalate esters coupled to gas chromatography-mass spectrometry
    Khataei, Mohammad Mahdi
    Yamini, Yadollah
    Ghaemmaghami, Mostafa
    MICROCHIMICA ACTA, 2020, 187 (04)
  • [10] Development of a Headspace-Solid Phase Micro Extraction Method for the Analysis of Volatile and Semi-Volatile Organic Compounds from Polyurethane Resins for Leather Finishing
    Flores, Antonia
    Sorolla, Silvia
    Casas, Concepcio
    Cuadros, Rosa
    Bacardit, Anna
    JOURNAL OF THE AMERICAN LEATHER CHEMISTS ASSOCIATION, 2021, 116 (08): : 290 - 297